Fusion G55815Recalls & reported problems
Cook Incorporated · Fusion Omni-tome Preloaded Sphincterotome · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002558151), received between Apr 26, 2016 and Apr 5, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Separation1
- Sharp Edges1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2022Sharp Edges; Sharp Edges
IN PREPARATION FOR A PROCEDURE, THE USER SELECTED A COOK FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME. IT WAS INITIALLY REPORTED THAT UPON INSPECTION, THE TIP OF THE SPHINCTEROTOME WAS CRACKED. THEY SIMPLY OPENED ANOTHER DEVICE AND CONTINUED WITH THE PROCEDURE. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. OUR EVALUATION OF THE RETURNED DEVICE ON 3/8/2022 DETERMINED THAT THE DISTAL TIP OF THE CATHETER IS CRACKED, IT IS ALSO JAGGED AND A PIECE OF THE CATHETER IS PROTRUDING OUT [SUBJECT OF REPORT]. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14014224 ↗MalfunctionMar 3, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME. IT WAS INITIALLY REPORTED THAT THIS TOME LEFT A FOREIGN BODY ON THE WIRE [GUIDE] AFTER IT WAS EXCHANGED OUT OF THE SYSTEM. A PIECE OF THE DEVICE WAS LEFT ON THE WIRE [GUIDE] AFTER IT WAS EXCHANGED OUT. THIS CAUSED US TO HAVE TO REMOVE THE WIRE [GUIDE] TO REMOVE THE FOREIGN BODY. PER ADDITIONAL INFORMATION RECEIVED ON 2/4/2022: THERE WAS A RING THAT WAS LEFT ON THE WIRE [METAL BAND DETACHES]. [SUBJECT OF REPORT]: THE METAL BAND DETACHED ON THE WIRE [GUIDE] AFTER IT WAS EXCHANGED OUT AND CAUSED THE PHYSICIAN TO HAVE TO REMOVE THE WIRE [GUIDE] TO REMOVE THE FOREIGN BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 13661287 ↗MalfunctionApr 26, 2016Material Separation; Material Separation
THIS REPORT SUMMARIZES ONE (1) MALFUNCTION EVENT. A REVIEW OF THE EVENT INDICATED THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME. DURING THE PROCEDURE, THE CUTTING WIRE SECURING COMPONENT (ANCHOR) SEPARATED FROM CATHETER. IT WAS REPORTED THAT THERE WAS NO SERIOUS COMPLICATION WITH THIS PROBLEM. THIS EVENT INVOLVED A PATIENT WITH NO PATIENT CONSEQUENCES. THIS PATIENT HAD PRE-EXISTING OBSTRUCTIVE JAUNDICE WITH COMMON BILE DUCT STONE. THIS PATIENT WAS REPORTED TO BE A (B)(6).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5609714 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page