Tri-Tome pc G22555Recalls & reported problems

Cook Incorporated · Tri-Tome PC Triple Lumen Sphincterotome · Class 2

All models of Tri-Tome pc (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Cook Tri-Tome PC triple lumen sphincterotome is a surgical tool used to cut tissue during ERCP procedures on the bile and pancreatic ducts. The 47 reports mostly describe the cutting wire breaking or detaching, and the bent tip not aligning as expected, so physicians swapped it out and finished with another device. No deaths and no injuries were reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
44
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 00827002225558), received between Feb 22, 2016 and Jun 4, 2025 — classified as 44 malfunctions.

The most frequently reported specific device problem is “Failure to Align” (36 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Align36
  • Material Separation4
  • Break2
  • Device Operates Differently Than Expected1
  • Failure to Cut1
  • Use of Device Problem1

Patient problems

  • No Consequences Or Impact To Patient33
  • No Known Impact Or Consequence To Patient4
  • No Patient Involvement4
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2025Material Separation; Material Separation

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. IT WAS INITIALLY REPORTED THAT THE DEVICE WAS TIGHTENED AND THE CUTTING WIRE BROKE/TORN. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. THE DEVICE WAS RECEIVED ON 9 MAY 2025. OUR EVALUATION OF THE RETURNED DEVICE ON 14 MAY 2025 DETERMINED THAT THE CUTTING WIRE SECURING COMPONENT (ANCHOR) HAS SEPARATED FROM THE CATHETER WITH DETACHMENT OF THE LONGER ANCHOR SEGMENT. HOWEVER, THE DETACHED SEGMENT WAS RETAINED WITHIN THE CATHETER AND ALL PORTIONS WERE INCLUDED IN THE RETURN. [SUBJECT OF REPORT] A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22142072 ↗
  • MalfunctionMay 8, 2025Material Separation; Material Separation

    DURING AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. IT WAS INITIALLY REPORTED THAT THE PHYSICIAN DETECTED THAT THE DEVICE WAS BROKEN AND IMMEDIATELY CHANGED TO A NEW ONE TO USE. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. OUR EVALUATION OF THE RETURNED DEVICE ON 15 APR 2025 DETERMINED THAT THE ANCHOR HAS SEPARATED FROM THE TIP. [SUBJECT OF REPORT]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21988145 ↗
  • MalfunctionMar 20, 2025Material Separation; Material Separation

    IN PREPARATION FOR A PROCEDURE, THE USER SELECTED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. IT WAS REPORTED THAT THE USER OPENED THE PACKAGE AND FOUND THE CUTTING WIRE TIP BROKEN. USER CHANGED TO ANOTHER OF THE SAME DEVICE TO COMPLETE THE PROCEDURE. THIS WAS INTERPRETED AS CUTTING WIRE BROKE IN ONE LOCATION WITHOUT DETACHMENT AND THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. PER OUR EVALUATION OF THE PRODUCT RETURNED ON 24 MAR 2025, THE CUTTING WIRE ANCHOR HAS SEPARATED FROM THE TIP OF THE CATHETER. [SUBJECT OF REPORT]. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21654374 ↗
  • MalfunctionDec 12, 2019Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED THREE (3) TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOMES. THE ALIGNMENT OF THE TIP WAS BENT OUT OF SHAPE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9460722 ↗
  • MalfunctionAug 12, 2019Use of Device Problem; Failure to Cut; Use of Device Problem; Failure to Cut

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. THE CUTTING WIRE WAS BROKEN DURING THE ERCP PROCEDURE. THE USER CHANGED TO ANOTHER OF THE SAME DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. ON (B)(6) 2019, ADDITIONAL INFORMATION WAS PROVIDED INDICATING THAT THE DEVICE ALLOWED COAGULATION/BURNING BUT DID NOT CUT PROPERLY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8885321 ↗
  • MalfunctionJul 11, 2019Failure to Align; Failure to Align

