Skip to content

CPR-D PADZ 8900-0800-01Recalls & reported problems

Zoll Medical Corporation · CPR-D*PADZ ONE PIECE ELECTRODE PAD WITH REAL CPR HELP · Class 3

All models of CPR-D PADZ (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll CPR-D PADZ one-piece electrode pad is used with automated external defibrillators. Most of the 43 reports describe the defibrillator failing to recognize the pads or failing its self-test during routine checks, plus pads that would not stick to a patient's skin. Reports include 2 deaths and 1 injury. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
2
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00847946000189), received between Aug 24, 2015 and Aug 28, 2026 — classified as 2 deaths, 1 injury, 40 malfunctions.

The most frequently reported specific device problem is “Unable to Obtain Readings” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unable to Obtain Readings18
  • Failure of Device to Self-Test8
  • Patient Device Interaction Problem7
  • Accessory Incompatible3
  • Component or Accessory Incompatibility1
  • Defective Component1
  • Device Difficult to Setup or Prepare1
  • Device Displays Incorrect Message1

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • No Patient Involvement9
  • No Information4
  • Burn(s)1
  • Cardiac Arrest1
  • Death1
  • Insufficient Information1
  • Necrosis1

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEVICE FAILED SELF TEST USING THESE ELECTRODE PADS.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50497964 ↗
  • MalfunctionOct 28, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A MALE PATIENT (AGE UNKNOWN), THE ELECTRODE PADS WOULD NOT ADHERE TO THE PATIENT'S SKIN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER SET OF PADS TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED.

    Patient problems: Cardiac Arrest

    MAUDE report 23403034 ↗
  • MalfunctionOct 15, 2025Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR PROMPTED A "UNIT FAILED" MESSAGE USING THESE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23285158 ↗
  • MalfunctionJan 31, 2025Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEVICE PROMPTED A "UNIT FAILED" MESSAGE USING THESE CPR-D PADZ. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21284005 ↗
  • MalfunctionDec 12, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR DOES NOT RECOGNIZED THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20915749 ↗
  • MalfunctionDec 10, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEVICE INAPPROPRIATELY DISPLAYED A "PLUG IN CABLE" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20892321 ↗
  • MalfunctionSep 19, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THESE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20262395 ↗
  • MalfunctionJun 18, 2024Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE ASSOCIATED DEFIBRILLATOR FAILED SELF TEST USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19557080 ↗
  • MalfunctionMay 20, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR DISPLAYED A "PLUG IN CABLE" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19351115 ↗
  • MalfunctionApr 3, 2024Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A MALE PATIENT (AGE UNKNOWN), THE ELECTRODE PADS WERE DAMAGED DURING REMOVAL FROM THE PACKAGING. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19031952 ↗
  • MalfunctionMar 21, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A 50-60-YEAR-OLD FEMALE PATIENT, THE ELECTRODES WOULD NOT ADHERE TO THE PATIENT'S SKIN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18949295 ↗
  • MalfunctionAug 17, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17568287 ↗
  • MalfunctionAug 15, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING INCOMING TESTING, THE ASSOCIATED DEFIBRILLATOR PROMPTED A "PLUG IN CABLE MESSAGE -- INAPPROPRIATE" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17548178 ↗
  • MalfunctionMay 23, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THESE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16983841 ↗
  • MalfunctionMar 21, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A MALE PATIENT (AGE UNKNOWN), THE ELECTRODES WOULD NOT ADHERE TO THE PATIENT'S SKIN. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16585190 ↗
  • MalfunctionMar 15, 2023Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEVICE FAILED SELF TEST USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16547538 ↗
  • MalfunctionMar 2, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THESE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16473335 ↗
  • MalfunctionDec 13, 2022Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEVICE PROMPTED A "UNIT FAILED" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15972695 ↗
  • MalfunctionOct 5, 2022Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEVICE FAILED SELF TEST USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15545545 ↗
  • MalfunctionOct 5, 2022Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEVICE FAILED SELF TEST USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15544966 ↗
  • MalfunctionSep 15, 2022Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THESE ATTACHED ELECTRODES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15432990 ↗
  • InjuryJul 8, 2022Patient Device Interaction Problem; Patient Device Interaction Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), AFTER REMOVING THE ELECTRODE PADS, AN ABRASION WAS FOUND ON THE PATIENT'S SKIN. COMPLAINANT INDICATED THAT THE PATIENT SUSTAINED BURN MARKS. NO INFORMATION AVAILABLE TO WHAT DEGREE THE BURNS ARE.

    Patient problems: Burn(s)

    MAUDE report 14959729 ↗
  • MalfunctionJun 30, 2022Patient Device Interaction Problem; Patient Device Interaction Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), SEVERAL SETS OF ELECTRODE PADS NEEDED TO BE REPLACED DURING THE RESUSITATION DUE TO NOT BEING ABLE TO ADHERE TO THE PATIENT. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 14873670 ↗
  • MalfunctionJun 13, 2022Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14678235 ↗
  • MalfunctionOct 29, 2021Accessory Incompatible; Accessory Incompatible

    COMPLAINANT ALLEGED THAT DURING INCOMING TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO CONNECT TO THE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12723076 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →