EchoTip G31518Recalls & reported problems
Cook Ireland Ltd · Echotip Ultra, Endoscopic Ultrasound Needle · Class 2
All models of EchoTip (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00827002315181), received between Nov 15, 2017 and Jul 26, 2019 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Retraction Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Retraction Problem4
- Bent1
- Break1
- Kinked1
- Material Deformation1
Patient problems
- No Known Impact Or Consequence To Patient7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 26, 2019Material Deformation; Material Deformation
DURING A PUNCTURE WITH 20G NEEDLE, NURSE + SURGEON ENCOUNTER GREAT DIFFICULTY TO PULL THE WIRE GUIDE. THE TIP TO ALLOW THE GRIP ON THE WIRE GUIDE BROKE. DEVICE REMOVED AND CHANGED WITH A 22G NEEDLE: SAME ISSUE. RESISTANCE OF THE WIRE GUIDE AND SEPARATION OF THE TIP. A SMALL SAMPLE WAS RETRIEVED WITH THE 2ND NEEDLE, THE BIOPSY WAS STOPPED. THE TEAM EMPHASISES T GUIDES ARE MORE AND MORE DIFFICULT TO PULL, IT IS THE 1ST TIME THE TIP OF THE GUIDES BREAKS. "AS PER COMPLAINT FORM": DIFFICULTY TO PULL THE STYLET , THE TIP OF STYLET WAS BROKEN THE DEVICE WAS EVALUATED ON THE 28-JUN-19 CONFIRMING THE NEEDLE WAS KINKED PROXIMALLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8830511 ↗MalfunctionDec 13, 2018Break; Break
AS PER CUSTOMER COMPLAINT FORM "WHILE DOING EUS-FNA, HANDLE OF EUS-FNA NEEDLE GOT DISENGAGED FROM THE ASSEMBLY".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8159199 ↗MalfunctionJun 14, 2018Retraction Problem; Retraction Problem
AS REPORTED TO CUSTOMER RELATIONS: "NEEDLES AND STICKING DURING ULTRASOUND PROCEDURES, WHEN TRYING TO ACTUATE THE NEEDLE THROUGH THE ENDOSCOPE AND OUT OF THE CATHETER. PROCEDURES COMPLETED BY USING ANOTHER DEVICE (MAY HAVE BEEN COOK, DIFFERENT GPN, MAY HAVE BEEN COMPETITOR DEVICE)." ADDITIONAL INFORMATION PROVIDED BY (B)(6) ON (B)(6) 2018: "I WAS NOT IN THE CASES AND CUSTOMER REPORTED TO ME INFORMATION I IN TURN REPORTED TO YOU. CUSTOMER DOES NOT HAVE FURTHER DETAIL. I WISH I COULD BE OF MORE HELP." ADDITIONAL INFORMATION RECEIVED: "WAS THE NEEDLE FULLY RETRACTED WHEN THE DEVICE WAS REMOVED FROM THE PATIENT? UNKNOWN."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7600882 ↗MalfunctionJun 14, 2018Retraction Problem; Retraction Problem
AS REPORTED TO CUSTOMER RELATIONS: "NEEDLES AND STICKING DURING ULTRASOUND PROCEDURES, WHEN TRYING TO ACTUATE THE NEEDLE THROUGH THE ENDOSCOPE AND OUT OF THE CATHETER. PROCEDURES COMPLETED BY USING ANOTHER DEVICE (MAY HAVE BEEN COOK, DIFFERENT GPN, MAY HAVE BEEN COMPETITOR DEVICE)." ADDITIONAL INFORMATION PROVIDED BY DM ON 24MAY2018: "I WAS NOT IN THE CASES AND CUSTOMER REPORTED TO ME INFORMATION I IN TURN REPORTED TO YOU. CUSTOMER DOES NOT HAVE FURTHER DETAIL. I WISH I COULD BE OF MORE HELP." ADDITIONAL INFORMATION RECEIVED "WAS THE NEEDLE FULLY RETRACTED WHEN THE DEVICE WAS REMOVED FROM THE PATIENT? UNKNOWN".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7600803 ↗MalfunctionJun 14, 2018Retraction Problem; Retraction Problem
AS REPORTED TO CUSTOMER RELATIONS: "NEEDLES AND STICKING DURING ULTRASOUND PROCEDURES, WHEN TRYING TO ACTUATE THE NEEDLE THROUGH THE ENDOSCOPE AND OUT OF THE CATHETER. PROCEDURES COMPLETED BY USING ANOTHER DEVICE (MAY HAVE BEEN COOK, DIFFERENT GPN, MAY HAVE BEEN COMPETITOR DEVICE)." ADDITIONAL INFORMATION PROVIDED BY DM ON 24MAY2018: "I WAS NOT IN THE CASES AND CUSTOMER REPORTED TO ME INFORMATION I IN TURN REPORTED TO YOU. CUSTOMER DOES NOT HAVE FURTHER DETAIL. I WISH I COULD BE OF MORE HELP." ADDITIONAL INFORMATION RECEIVED: "WAS THE NEEDLE FULLY RETRACTED WHEN THE DEVICE WAS REMOVED FROM THE PATIENT? UNKNOWN".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7600766 ↗MalfunctionJun 14, 2018Retraction Problem; Retraction Problem
