EchoTip G52011Recalls & reported problems
Cook Ireland Ltd · EchoTip Ultra, Endobronchial High Definition Ultrasound Needle · Class 2
All models of EchoTip (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00827002520110), received between Jun 8, 2017 and Jul 22, 2025 — classified as 3 injuries, 12 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Kinked5
- Retraction Problem3
- Bent2
- Improper or Incorrect Procedure or Method2
- Fracture1
- Material Deformation1
- Material Fragmentation1
Patient problems
- No Known Impact Or Consequence To Patient7
- No Consequences Or Impact To Patient5
- Foreign Body In Patient4
- No Patient Involvement1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
InjuryJul 22, 2025Break; Improper or Incorrect Procedure or Method; Break; Improper or Incorrect Procedure or Method
DURING A BRONCHOSCOPY PROCEDURE WITH NEEDLE SUCTION OF THE LIMPH NODE, THE NEEDLE BROKE, AND A FRAGMENT REMAINED IN THE AIRWAYS. IT WAS REMOVED WITH A CLIP. AS PER CUSTOMER COMPLAINT FORM: "DURING THE BRONCOSCOPY PROCEDURE WITH NEEDLE, THE NEEDLE BROKE AND REMAINED IN THE PATIENT AIRWAY, IT WAS REMOVED WITH FORCEPS."
Patient problems: Foreign Body In Patient
MAUDE report 22564149 ↗MalfunctionMar 30, 2022Material Fragmentation
PATIENT WAS HAVING AN EBUS PROCEDURE DONE IN THE OPERATING ROOM UNDER GENERAL ANESTHESIA, INTUBATED. SEVERAL PASSES HAD BEEN DONE ON THE SAME LYMPH NODE USING THE EBUS NEEDLE. WITH THE LAST PASS USING THIS NEEDLE, REMOVAL OF THE NEEDLE THROUGH THE SCOPE MET MUCH RESISTANCE. ASSISTING RT WITHDREW THE NEEDLE AND UPON EXAMINATION IT APPEARED THAT PART OF THE NEEDLE HAD BROKEN OFF. PHYSICIAN USED THE BRONCHOSCOPE TO EXAMINE ALL AREAS THAT ARE VISIBLE USING THE SCOPE AND NO NEEDLE WAS OBSERVED. PROCEDURE WAS CONTINUED USING ANOTHER NEEDLE AND OTHER INSTRUMENTS INCLUDING FLUOROSCOPY. PHYSICIAN HAD THE TECH DO A FULL FLUORO SCAN OF THE LUNGS AND NO NEEDLE WAS OBSERVED. THERE WAS NO BLEEDING UPON COMPLETION OF THE PROCEDURE. AFTER EXTUBATION AND WHILE IN RECOVERY, THE PATIENT WAS COUGHING UP BLOOD TINGED SPUTUM. PACU RN NOTED A METAL OBJECT ON HIS LIP. PULMONOLOGIST AND ANESTHESIA AWARE. REPEAT CXR COMPLETED- NO NEEDLE NOTED. FAIR LEVEL OF CONFIDENCE THAT REMAINDER OF NEEDLE WAS COUGHED OUT. FDA SAFETY REPORT ID # (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 13966173 ↗InjuryJan 6, 2020Break; Break
THE NEEDLE BROKE INSIDE THE PATIENT DURING THE SURGERY. "AS PER COMPLAINT FORM": THE DOCTOR WAS ENTERING THE STATION 10R. AT THE SECOND PUNCTURE THEY VERIFIED THAT THE NEEDLE BROKE AND WAS VISIBLE AT THE ENTRANCE OF THE MAIN BRONCHUS OF LEFT. THE PATIENT UNDERWENT AN RX AND A CT SCAN AND THE NEEDLE WAS IDENTIFIED IN THE BRONCHIAL TREE. TRANSLATION RECEIVED 12-DEC-2019: EXAM EXECUTED WITH DEDICATED INSTRUMENT FOR EBUS TBNA, ROUND SHAPE LYMPH NODES VISUALIZED WITH DIAMETER OVER 1 CM IN STATION 4R AND 10R EXECUTED TBNA IN THESE STATIONS. DURING THE PROCEDURE IN A SECOND ATTEMPT IN STATION 10R WE VERIFIED THE RUPTURE OF COOK 22 G NEEDLE. THE FRAGMENT OF NEEDLE (AROUND 2 CM) IS THEN VISUALIZED IN THE ENTRY OF THE LEFT MAIN BRONCHUS FOR AN INSTANT THEN IS NO MORE VISIBLE IN THE BRONCHIAL TREE. THE PATIENT UNDERGO AN RX EXAM AND THEN TO A TAC THORAX (CT SCAN?). THE NEEDLE PORTION INITIALLY IS DETECTED IN THE ESOPHAGUS (MAYBE EXPELLED BY THE PATIENT WITH COUGH AND THEN INGESTED) AND THEN DURING A SECOND CT SCAN, VISUALIZED IN THE STOMACH. DR (B)(6) EXECUTE A EGDS AND REMOVED THE FOREIGN BODY.