    IN PREPARATION FOR A PROCEDURE, THE USER SELECTED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. THE CUTTING WIRE IS TWISTED. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 8782408 ↗
  • MalfunctionMay 7, 2019Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. THE USER OPENED THE PACKAGE AND FLUSHED THE DEVICE WITH SUFFICIENT LUBRICATION. THE USER CONCERNS [CHECKED] THE ORIENTATION OF THE CUTTING WIRE AND FOUND OUT THE CUTTING WIRE ORIENTATION WAS POINTING AT 2 O'CLOCK INSIDE THE ENDOSCOPE [INCORRECT CUTTING WIRE ORIENTATION]. THE USER CANNOT ADJUST THE ORIENTATION AND RETRACTED THE DEVICE FROM THE ENDOSCOPE. THE USER CHANGED TO ANOTHER SAME DEVICE TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8588199 ↗
  • MalfunctionApr 2, 2019Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. WHEN PERFORMING THE USE [SPHINCTEROTOMY], A DEFECT WAS NOTED IN THE TIP THAT DOES NOT ALLOW CANNULATION OF THE PAPILLA IN THE PATIENT. THE TIP WAS NOTED TO GO TO ONE (1) SIDE AND NOT UPWARDS [INCORRECT CUTTING WIRE ORIENTATION] AS IS THE MECHANISM OF ACTION. FOR THIS REASON, ANOTHER PAPILLOTOME WAS OPENED WITH WHICH THE PROCEDURE WAS PERFORMED. OUR ATTEMPTS TO COLLECT SPECIFIC PATIENT OUTCOME INFORMATION WERE UNSUCCESSFUL. BECAUSE THE COMPLAINANT STATED THAT THERE WAS NO ADVERSE EFFECTS TO THE PATIENT, WE CAN CONCLUDE THAT THIS INFORMATION DOES NOT REASONABLY SUGGEST THE PATIENT WAS ADVERSELY IMPACTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8474967 ↗
  • MalfunctionJun 29, 2018Material Separation; Material Separation

    IN PREPARATION FOR A PROCEDURE, THE USER SELECTED A COOKTRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. WHEN THE NURSE OPENED THE PACKAGE, IT WAS FOUND THAT A SECTION OF CUTTING WIRE FELL OFF [CUTTING WIRE SECURING COMPONENT SEPARATED]. THE PROCEDURE WAS FINISHED BY CHANGING TO ANOTHER DEVICE. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 7652534 ↗
  • MalfunctionNov 13, 2017Failure to Align; Failure to Align

    IN PREPARATION FOR A PROCEDURE, THE USER SELECTED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. AS REPORTED TO CUSTOMER RELATIONS: "THE SPHINCTEROTOME WAS TWISTED IMMEDIATELY AFTER BEING EXPOSED [INCORRECT CUTTING WIRE ORIENTATION]."

    Patient problems: No Patient Involvement

    MAUDE report 7026313 ↗
  • MalfunctionNov 13, 2017Failure to Align; Failure to Align

    IN PREPARATION FOR A PROCEDURE, THE USER SELECTED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. AS REPORTED TO CUSTOMER RELATIONS: "THE SPHINCTEROTOME WAS TWISTED IMMEDIATELY AFTER BEING EXPOSED [INCORRECT CUTTING WIRE ORIENTATION]."

    Patient problems: No Patient Involvement

    MAUDE report 7026304 ↗
  • MalfunctionOct 9, 2017Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. AS REPORTED TO CUSTOMER RELATIONS: WHEN USING A TRI-25M ON (B)(6) 2017, THE DOCTOR NOTICED THAT THE OUTPUT OF THE CUTTING WIRE WAS LATERALIZED [MALFUNCTION UNKNOWN FOLLOWING ANALYSIS OF PICTURE PROVIDED BY USER], THUS MAKING THE PAPILLOTOMY DIFFICULT AND MAKING IT IMPOSSIBLE TO USE. THE FOLLOWING CLARIFICATION WAS RECEIVED ON (B)(6) 2017 - THE PHYSICIAN NOTICED THAT THE TIP OF THE SPHINCTEROTOMY WAS TURNED TO AN UNUSUAL POSITION, MAKING THE CANNULATION HARDER [INCORRECT CUTTING WIRE ORIENTATION].