AS REPORTED TO CUSTOMER RELATIONS: "NEEDLES AND STICKING DURING ULTRASOUND PROCEDURES, WHEN TRYING TO ACTUATE THE NEEDLE THROUGH THE ENDOSCOPE AND OUT OF THE CATHETER. PROCEDURES COMPLETED BY USING ANOTHER DEVICE (MAY HAVE BEEN COOK, DIFFERENT GPN, MAY HAVE BEEN COMPETITOR DEVICE)." ADDITIONAL INFORMATION PROVIDED BY DM ON 24MAY2018: "I WAS NOT IN THE CASES AND CUSTOMER REPORTED TO ME INFORMATION I IN TURN REPORTED TO YOU. CUSTOMER DOES NOT HAVE FURTHER DETAIL. I WISH I COULD BE OF MORE HELP." ADDITIONAL INFORMATION RECEIVED: "WAS THE NEEDLE FULLY RETRACTED WHEN THE DEVICE WAS REMOVED FROM THE PATIENT? UNKNOWN".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7600764 ↗MalfunctionNov 15, 2017Bent; Kinked; Bent; Kinked
THIS FOLLOW UP MDR IS BEING SUBMITTED 30 DAYS FROM THE INITIAL REPORT TO INCLUDE DEVICE EVALUATION DETAIL UPDATES TO THE INVESTIGATION. THE NEEDLE IS BENT AND SHEATH DENTED. "AS PER COMPLAINT FORM": WHEN UNPACKING THE TEAM FOUND THAT THE SHEATH ADAPTER WASN'T EASY TO PUSH, WHEN LOOKING CLOSER THEY FOUND THE SHEATH SLIGHTLY BENT AND DENTED. THIS WAS THERE BEFORE UNPACKING THE NEEDLE. THEY TRIED TO PUSH THE NEEDLE THROUGH THE WORKING CHANNEL, BUT WITHOUT SUCCESS. THE STYLET COULD ONLY BE PULLED WITH RESISTANCE AND THEY WERE NOT ABLE TO REINSERT IT. A SECOND BEND (CLOSER TO THE TIP) HAPPENED, WHEN THEY REPACKED THE NEEDLE, SO IT HAS NOTHING TO DO WITH THIS COMPLAINT. ADDITIONAL INFORMATION RECEIVED 26OCT 2017"THE NEEDLE HASN'T BEEN USED IN THE PATIENT. IN THIS HOSPITAL THEY ARE INTRODUCING THE NEEDLE IN THE WORKING CHANNEL BEFORE THEY INTRODUCE THE SCOPE IN THE PATIENT. THEY FOUND THE FIRST KINK (THE ONE BELOW THE SHEATH ADJUSTOR) WHEN UNPACKING THE NEEDLE AND FOUND IT HARD TO MOVE THE STYLET. THE SECOND KINK (CLOSE TO THE DISTAL END OF THE NEEDLE) HAPPENED WHEN THEY TRIED TO REPACK THE NEEDLE IN THE PACKAGING AFTER THEY FOUND OUT IT WAS NOT USABLE. " ADDITIONAL INFORMATION RECEIVED -
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7032790 ↗
FDA clearance and classification
The FDA regulates product code FCG (biopsy needle) under 21 CFR 876.1075 as a Class II device. 21 CFR 876.1075 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the EchoTip G31518
Has the EchoTip G31518 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002315181), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the EchoTip G31518?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Nov 15, 2017 and Jul 26, 2019. In those reports, the most common specific device problems are “Retraction Problem” (4), “Bent” (1) and “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EchoTip G31518?
None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the EchoTip G31518, and is it still sold?
The FDA's device database lists Cook Ireland Ltd as the company responsible for it (brand EchoTip, model G31518). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Biopsy needle devices
- MAGNUM MN1410Bard Peripheral Vascular, Inc. · Biopsy needle1 recall119 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4021Olympus Medical Systems Corp. · Biopsy needle1 recall94 reports
- EchoTip G31520Cook Ireland Ltd · Biopsy needle1 recall84 reports
- Single Use Aspiration Needle NA-201SX-4021Olympus Medical Systems Corp. · Biopsy needle1 death126 reports
- EchoTip ProCore G34785Cook Ireland Ltd · Biopsy needle97 reports
- MAGNUM MN1820Bard Peripheral Vascular, Inc. · Biopsy needle1 recall37 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4022Olympus Medical Systems Corp. · Biopsy needle1 recall28 reports
- Interject M00518250Boston Scientific Corporation · Biopsy needle70 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page