Skip to content

Prodisc C SK PDSXL6Recalls & reported problems

Centinel Spine, LLC · Prodisc C SK, Extra Large, 6mm · Class 3 · Implant

All models of Prodisc C SK (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843193113986), received between Mar 6, 2026 and Mar 6, 2026 — classified as 1 injury.

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (P070001); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 20, 2026Class IIOpen, Classified

    Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.

    Z-2051-2026 · Root cause: Labeling mix-ups · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Inadequate Pain Relief1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 6, 2026Adverse Event Without Identified Device or Use Problem

    A PATIENT (55 YEARS, FEMALE) WAS IMPLANTED WITH TWO PDC SK DEVICES AT C5-6 AND C6-7 ON (B)(6) 2024. PATIENT HAD LINGERING PAIN AND THE SURGEON WAS CONCERNED ABOUT A POSSIBLE INFECTION. AN EXPLORATORY REVISION SURGERY WAS SCHEDULED. THERE WERE NO VISIBLE SIGNS OF INFECTION BUT THE PATIENT HAD POOR BONE QUALITY. C6-7 WAS REMOVED AND THE AREA SWABBED FOR CULTURES, DURING THE REMOVAL SURGERY THE SURGEON DECIDED TO REMOVE THE C5-6 LEVEL AS WELL. THE SK DEVICES WERE REPLACED WITH CAGES AND THEN THE PATIENT WAS CLOSED AND FLIPPED AND POSTERIOR FIXATION WAS ADDED. PATIENT WAS GIVEN ORAL ANTIBIOTICS TO START WHILE WAITING FOR CULTURES TO BE RETURNED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24532155 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P070001. One submission often covers many models, so check the product and lot details in each record.

  • Z-2050-2026 · Mar 20, 2026 · Open, Classified
    Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.
  • Z-1426-2026 · Dec 31, 2025 · Open, Classified
    Brand Name: Prodisc Product Name: Prodisc C SK Model/Catalog Number: PDSL6 Software Version: N/A Product Description: Total cervical disc replacement Component: No
  • Z-1406-2021 · Mar 5, 2021 · Terminated
    Prodisc C - prosthesis, intervertebral disc - Product Usage: indicated in skeletally mature patients for reconstruction of the disc from C3-C7 following single-level discectomy for

Other Prosthesis, Intervertebral Disc devices

All Prosthesis, Intervertebral Disc devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page