Skip to content

Echotip ClearCore G59742Recalls & reported problems

Cook Ireland Ltd · Echotip ClearCore EUS Biopsy Needle · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002597426), received between Aug 11, 2025 and Nov 17, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Failure to Obtain Sample” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Obtain Sample1
  • Material Twisted/Bent1
  • Off-Label Use1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 17, 2025Failure to Obtain Sample; Material Twisted/Bent; Failure to Obtain Sample; Material Twisted/Bent

    AN INITIAL REPORT IS BEING SUBMITTED DUE TO COMPLETION OF THE DEVICE EVALUATION ON 10-OCT-2025. PER REP - (B)(6) 2025 - INITIALLY, THEY BIOPSIED THE LEFT LIVER LOBE AND WERE SUCCESSFUL IN OBTAINING A SAMPLE VIA A TRANS-GASTRIC APPROACH. THE NEEDLE WAS THEN REPOSITIONED TO SAMPLE RIGHT LIVER LOBE VIA A TRANS-DUODENAL APPROACH. HOWEVER, RESISTANCE WAS FELT WHEN EXTENDING THE NEEDLE OUT OF THE SCOPE. WHEN TRYING TO PUNCTURE TO OBTAIN THE SAMPLE, THE NEEDLE FAILED TO PUNCTURE AND PUSHED THE SAMPLE AWAY. TWO ATTEMPTS WERE TRIED AND UPON THE SECOND FAILURE, THE NEEDLE WAS REMOVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A COMPETITIVE NEEDLE. 1. WAS GAINING ACCESS TO THE TARGET SITE DIFFICULT? N/A, YES, NO. NO. 2. WAS PUNCTURE OF THE TARGETED SITE DIFFICULT? YES. 3. WHAT IS THE SCOPE MANUFACTURER AND MODEL NUMBER THAT WAS USED? FUJI. 4. WAS THE SCOPE RECENTLY SERVICE/REPAIRED? UNKNOWN. 5. WAS THE DEVICE USED IN A TORTUOUS POSITION? SOMEWHAT. TRANSDUODENAL PUNCTURE ATTEMPTED. 6. WAS FORCE REQUIRED TO REMOVE THE DEVICE? N/A, YES, NO NO. 7. WERE ANY OTHER DEFECTS (OTHER THAN THE COMPLAINT ISSUE) OBSERVED ON THE DELIVERY SYSTEM (E.G., KINKS) PRIOR TORETURN? (IF YES, PLEASE SPECIFY WHAT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23567952 ↗
  • MalfunctionAug 11, 2025Off-Label Use; Off-Label Use

    A SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON 25-SEP-2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22769685 ↗

Other Biopsy needle devices

All Biopsy needle devices →

Sources and updates