Patient problems: Foreign Body In Patient
MAUDE report 9553400 ↗MalfunctionJan 6, 2020Use of Device Problem; Use of Device Problem
THEY TRIED TO PUSH THE NEEDLE A FEW TIMES MORE. IN THIS CASE NEEDLE HAS BEEN WARPED. THAN THEY DECIDED TO OPENED NEW PRODUCT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9552469 ↗InjuryDec 24, 2019Break; Break
NEEDLE'S FRACTURING DURING THE PROCEDURE, IF THE NEEDLE NEEDS TO PASS FROM HARD CARTILAGE TO REACH THE TISSUE, NEEDLE SHAPE CAN CHANGE. ESPECIALLY, ASPIRATION MOMENT IS MOST IMPORTANT TO AVOID THE NEEDLE FRACTURING. BECAUSE NITINOL STYLET SUPPORT THE LUMEN FOR THE FIRST PUNCHING BUT ASPIRATION MOMENT, UNFORTUNATELY LUMEN STAYS EMPTY. IN THIS CASE, YOU CAN FRACTURE THE NEEDLE EASY. EVERY PROBLEM MAY BE ABOUT USING, BUT SUCH A FRACTURE PROBLEM, I CANNOT SAY ANYTHING TO SOLVE THIS PROBLEM. AS PER COMPLAINT FORM: " NEEDLE HAS BEEN PULL IT OUT FIRST. FRACTURED NEEDLE PART HAS BEEN TAKEN WITH FORCEPS."
Patient problems: Foreign Body In Patient
MAUDE report 9516356 ↗MalfunctionDec 16, 2019Break; Fracture; Improper or Incorrect Procedure or Method; Break; Fracture; Improper or Incorrect Procedure or Method
THEY INFORMED ME THAT THE DOCTOR PERFORMED THE FIRST ATTEMPT AND THE NEEDLE SEEMED GOOD BUT WHEN THEY LOOKED AT IT WITH THE ENDOSCOPE, THE IMAGE SHOWED THAT THE TIP WAS SLIGHTLY BENT. THEY TRIED TO TAKE IT OUT FROM THE PATIENT AND WHEN THEY PLACED IT ON THE STERILE TABLE, THE TIP BROKE IMMEDIATELY. THE BROKEN PART IS BEING RETURNED ALONG WITH THE NEEDLE ITSELF IN ITS ORIGINAL BOX.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9472222 ↗MalfunctionAug 21, 2019Retraction Problem; Retraction Problem
THIS REPORT IS BEING SUBMITTED AS A CANCELLATION REPORT. INITIAL REPORT DETAILS: USER CHECK THE DEVICE FUNCTION BEFORE USE WHEN FOUND OUT THE NEEDLE IS DIFFICULT TO PULL OUT. AFTER THE NEEDLE WAS PULL OUT , IT CANNOT BE RETRACTED AFTERWARD. NO USE ON PATIENT. ADDITIONAL INFORMATION PROVIDED ON 06-SEP-19 INDICATED THAT IT IS STYLET WHICH WAS DIFFICULT TO PULL OUT AND NOT THE NEEDLE. BASED ON REVIEW OF ADDITIONAL INFORMATION PROVIDED AND THE COMPLETED INVESTIGATION THIS COMPLAINT NO LONGER MEETS THE FDA MDR REPORTING CRITERIA OF A MALFUNCTION OR SERIOUS INJURY REPORT.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 8915484 ↗MalfunctionSep 4, 2018Break; Break
NEEDLE BROKE OFF DURING PROCEDURE. ACTION TAKEN: BIOPSY FORCEPS USED TO RETRIEVE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7842623 ↗MalfunctionJul 3, 2018Kinked; Retraction Problem; Material Deformation; Kinked; Retraction Problem; Material Deformation
