Skip to content

Cotton Cannulatome II G22537Recalls & reported problems

Cook Incorporated · Cotton Cannulatome II PC Double Lumen Sphincterotome · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002225374), received between Mar 17, 2016 and Nov 11, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation3
  • Failure to Align1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2025Material Separation; Material Separation

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY [ERCP], THE PHYSICIAN USED A COOK COTTON CANNULATOME II PC DOUBLE LUMEN SPHINCTEROTOME. IT WAS REPORTED THAT THE CUTTING WIRE WAS BROKEN SO THE PHYSICIAN CHANGED TO ANOTHER ONE. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. THE DEVICE WAS EVALUATED ON 16OCT2025 AND IT WAS DETERMINED THAT THE ANCHOR HAD SEPARATED FROM THE TIP [SUBJECT OF REPORT]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23527960 ↗
  • MalfunctionJun 10, 2020Material Separation; Material Separation

    PRIOR TO AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE NURSE ASSISTANT PREPARED A COOK COTTON CANNULATOME II PC DOUBLE LUMEN SPHINCTEROTOME. THE NURSE ASSISTANT THAT OPENED THE PACKAGE NOTICED THE CUTTING WAS SEPARATED AT THE SOLDER JOINT INSIDE THE TIP OF THE CATHETER AND WAS STARTING TO COME OUT OF THE CATHETER [CUTTING WIRE SECURING COMPONENT SEPARATION]. IT WAS NEVER ADVANCED INTO THE ENDOSCOPE. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 10139096 ↗
  • MalfunctionMar 28, 2016Failure to Align; Failure to Align

    THIS FOLLOW UP MEDWATCH REPORT IS BEING SUBMITTED TO PROVIDE NEW INFORMATION [DEVICE EVALUATION] AND CORRECTED INFORMATION. DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. DURING THE ARC VOLTAGE [APPLICATION OF CURRENT AND BOWING OF THE TIP], THE SPHINCTEROTOME DEFLECTS TO THE LEFT [INCORRECT CUTTING WIRE ORIENTATION]. EIGHT (8) SPHINCTEROTOMES WERE USED IN ORDER TO CARRY OUT THE PROCEDURE. FIVE (5) OF THE SPHINCTEROTOMES USED IN THE PROCEDURE ARE INCLUDED IN THIS REPORT. SEE RELATED MEDWATCH REPORT NUMBERS 1037905-2016-00066 & 1037905-2016-00067.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5528972 ↗
  • MalfunctionMar 17, 2016Material Separation; Material Separation

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK COTTON CANNULATOME II PC DOUBLE LUMEN SPHINCTEROTOME. UPON ACTIVATING CUTTING ENERGY [CUTTING FUNCTION OF THE ELECTROSURGICAL GENERATOR], THE CUTTING WIRE GOT TORN FROM ONE SIDE [CUTTING WIRE BROKE IN ONE LOCATION]. ON (B)(4) 2016, WE RECEIVED THE FOLLOWING INFORMATION: THE CUTTING WIRE ANCHOR DISCONNECTED AT THE PATIENT END OF DEVICE. NO SECTION OF THE DEVICE REMAINED IN THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5507032 ↗

Other Unit, electrosurgical, endoscopic (with or without accessories) devices

All Unit, electrosurgical, endoscopic (with or without accessories) devices →

Sources and updates