VOCSN V+Pro, English (REF PRT-01185-002)Recalls & reported problems

Ventec Life Systems, Inc. · V+Pro, English: The VOCSN Unified Respiratory System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It may be used in invasive and non- invasive applications. The VOCSN is intended for pediatric through adult patients weighing at least 5 kg. It is intended for use in home, hospital, institutional and transport settings, including portable applications. www.VentecLife.com · Class 2

All models of VOCSN (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Ventec VOCSN V+Pro ventilator provides breathing support for patients. Of its 570 reports, all describe malfunctions, most often delivering the wrong pressure or breath volume, with unexpected shutdowns also mentioned. No injuries or deaths are reported. One open recall, started in February 2025, covers units serviced with incorrect parts that can cause unexpected shutdowns or inaccurate breath monitoring.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
570
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 570 adverse event reports that list this exact device (DI 00850018761154), received between Mar 4, 2022 and Jun 5, 2026 — classified as 570 malfunctions.

The most frequently reported specific device problem is “Output Problem” (519 reports).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K162877); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 11 reportsDec 2023: 5 reportsJan 2024: 20 reportsFeb 2024: 17 reportsMar 2024: 23 reportsApr 2024: 15 reportsMay 2024: 14 reportsJun 2024: 13 reportsJul 2024: 8 reportsAug 2024: 17 reportsSep 2024: 17 reportsOct 2024: 52 reportsNov 2024: 32 reportsDec 2024: 6 reportsJan 2025: 17 reportsFeb 2025: 0 reportsMar 2025: 36 reportsApr 2025: 10 reportsMay 2025: 12 reportsJun 2025: 33 reportsJul 2025: 33 reportsAug 2025: 4 reportsSep 2025: 10 reportsOct 2025: 4 reportsNov 2025: 76 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 6 reportsMar 2026: 24 reportsApr 2026: 9 reportsMay 2026: 15 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202676
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 3, 2025Class IOpen, Classified

    Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.

    Z-1196-2025 · Root cause: Component change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Output Problem519
  • Unexpected Shutdown32
  • Connection Problem31
  • Inaccurate Delivery9
  • Failure to Power Up5
  • Device Alarm System4
  • Display or Visual Feedback Problem4
  • Defective Alarm1

Patient problems

  • No Clinical Signs, Symptoms or Conditions570

Recent adverse event reports

The 25 most recent of 570 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT IT WAS UNABLE TO MAINTAIN PEEP (POSITIVE END EXPIRATORY PRESSURE) AND THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25417574 ↗
  • MalfunctionMay 31, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. AS PART OF THE DEVICE INVESTIGATION UPON RECEIVING THE DEVICE, VENTEC REVIEWED A COPY OF THE AUTHORIZED SERVICE PROVIDER'S (ASP'S) SERVICE REPORT WHICH STATED THAT THE DEVICE HAD MEASURED HIGHER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET AND THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25344212 ↗
  • MalfunctionMay 31, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT MEASURED LOWER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25344203 ↗
  • MalfunctionMay 28, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS DELIVERING LOWER THAN SET PEEP (POSITIVE END EXPIRATORY PRESSURE) AND LOW VTE (EXHALED TIDAL VOLUME). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326262 ↗
  • MalfunctionMay 28, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE'S EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. ADDITIONALLY, IT WAS REPORTED THAT THE DEVICE'S BREATH RATE WAS OUT OF SPECIFICATION. HOWEVER, THE ASP COULD NOT ADVISE WHETHER IT WAS DELIVERING BELOW OR ABOVE SPEC, SO IT IS BEING REPORTED OUT OF AN ABUNDANCE OF CAUTION. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25325981 ↗
  • MalfunctionMay 28, 2026Output Problem

    AN AUTHORIZED THIRD PARTY SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE VENTILATOR PROVIDED AN ALARM INDICATING HIGHER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET AND THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25325949 ↗
  • MalfunctionMay 27, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS DISPLAYING THE PATIENT CIRCUIT DISCONNECT ALARM. THERE WAS NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25310767 ↗
  • MalfunctionMay 18, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS LCD TOUCHSCREEN WAS UNRESPONSIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25211459 ↗
  • MalfunctionMay 15, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE'S EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25194679 ↗
  • MalfunctionMay 15, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE MEASURED LOWER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25194664 ↗
  • MalfunctionMay 15, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT IT WAS UNABLE TO MAINTAIN PEEP (POSITIVE END EXPIRATORY PRESSURE), THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW, AND THAT ITS BREATH RATE WAS NOT AS EXPECTED (LOW). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25194569 ↗
  • MalfunctionMay 15, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS DISPLAYING THE PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25194503 ↗
  • MalfunctionMay 14, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE DISPLAYED AN ALARM INDICATING LOW INSPIRATORY PRESSURE. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25179087 ↗
  • MalfunctionMay 11, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS LCD TOUCHSCREEN WAS UNRESPONSIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25144113 ↗
  • MalfunctionMay 8, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS UNABLE TO MAINTAIN SET PEEP (POSITIVE END EXPIRATORY PRESSURE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25129270 ↗
  • MalfunctionMay 5, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT MEASURED LOWER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET AND THAT IT WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25088982 ↗
  • MalfunctionApr 29, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS DELIVERING LOWER THAN SET PEEP (POSITIVE-END EXPIRATORY PRESSURE) AND LOW VTE (EXHALED TIDAL VOLUME). THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25038470 ↗
  • MalfunctionApr 29, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS DELIVERING LOWER THAN SET PEEP (POSITIVE END EXPIRATORY PRESSURE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25038254 ↗
  • MalfunctionApr 28, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023887 ↗
  • MalfunctionApr 28, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023883 ↗
  • MalfunctionApr 28, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE PROVIDED AN ALARM INDICATING LOW INSPIRATORY PRESSURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023849 ↗
  • MalfunctionApr 28, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023473 ↗
  • MalfunctionApr 28, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS DELIVERING LOW VTE (EXHALED TIDAL VOLUME). THERE WAS NO REPORTED PATIENT USE ASSOCIATED WITH THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023180 ↗
  • MalfunctionApr 24, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT A SECTION OF ITS LCD TOUCHSCREEN WAS UNRESPONSIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24975447 ↗
  • MalfunctionApr 22, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT THE VENTILATOR WAS UNABLE TO MAINTAIN PEEP (POSITIVE END EXPIRATORY PRESSURE), RESULTING IN LOWER PEEP THAN SET. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24960737 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K162877. One submission often covers many models, so check the product and lot details in each record.

  • Z-2525-2026 · May 8, 2026 · Open, Classified
    VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
  • Z-1942-2026 · Mar 25, 2026 · Open, Classified
    VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
  • Z-1070-2024 · Jan 9, 2024 · Open, Classified
    VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides c
  • Z-1902-2021 · May 10, 2021 · Open, Classified
    Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900

Other Filter, Bacterial, Breathing-Circuit devices

All Filter, Bacterial, Breathing-Circuit devices →