¿WE USED AN EBUS NEEDLE THIS MORNING WHICH BENT DURING THE PROCEDURE AND WE WERE UNABLE TO PULL IT BACK THROUGH THE SCOPE SO PULLED THE WHOLE SCOPE OUT WITH NEEDLE BENT, THROUGH THE MOUTH. WHEN WE LOOKED AT THE NEEDLE IT WAS VERY WEAK AT THE POINT WHERE IT BENT (2CM FROM TIP) AND I THINK IF WE'D TRIED HARDER TO PULL IT OUT IT WOULD HAVE SNAPPED IN THE TRACHEA. WE HAVE SAVED THE NEEDLE, WOULD YOU LIKE US TO SEND IT BACK?¿ "AS PER COMPLAINT FORM": WE USED AN EBUS NEEDLE THIS MORNING WHICH BENT DURING THE PROCEDURE AND WE WERE UNABLE TO PULL IT BACK THROUGH THE SCOPE SO PULLED THE WHOLE SCOPE OUT WITH NEEDLE BENT, THROUGH THE MOUTH. WHEN WE LOOKED AT THE NEEDLE IT WAS VERY WEAK AT THE POINT WHERE IT BENT (2CM FROM TIP) AND I THINK IF WE'D TRIED HARDER TO PULL IT OUT IT WOULD HAVE SNAPPED IN THE TRACHEA. 1ST PASS SAMPLE ACQUIRED USING NEEDLE. QUITE DIFFICULT TO GET NEEDLE THROUGH TRACHEAL WALL. 2ND PASS NEEDLE DID NOT EASILY PASS THROUGH TRACHEAL WALL AND I INSTRUCTED SPR WHO WAS HOLDING THE SCOPE TO GENTLY ADVANCE THE SCOPE INTO THE PATIENT TO APPLY ADDITIONAL PRESSURE. STILL UNABLE TO GET NEEDLE INTO NODE THEREFORE ATTEMPTED TO WITHDRAW NEEDLE FROM TRACHEAL WALL, CLEAR THAT IT WAS BE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7657612 ↗MalfunctionMay 18, 2018Retraction Problem; Retraction Problem
ADDITIONAL INFORMATION RECEIVED CONFIRMATION THAT THE NEEDLE WAS NOT FULLY RETRACTED PRIOR TO PATIENT REMOVAL. DURING EBUS PROCEDURE, TUMOR WAS LOCATED BY ULTRASONIC LOCALIZATION. USER TRY TO DO BIOPSY WITH ECHO-HD-22-EBUS-O BUT FOUND OUT THE CORE NEEDLE WAS VERY DIFFICULT TO PULL OUT AND RETRIEVE BACK TO ULTRASOUND NEEDLE. USER WITHDRAW THE DEVICE FROM EBUS AND DETECTED A BENDING 10CM FROM OUTER SHEATH WHICH CAUSE THE DIFFICULTY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7524987 ↗MalfunctionApr 30, 2018Kinked; Material Perforation; Kinked; Material Perforation
THE NEEDLE AND SHEATH "SEPARATED" AFTER FIRST "ATTEMPT" WHICH AFFECTED SECOND BIOPSY. OPERATOR CHANGED ANOTHER SAME PRODUCT TO FINISH THE BIOPSY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7472137 ↗MalfunctionMar 22, 2018Kinked; Kinked
REP ADVISED THAT THE STILLED WOULD RUN THROUGH THE DEVICE AFTER THEIR FIRST PASS. "AS PER COMPLAINT FORM": FIRST PASS SUCCESSFUL HOWEVER STYLET WOULD NOT ENTER BACK INTO THE BORE OF THE NEEDLE AGAIN.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7360876 ↗MalfunctionFeb 14, 2018Bent; Kinked; Bent; Kinked
REPORTED AS - DIFFICULTY TO GET OUT THE SHEATH DESPITE MANY ATTEMPTS AND THE PROCEDURE WAS ABORTED. LAB EVALUATION CONDUCTED ON THE 19 JAN 2018 DISCOVERED THE NEEDLE WAS KINKED PROXIMALLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7270098 ↗MalfunctionDec 27, 2017Bent; Kinked; Bent; Kinked
PRODUCT DID NOT WORK AS EXPECTED. REP DOESN'T HAVE FURTHER INFO. PRODUCT AVAILABLE IN PURCHASING DEPT. "AS PER COMPLAINT FORM": CUSTOMER OPENED THE PACKAGE. REMOVED THE SYSTEM (EBUS NEEDLE) OUT OF THE PACKAGE. HE TRIED TO MOVE THE SLIDING SHEATH ADJUSTER IN-FRONT OF THE ENDOSCOPE. BUT IT'S NOT WORKING, THE SHEATH ADJUSTER STUCK AT THE SYSTEM. AFTER THIS HE TRIED TO REMOVE THE STYLET. BUT IT WASN'T POSSIBLE.HE PULLED HARDER, WITHOUT SUCCESS. AFTER THE THIRD TRY THE STYLET CAME OUT OF THE SYSTEM AND WAS COMPLETELY DESTROYED. THE CUSTOMER USED A NEW COOK SET WITH THE SAME LOT NUMBER AND THE PRODUCT WORKS WELL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7147532 ↗MalfunctionJun 8, 2017Break; Break
AS REPORTED TO CUSTOMER RELATIONS: " ON THE THIRD PASS OF THE NEEDLES THEY BENT. THE PHYSICIAN HAD A QUANTITY OF 3 NEEDLES OF THE SAME LOT NUMBER BEND. A FOURTH NEEDLE WAS USED AND WORKED AS INTENDED. THERE WAS NO HARM TO THE PATIENT." UPON EVALUATION TWO FAILURE MODES WERE DISCOVERED: NEEDLE KINK/BENT AND NEEDLE BROKE [3001845648-2017-00210] (B)(4). PER THE DISTRICT MANAGER: "THE CUSTOMER REPLIED THAT THEY WEREN¿T SURE WHEN THE NEEDLE BROKE, BUT ASSUMED IT WAS DURING BIOPSY. THEY STATED THAT THE NEEDLE TIP CAME OUT OF THE SCOPE DURING CLEANING. THE SCOPE DID NOT PASS THE LEAK TEST AND HAD TO BE SENT OUT FOR REPAIR. THEY ALSO STATED THERE WAS NO PATIENT HARM OR PIECES THAT NEEDED TO BE RETRIEVED WHILE INSIDE THE PATIENT."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6621254 ↗
FDA clearance and classification
The FDA regulates product code FCG (biopsy needle) under 21 CFR 876.1075 as a Class II device. 21 CFR 876.1075 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the EchoTip G52011
Has the EchoTip G52011 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002520110), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the EchoTip G52011?
The FDA's MAUDE database lists 15 adverse event reports for this device, received between Jun 8, 2017 and Jul 22, 2025. In those reports, the most common specific device problems are “Break” (6), “Kinked” (5) and “Retraction Problem” (3); the most common patient problem is “Foreign Body In Patient” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EchoTip G52011?
None of the reports are classified as deaths, 3 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the EchoTip G52011, and is it still sold?
The FDA's device database lists Cook Ireland Ltd as the company responsible for it (brand EchoTip, model G52011). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Biopsy needle devices
- MAGNUM MN1410Bard Peripheral Vascular, Inc. · Biopsy needle1 recall119 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4021Olympus Medical Systems Corp. · Biopsy needle1 recall94 reports
- EchoTip G31520Cook Ireland Ltd · Biopsy needle1 recall84 reports
- Single Use Aspiration Needle NA-201SX-4021Olympus Medical Systems Corp. · Biopsy needle1 death126 reports
- EchoTip ProCore G34785Cook Ireland Ltd · Biopsy needle97 reports
- MAGNUM MN1820Bard Peripheral Vascular, Inc. · Biopsy needle1 recall37 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4022Olympus Medical Systems Corp. · Biopsy needle1 recall28 reports
- Interject M00518250Boston Scientific Corporation · Biopsy needle70 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page