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6928992 ↗
  • MalfunctionSep 15, 2017Failure to Align; Failure to Align

    DURING A ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) THE PHYSICIAN USED A TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. THE TIP OF THE SPHINCTEROTOME IS TWISTED [INCORRECT CUTTING WIRE ORIENTATION]. THE FOLLOWING ADDITIONAL INFORMATION WAS RECEIVED ON 9/1/17: THE ORIENTATION OF SPHINCTEROTOME DID NOT LET THE DOCTOR DO THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6867941 ↗
  • MalfunctionJun 9, 2017Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. THE USER TRIED TO DO AN ERCP. THE PHYSICIAN, STATED THE SPHINCTEROTOME HAD VERY BAD ORIENTATION AND COULD NOT BE USED, VERY TWISTED. THE PATIENT HAD A VERY DISTORTED ANATOMY. ANOTHER DEVICE WAS USED WITH NO PROBLEMS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6629921 ↗
  • MalfunctionJun 2, 2017Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. THE ORIENTATION POINTED TO THE 9 O'CLOCK [POSITION], REAL WRONG ORIENTATION [INCORRECT CUTTING WIRE ORIENTATION].

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6608821 ↗
  • MalfunctionMar 20, 2017Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED TWO (2) COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOMES. AFTER INTRODUCING THE PRODUCT INTO THE ENDSOSCOPE, IT CAME OUT WRONG [INCORRECT CUTTING WIRE ORIENTATION].

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6417848 ↗
  • MalfunctionMar 20, 2017Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED TWO COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOMES. AFTER INTRODUCING THE PRODUCT INTO THE ENDOSCOPE, IT CAME OUT WRONG [INCORRECT CUTTING WIRE ORIENTATION].

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6417845 ↗
  • MalfunctionMar 16, 2017Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. THE DEVICE PRESENTS MIS-ORIENTATION [INCORRECT CUTTING WIRE ORIENTATION].

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6411352 ↗
  • MalfunctionMar 16, 2017Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. THERE WAS ORIENTATION PROBLEM. THE ORIENTATION WAS NOT CORRECT [INCORRECT CUTTING WIRE ORIENTATION].

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6410783 ↗
  • MalfunctionJan 10, 2017Break; Break

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. AS REPORTED TO CUSTOMER RELATIONS: DURING THE PAPILLOTOMY, THE PAPILLOTOME'S CUTTING WIRE RUPTURED [CUTTING WIRE BROKE IN TWO LOCATIONS], MAKING IT IMPOSSIBLE TO USE. IT WAS NECESSARY TO USE THE PRE-CUT PAPILLOTOME [NEEDLE KNIFE]. THE FOLLOWING CLARIFICATION WAS RECEIVED 1/06/2017 FROM THE DISTRIBUTOR: "[IN] THE MEDICAL REPORT WE RECEIVED ON THIS CASE, THERE WAS NO INFORMATION FROM WHICH PART OF THE DEVICE WAS INSIDE THE PATIENT OR EVEN THAT AN ADDITIONAL PROCEDURE WAS REQUIRED. WE SOON CONCLUDED THAT THERE WAS NO HARM TO THE PATIENT."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6236853 ↗
  • MalfunctionNov 11, 2016Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. ONCE INTRODUCED INTO THE ENDOSCOPE, THE SPHINCTEROTOME ORIENTATION WAS TO THE LEFT [INCORRECT CUTTING WIRE ORIENTATION].

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6096796 ↗
  • MalfunctionNov 11, 2016Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED TWO (2) COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOMES. ONCE INTRODUCED INTO THE ENDOSCOPE, BOTH OF THE SPHINCTEROTOMES' ORIENTATION WAS TO THE LEFT [INCORRECT CUTTING WIRE ORIENTATION].

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6096795 ↗
  • MalfunctionOct 14, 2016Break; Break

    THIS CORRECTION FOLLOW UP REPORT IS BEING SENT TO CANCEL THE INITIAL REPORT SUBMITTED RELATING TO THIS EVENT. PLEASE REFERENCE THE ADDITIONAL MANUFACTURER NARRATIVE NOTES SECTION FOR THIS JUSTIFICATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6029595 ↗
  • MalfunctionAug 26, 2016Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. AS REPORTED TO CUSTOMER RELATIONS: "WHEN TRYING TO MAKE CANNULATION, THE TIP OF THE SPHINCTEROTOME IS CROOKED, YOU CAN NOT MAKE THE CUT." [INCORRECT CUTTING WIRE ORIENTATION/TIP POINTED WRONG DIRECTION].

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5906234 ↗
  • MalfunctionAug 9, 2016Failure to Align; Failure to Align

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME. THE SPHINCTEROTOME CAME OUT [OF THE ENDOSCOPE] WITH BAD ORIENTATION. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5856970 ↗

Other Unit, electrosurgical, endoscopic (with or without accessories) devices