Skip to content

Arndt G44114Recalls & reported problems

Cook Incorporated · Arndt Endobronchial Blocker Set With Spherical Balloon · Class 2

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002441149), received between Nov 10, 2016 and Feb 17, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inflation Problem5
  • Material Rupture4
  • Deflation Problem3
  • Gas/Air Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 17, 2026Material Rupture

    ADDITIONAL INFORMATION WAS PROVIDED ON 25FEB2026. THE PATIENT DID NOT HAVE TORTUOUS ANATOMY. IT IS UNKNOWN IF THE DEVICE WAS TEST INFLATED BEFORE INSERTION BECAUSE IT WAS AN EMERGENCY AND NO ONE IS ABLE TO REMEMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24367763 ↗
  • MalfunctionJan 2, 2026Deflation Problem; Inflation Problem

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23945898 ↗
  • MalfunctionDec 10, 2025Inflation Problem; Inflation Problem

    NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23768193 ↗
  • MalfunctionJul 22, 2025Inflation Problem; Material Rupture; Inflation Problem; Material Rupture

    THE DEVICE WAS RETURNED FOR INVESTIGATION ON 13NOV2025. DURING PRELIMINARY DEVICE EVALUATION, IT WAS DETERMINED THE BALLOON OF THE ARNDT ENDOBRONCHIAL BLOCKER CATHETER WAS TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22566493 ↗
  • MalfunctionJun 24, 2025Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT THE BALLOON FROM AN ARNDT ENDOBRONCHIAL BLOCKER SET LEAKED. THE DEVICE WAS REQUIRED FOR A CRYOPULMONARY BIOPSY. THE BALLOON WAS TEST INFLATED WITH 4 ML WHEN THE LEAK WAS DISCOVERED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER, NEW DEVICE AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22322595 ↗
  • MalfunctionJun 10, 2025Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT THE BLOCKER BALLOON FROM AN ARNDT ENDOBRONCHIAL BLOCKER SET WOULD NOT RETAIN AIR. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22179660 ↗
  • MalfunctionMar 26, 2025Inflation Problem; Material Rupture; Inflation Problem; Material Rupture

    IT WAS REPORTED THAT THE BALLOON INCLUDED IN THE ARNDT ENDOBRONCHIAL BLOCKER SET FAILED TO INFLATE AND LEAKED DURING AN AIRWAY CRYOBIOPSY PROCEDURE IN A PATIENT OF UNKNOWN AGE AND GENDER. THE BALLOON WAS INFLATED WITH 5ML OF AIR. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A NEW DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21701082 ↗
  • MalfunctionMar 20, 2025Material Rupture; Gas/Air Leak; Material Rupture; Gas/Air Leak

    IT WAS REPORTED THAT AN ARNDT ENDOBRONCHIAL BLOCKER SET WAS REQUIRED FOR A FROZEN LUNG BIOPSY PROCEDURE. DURING DEVICE TESTING, PRIOR TO PATIENT CONTACT, THE BALLOON WAS FILLED WITH 10ML OF AIR AND LEAKAGE WAS DISCOVERED. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21654175 ↗
  • MalfunctionNov 10, 2016Inflation Problem; Inflation Problem

    CUSTOMER REPORTED THAT A PATIENT UNDERWENT A PLANNED ESOPHAGECTOMY FOR AN ESOPHAGEAL REPAIR. AN ARNDT ENDOBRONCHIAL BLOCKER SET WAS USED TO SELECTIVELY BLOCK ONE LUNG DURING THE PROCEDURE. THE DEVICE WAS SET WITHIN THE MAIN BRONCHUS OF THE RIGHT LUNG AND REPORTEDLY FAILED TO INFLATE EVENLY. THE EVENT ALLEGEDLY CAUSED AN IMPRESSION / INDENTATION OF THE LUNG TISSUE ( BRONCHUS) WITH LOCALIZED CAPILLARY BLEEDING. NO INTERVENTION WAS NEEDED TO ALLEVIATE THE IMPRESSION OR TO CONTROL THE CAPILLARY OOZE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Consequences Or Impact To Patient

    MAUDE report 6092502 ↗

Other TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR) devices

All TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.