Cotton Cannulatome II G49196Recalls & reported problems
Cook Incorporated · Cotton Cannulatome II PC Protector · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002491960), received between Apr 8, 2016 and Nov 21, 2025 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
- Break1
- Failure to Cut1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2025Material Separation; Material Separation
DURING AN ENDOSCOPIC PROCEDURE IN THE DUODENUM AND BILE DUCT, THE PHYSICIAN USED A COOK COTTON CANNULATOME II PC PROTECTOR SPHINCTEROTOMES. IT WAS REPORTED THAT THE TIP OF THE CUTTING WIRE CAME UNDONE. [CUTTING WIRE SECURING COMPONENT SEPARATES FROM CATHETE-SUBJECT OF REPORT]. THE PAPILLA WAS LOCATED IN A DIVERTICULUM AND WAS ACTUALLY PROBED BLINDLY. AFTER UNSUCCESSFUL VISUALIZATION, THE PAPILLOTOME WAS REMOVED AND IT TURNED OUT THAT THE CUTTING WIRE HAD COME LOOSE. THE EXAMINATION WAS TERMINATED AND WILL THEN BE PERFORMED VIA PTCD [PERCUTANEOUS TRANSHEPATIC BILIARY DRAINAGE]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23619659 ↗MalfunctionJul 3, 2024Material Separation; Material Separation
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK COTTON CANNULATOME II PC PROTECTOR. IT WAS INITIALLY REPORTED THAT THE PHYSICIAN TOOK THE DEVICE AND A COMPETITOR'S WIRE GUIDE TO GET ACCESS TO DUODENAL PAPILLA, THEN WHEN THE PHYSICIAN WAS READY TO CUT THE DUODENAL PAPILLA TOWARD THE COMMON BILE DUCT DIRECTION, THE ORIENTATION OF CUTTING WIRE WAS DEVIATED TO PANCREAS DUCT DIRECTION, USER REPEATEDLY PULLING THE CUTTING WIRE WHILE THE CUTTING WIRE BROKEN. THE PHYSICIAN CHANGED TO ANOTHER DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. PER OUR PRODUCT EVALUATION ON 7 JUN 2024, IT WAS NOTED THAT THE CUTTING WIRE SECURING COMPONENT HAS SEPARATED FROM THE CATHETER WITHOUT DETACHMENT. [SUBJECT OF REPORT] A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19666427 ↗InjuryApr 8, 2016Break; Failure to Cut; Break; Failure to Cut
ON (B)(6) 2016, THE FOLLOWING WAS REPORTED TO A COOK REPRESENTATIVE FROM THE USER FACILITY: DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK COTTON CANNULATOME II PC PROTECTOR. THE CUTTING WIRE GOT TORN OFF AND DETACHED COMPLETELY FROM THE CATHETER. LATER, THE CUTTING WIRE WAS REMOVED WITH FORCEPS FROM THE DUODENUM. ON (B)(6) 2016, THE FOLLOWING WAS REPORTED TO COOK VIA A [(B)(6)] USER REPORT: IT WAS PLANNED TO REMOVE GALL STONES FROM A FEMALE PATIENT. FOR THIS A CUTTING WIRE (PAPILLOTOME) BY COOK WAS USED AMONG OTHERS, IN CONNECTION WITH A HIGH FREQUENCY MACHINE BY ERBE. CUTTING WAS NOT POSSIBLE, ONLY COAGULATION DESPITE CUTTING MODUS. FURTHERMORE, THE CUTTING WIRE TORE. THE TORN OFF PIECE WAS UNABLE TO BE RETRIEVED, EVEN WITH A FOREIGN-BODY FORCEPS; IT COULD NOT BE LOCATED. UPON REPLACEMENT OF THE CUTTING WIRE, NORMAL PAPILLOTOMY OF THE BILE DUCT SPHINCTER WAS POSSIBLE. PHOTOS OF THE TORN OFF CUTTING WIRE AND THE DEFECTIVE CUTTING WIRE ITSELF ARE WITH ME IN THE MEDICAL TECHNOLOGY DEPARTMENT. WE WILL NOTIFY COOK. WE HAVE REQUESTED THE PHOTOS OF THE CUTTING WIRE FROM THE USER FACILITY, BUT THEY HAVE NOT YET BEEN RECEIVED. ON (B)(6) 2016, WE
Patient problems: Device Embedded In Tissue or Plaque; No Consequences Or Impact To Patient
MAUDE report 5563590 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page