OsteoMed 450-0005-00Recalls & reported problems

Osteomed LLC · Power Control Console 2i with Irrigation · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845694012119), received between Aug 11, 2021 and Aug 11, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unintended Electrical Shock” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Unintended Electrical Shock1

Patient problems

  • Electric Shock1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2021Unintended Electrical Shock; Unintended Electrical Shock

    ON (B)(6) 2021 OSTEOMED WAS NOTIFIED THAT, UNIT 420-0021-00 AND SN (B)(4) WAS SHOCKING THE PATIENTS. PER THE CUSTOMER:"WE HAVE THIS UNIT THAT IS SHOCKING THE PATIENTS. WE SEEM TO THINK IT IS THE CORD BECAUSE IT HAS A KINK IN IT. " ON 6/8/2021 OSTEOMED REACHED OUT TO THE REPORTING CUSTOMER AND COLLECTED: I SPOKE WITH THE OFFICE MANAGER THIS MORNING, WHO ALSO REPORTED THE COMPLAINT. SHE CONFIRMED THAT THERE WAS NO PATIENT HARM DUE TO THE "SHOCKING SENSATION" THAT WAS REPORTED. (HARM= BURNS, CUTS, RED MARKS OR ANY INJURY THAT WOULD REQUIRE TREATMENT). THE SURGEON REPORTED THAT EVERY TIME THE DRILL TOUCHED THE PATIENT THERE WAS A SENSATION THAT WAS NOT REPORTED AS PAINFUL. TO RESOLVE, THE SURGEON TURNED OFF THE MACHINE AND USED ANOTHER ONE. THERE WERE NO DELAYS IN SURGERY TO REPORT. (B)(6) COULD ALSO NOT THINK OF ANY DAMAGE TO THE POWER CONSOLE TO REPORT. MDR DETERMINATION FORM COMPLETED. BASED ON INFORMATION RECEIVED, THERE IS NO INFORMATION THAT REPORTS ANY HARM TO A PATIENT OR CONFIRMED MALFUNCTION OF THE PRODUCT. COMPLAINT DETERMINED NOT TO BE MDR REPORTABLE. ON 6/14/2021 AS A FOLLOW UP: THE CUSTOMER CONTACTED THE SALES REPRESENTATIVE WHO REPORTED THAT THE CUSTOMER HAD EXPERIENCED

    Patient problems: Electric Shock

    MAUDE report 12310900 ↗

Other HANDPIECE, ROTARY BONE CUTTING devices