VOCSN V+Pro, English (REF PRT-01185-000)Recalls & reported problems

Ventec Life Systems, Inc. · V+Pro, English: The VOCSN Unified Respiratory System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It may be used in invasive and non- invasive applications. The VOCSN is intended for pediatric through adult patients weighing at least 5 kg. It is intended for use in home, hospital, institutional and transport settings, including portable applications. www.VentecLife.com · Class 2

All models of VOCSN (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Ventec VOCSN V+Pro is a breathing ventilator. Its 327 reports describe low oxygen readings, unexpected shutdowns, connection problems, and cracked screens. The reports include 2 deaths and 8 injuries. Two recalls name this device, both open: one over an oxygen leak that can occur under high pressure, and one over incorrect parts used in servicing that can cause unexpected shutdowns. Reports fell to 44 in the last 12 months from 82 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
325
Deaths reported
2
Injuries reported
7
Recalls naming it
2

Summary

The FDA's MAUDE database contains 325 adverse event reports that list this exact device (DI 00855573007877), received between Aug 27, 2021 and May 19, 2026 — classified as 2 deaths, 7 injuries, 316 malfunctions.

The most frequently reported specific device problem is “Output Problem” (239 reports).

2 FDA recall records name this device's identifier, 2 still open.

3 other recalls cover devices cleared under the same FDA submission (K162877); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 8 reportsJan 2024: 28 reportsFeb 2024: 17 reportsMar 2024: 11 reportsApr 2024: 8 reportsMay 2024: 6 reportsJun 2024: 3 reportsJul 2024: 8 reportsAug 2024: 6 reportsSep 2024: 6 reportsOct 2024: 1 reportsNov 2024: 9 reportsDec 2024: 3 reportsJan 2025: 12 reportsFeb 2025: 0 reportsMar 2025: 9 reportsApr 2025: 10 reportsMay 2025: 11 reportsJun 2025: 12 reportsJul 2025: 5 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 6 reportsNov 2025: 8 reportsDec 2025: 5 reportsJan 2026: 3 reportsFeb 2026: 7 reportsMar 2026: 7 reportsApr 2026: 7 reportsMay 2026: 5 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202628
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 25, 2026Class IOpen, Classified

    Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

    Z-1942-2026 · Root cause: Process change control · Recall details · FDA record ↗

  2. Feb 3, 2025Class IOpen, Classified

    Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.

    Z-1196-2025 · Root cause: Component change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Output Problem239
  • Unexpected Shutdown49
  • Connection Problem37
  • Display or Visual Feedback Problem14
  • Inaccurate Delivery11
  • Device Alarm System4
  • Use of Device Problem4
  • No Display/Image3

Patient problems

  • No Clinical Signs, Symptoms or Conditions320
  • Distress2
  • Cardiac Arrest1
  • Diaphoresis1
  • Discomfort1
  • Insufficient Information1
  • Low Oxygen Saturation1
  • Sleep Dysfunction1

Recent adverse event reports

The 25 most recent of 325 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25225240 ↗
  • MalfunctionMay 19, 2026Display or Visual Feedback Problem

    IT WAS REPORTED TO VENTEC THAT THE LCD TOUCHSCREEN ON THE DEVICE WAS CRACKED AND UNRESPONSIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25225067 ↗
  • MalfunctionMay 13, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS REBOOTING BY ITSELF. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25164330 ↗
  • MalfunctionMay 13, 2026Display or Visual Feedback Problem

    IT WAS REPORTED TO VENTEC THAT THE LCD TOUCHSCREEN ON THE DEVICE WAS CRACKED AND UNRESPONSIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25162685 ↗
  • MalfunctionMay 12, 2026Inaccurate Delivery

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS INSPIRATORY TIDAL VOLUME (VTI) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25156331 ↗
  • MalfunctionApr 30, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE'S EVALUATION, VENTEC OBSERVED THAT IT WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25053397 ↗
  • MalfunctionApr 20, 2026Unexpected Shutdown

