Prodisc C SK recallZ-2051-2026 · Centinel Spine, Inc. · initiated Mar 20, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Centinel Spine, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.
FDA root cause category: Labeling mix-ups
What the company is doing
Centinel Spine notified distributors on about 03/20/2026 via telephone calls and emailed letter. Surgeons were sent letters via FedEx on 03/24/2026 and by email on 03/25/2026 & 03/26/2026. Distributors were instructed to identify any affected units on hand, provide photographs documenting such, and arrange for the return of all affected units. Additionally, distributors have been instructed to diligently check prodisc C SK inventory prior to any scheduled surgery to ensure that the recalled lot numbers are not available for implantation. This includes both consignment and loaner inventory. Acknowledgment of the letter has been requested with return envelopes provided. For surgeons, they were instructed to monitor any patient(s) that have received the recalled device for any unexpected device-related symptoms throughout the regularly scheduled check-in procedures. Centinel Spine intends to remove from the market all remaining implants within the lots effected by this recall and issue a replacement product with corrected labeling for any remaining products that are subject to this recall.
Recall details
- Recall number
- Z-2051-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 20, 2026
- Posted by the FDA
- May 1, 2026
- Recalling firm
- Centinel Spine, Inc.
- Quantity
- 30 units
- Product code
- MJO
- FDA submissions
- P070001, P070001S019, P070001S025
- Distribution
- US Nationwide distribution in the states of AZ, CA, GA, LA, MO, NY, TN, TX.
Product description
Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
Code information (models, lots, serial numbers)
Model Number: PDSXL6; UDI-DI: 00843193113986; Lot Number: 2026-0027
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Prodisc C SK PDSXL6Centinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall1 report
Product lines
Other devices under the same FDA submission
45 devices cleared under P070001, P070001S019, P070001S025 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Prodisc C SK PDSM5Centinel Spine, LLC · Prosthesis, Intervertebral Disc2 recalls4 reports
- Prodisc C 09.820.055SCentinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall9 reports
- Prodisc C SK PDSXL5Centinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall3 reports
- Prodisc C SK PDSL6Centinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall1 report
- Prodisc C Vivo PDVM5Centinel Spine, LLC · Prosthesis, Intervertebral Disc21 reports
- Prodisc C Vivo PDVL5Centinel Spine, LLC · Prosthesis, Intervertebral Disc20 reports
- Prodisc C Vivo PDVXL5Centinel Spine, LLC · Prosthesis, Intervertebral Disc7 reports
- Prodisc C Vivo PDVLD5Centinel Spine, LLC · Prosthesis, Intervertebral Disc6 reports
- Prodisc C 09.820.026SCentinel Spine, LLC · Prosthesis, Intervertebral Disc5 reports
- Prodisc C 09.820.036SCentinel Spine, LLC · Prosthesis, Intervertebral Disc5 reports
- Prodisc C 09.820.045SCentinel Spine, LLC · Prosthesis, Intervertebral Disc5 reports
- Prodisc C Vivo PDVMD5Centinel Spine, LLC · Prosthesis, Intervertebral Disc5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98721).
- Z-2050-2026 · Open, ClassifiedProdisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page