SureTome SW G61350Recalls & reported problems
Cook Incorporated · SureTome SW Sphincterotome with DomeTip · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002613508), received between Mar 4, 2026 and Sep 17, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Accessory Incompatible” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Accessory Incompatible1
- Material Separation1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Perforation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Material Separation
During an endoscopic retrograde cholangiopancreatography (ERCP) procedure, the physician used five (5) Cook SureTome SW Sphincterotomes. It was reported that during a very difficult 5-hour ERCP with challenging patient anatomy and significant scope retroflexion, five SureTome sphincterotomes were used. One of which, was passed down the scope and when exiting tip of scope in patient the doctor noticed the tip looked very malformed. When the device was tested by bowing it, the cutting wire broke. This instance looks like it was a very tough procedure in which multiple sphincterotomes had to be used and one was broken because of scope position.A section of the device did not remain inside the patient¿s body. The patient did not require any additional procedures due to this occurrence. According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50804263 ↗InjuryMar 4, 2026Accessory Incompatible; Physical Resistance/Sticking
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK SURETOME SW SPHINCTEROTOME. IT WAS REPORTED THAT THERE WAS SIGNIFICANT RESISTANCE WHEN ADVANCING THE GUIDE WIRE THROUGH THE SPHINCTEROTOME. THIS REDUCED ABILITY TO CONTROL THE WIRE MAY HAVE CONTRIBUTED TO A SMALL PERFORATION OF AN INTRAHEPATIC DUCT. AS A RESULT, A PLASTIC STENT WAS PLACED TO FACILITATE DRAINAGE AND PROMOTE HEALING. FURTHER INFORMATION WAS RECEIVED STATING THAT THE PHYSICIAN WAS USING A SURETOME WITH AN ACROBAT 2 ANGLED 035 WIRE GUIDE. THEY FLUSHED THE RACETRACK OF THE WIRE GUIDE AND THE SURETOME LUMEN AND LOADED THE WIRE, AND SOME RESISTANCE WITH NOTED. THEY TIPPED THE WIRE GUIDE TIP IN WATER TO ASSIST. THEY WERE ABLE TO GET THE WIRE GUIDE LOADED. THEY ACCESSED THE DUCT. THE USER NOTED THE WIRE GUIDE FELT "SLIPPERY" BUT THEY WERE ABLE TO MAINTAIN ACCESS. THEY STATED THAT WHEN THEY ADVANCED THE TOME INTO THE DUCT THE WIRE TRAVELED WITH IT. THEY INJECTED SOME CONTRAST AND THERE WAS A SLIGHT BLUSHING INDICATIVE OF A SMALL INTRA HEPATIC DUCT PERFORATION. THEY THEN SWITCHED WIRE GUIDES TO AN ANGLED 035 TERUMO WIRE GUIDE. THE PHYSICIAN NOTED RESISTANCE AFTER THE WIRE SWIT
Patient problems: Perforation
MAUDE report 24510162 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page