Prodisc L PDL-M-IP08SRecalls & reported problems

Centinel Spine, LLC · INFERIOR ENDPLATE MEDIUM 8°-STERILE · Class 3 · Implant

All models of Prodisc L (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00843193114976), received between Oct 2, 2024 and Mar 25, 2026 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Malposition of Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Malposition of Device3
  • Adverse Event Without Identified Device or Use Problem1
  • Collapse1
  • Expulsion1
  • Insufficient Device Problem Information1

Patient problems

  • Insufficient Information2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Limb Pain1
  • Movement Disorder1
  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Nervous System Problem1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 25, 2026Malposition of Device

    A PATIENT WAS IMPLANTED WITH A PRODISC L DEVICE AT L5-S1 ON (B)(6) 2026. THE PATIENT FELT TIPPED FORWARD AND IT WAS FOUND THAT THE DEVICE WAS KYPHOTIC. A REVISION WAS COMPLETED ON (B)(6) 2026 AND THE SUPERIOR ENDPLATE WAS REPOSITIONED A LITTLE DEEPER WHICH SEEMED TO CORRECT THE KYPHOSIS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24690326 ↗
  • InjuryJan 12, 2026Malposition of Device

    A PATIENT (FEMALE, 42 YEARS) WAS IMPLANTED WITH A PRODISC L DEVICE AT L5-S1 ON (B)(6) 2025. AFTER IMPLANTATION SURGERY THE PATIENT WAS EXPERIENCING NEUROLOGICAL ISSUES AND IT WAS FOUND THAT THE DEVICE WAS OFF OF MIDLINE BY 1.5 MM. THE IMPLANT WAS REMOVED ON (B)(6) 2025 AND THE PATIENT WAS FUSED WITH A CAGE AND PLATE.

    Patient problems: Unspecified Nervous System Problem

    MAUDE report 24036141 ↗
  • InjurySep 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WAS IMPLANTED WITH TWO PRODISC L DEVICES AT L4-5 AND L5-S1 ON (B)(6) 2025. THE SURGERY WENT WELL AND WITHOUT CONCERN, THE PATIENT WAS FULLY DECOMPRESSED AT BOTH LEVELS AND IMPLANTS WERE WELL PLACED. THE PRE-OP PATIENT COMPLAINTS WERE RESOLVED BUT THE PATIENT PRESENTED 24 HOURS AFTER SURGERY WITH A NEW LEFT FOOT DROP. A POST-OP MRI WAS COMPLETED TO ASSESS THE CANAL AND THE ASSUMPTION THAT THE FOOT DROP IS A STRETCH INJURY OR A RETRACTOR INJURY. THE SURGEON TREATED THE PATIENT WITH STEROIDS AND THE PATIENT IS SLOWLY REGAINING SOME FUNCTION. SURGEON REQUESTED A PEER-TO-PEER DISCUSSION TO ENSURE THAT HE IS TREATING PROPERLY. DISCUSSION OUT OF THE PEER-TO-PEER MEETING RESULTED IN THE FOLLOWING CONCLUSIONS. THERE IS NO CONCERN WITH THE IMPLANT OR THE IMPLANT PLACEMENT. THIS IS AN UNCOMMON BUT POTENTIAL COMPLICATION FOR WHICH THERE ARE NO RELIABLE PRE-OPERATIVE TESTS OR INDICATORS. PATIENT WILL HAVE A CT MYELOGRAM DONE TO RULE OUT ANY COMPRESSION UNABLE TO BE IDENTIFIED WITH THE EXISTING POST OP MRI. PATIENT WILL CONTINUE MEDICATION AND THERAPY IN HOPES OF REGAINING FUNCTION IN THE FOOT.

    Patient problems: Movement Disorder

    MAUDE report 23008501 ↗
  • InjuryJun 12, 2025Malposition of Device; Malposition of Device

    A PATIENT WAS IMPLANT WITH A PDL DEVICE AT L5-S1 ON (B)(6) 2025. FOLLOWING SURGERY, THE PATIENT HAD DEVELOPED LEG PAIN THAT WAS NOT PRESENT PRIOR TO SURGERY. IMAGING SHOWED THAT THE IMPLANT WAS TOO FAR FROM POSTERIOR. THE SURGEON REVISED THE IMPLANT ON (B)(6) 2025 BY REMOVING THE INLAY, REPOSITIONING THE ENDPLATES AND INSERTING A NEW INLAY.

    Patient problems: Limb Pain

    MAUDE report 22193244 ↗
  • InjuryMar 3, 2025Expulsion; Expulsion

    PATIENT WAS IMPLANTED WITH A PDL IMPLANT AT L5-S1 ON (B)(6)2025. IMPLANT WAS REMOVED ON (B)(6)2025 AND THE PATIENT WAS FUSED. IMPLANT WAS REMOVED DUE TO IMPLANT EXPULSION, NO PATIENT SYMPTOMS WERE PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 21510216 ↗
  • InjuryDec 10, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A PATIENT WAS IMPLANTED WITH A PRODISC L DEVICE AT L5/S1 ON (B)(6) 2024. THE DEVICE WAS REMOVED ON (B)(6) 2024. NO REASON FOR REMOVAL WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 20893992 ↗
  • MalfunctionOct 2, 2024Collapse; Collapse

    A PATIENT WAS IMPLANTED WITH A PRODISC L DEVICE AT L5-S1 ON (B)(6) 2024. AT A FOLLOW UP APPOINTMENT IT WAS SUSPECTED THAT THE IMPLANT WAS SUBSIDED. ADDITIONAL IMAGING WAS REQUESTED TO BETTER VISUALIZE THE DEVICE. A CT SCAN WAS COMPLETED IT WAS CONFIRMED THAT THERE WAS A FRACTURED L5 ENDPLATE AND SOME SUBSIDENCE OF THE IMPLANT THERE. THE PATIENT IS CURRENTLY ASYMPTOMATIC AND HAPPY WITH THE IMPLANT. THE SURGEON WILL FOLLOW THE PATIENT WITH ADDITIONAL RADIOGRAPHS EVERY 3-6 MONTHS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20350780 ↗

Other Prosthesis, Intervertebral Disc devices

All Prosthesis, Intervertebral Disc devices →