    IT WAS REPORTED TO VENTEC THAT THE DEVICE CONTINUOUSLY REBOOTS BY ITSELF. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24933679 ↗
  • MalfunctionApr 13, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT HAD NO VENTILATION OUTPUT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24853509 ↗
  • MalfunctionApr 7, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE HAD NO VENTILATION OUTPUT. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT; HOWEVER, THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24811342 ↗
  • MalfunctionApr 7, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE HAD NO VENTILATION OUTPUT. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT; HOWEVER, THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24811202 ↗
  • MalfunctionApr 2, 2026Output Problem

    IT WAS REPORTED TO VENTEC THAT THE DEVICE WAS DELIVERING LOWER THAN SET PEEP (POSITIVE END EXPIRATORY PRESSURE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24769793 ↗
  • MalfunctionApr 1, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW (0) AND THAT IT WAS PROVIDING ERRATIC BREATHS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24760702 ↗
  • MalfunctionMar 30, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS REBOOTING BY ITSELF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24721422 ↗
  • MalfunctionMar 25, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THE DEVICE POWERED OFF UNEXPECTEDLY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24692533 ↗
  • MalfunctionMar 24, 2026Output Problem

    IT WAS REPORTED TO VENTEC THAT THE DEVICE HAD DISPLAYED THE PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24678018 ↗
  • MalfunctionMar 18, 2026Unexpected Shutdown

    IT WAS REPORTED TO VENTEC THAT THE DEVICE THAT THE DEVICE HAD POWERED OFF UNEXPECTEDLY DURING USE. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT; HOWEVER, THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE. VENTEC HAS MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION ABOUT THE PATIENT, BUT NO FURTHER INFORMATION HAS BEEN RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24634714 ↗
  • MalfunctionMar 18, 2026Unexpected Shutdown

    IT WAS REPORTED TO VENTEC THAT THE DEVICE CONTINUOUSLY REBOOTS BY ITSELF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24634556 ↗
  • MalfunctionMar 5, 2026Device Alarm System; Output Problem

    IT WAS REPORTED TO VENTEC THAT THE DEVICE HAD NO VENTILATION OUTPUT. IN ADDITION, THE REPORTER ADVISED THAT THE DEVICE DID NOT PROVIDE A PATIENT CIRCUIT DISCONNECT ALARM AS EXPECTED WHEN THE PATIENT HAD BECOME DISCONNECTED FROM THE DEVICE. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT, HOWEVER, THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED EVENT. A BACKUP DEVICE WAS AVAILABLE AND THE PATIENT SURVIVED THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24515795 ↗
  • MalfunctionMar 2, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24492160 ↗
  • MalfunctionFeb 23, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS REBOOTING BY ITSELF. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24424312 ↗
  • MalfunctionFeb 23, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT UNEXPECTEDLY POWERED OFF BY ITSELF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24423731 ↗
  • MalfunctionFeb 16, 2026Power Problem

    IT WAS REPORTED TO VENTEC THAT THE "VENT IN OPT". VENTEC REQUESTED CLARIFICATION REGARDING THE REPORTED ISSUE BUT WAS ADVISED THAT NO FURTHER CLARIFICATION WAS POSSIBLE. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24356626 ↗
  • MalfunctionFeb 13, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24337678 ↗
  • MalfunctionFeb 6, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24284927 ↗
  • MalfunctionFeb 4, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS REBOOTING BY ITSELF. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24264763 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K162877. One submission often covers many models, so check the product and lot details in each record.

  • Z-2525-2026 · May 8, 2026 · Open, Classified
    VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
  • Z-1070-2024 · Jan 9, 2024 · Open, Classified
    VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides c
  • Z-1902-2021 · May 10, 2021 · Open, Classified
    Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900

Other Filter, Bacterial, Breathing-Circuit devices

All Filter, Bacterial, Breathing-Circuit devices →