VOCSN V+O+C+S+N+Pro, EnglishRecalls & reported problems
Ventec Life Systems, Inc. · V+O+C+S+N+Pro, English: The VOCSN Unified Respiratory System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It may be used in invasive and non- invasive applications. The VOCSN is intended for pediatric through adult patients weighing at least 5 kg. It is intended for use in home, hospital, institutional and transport settings, including portable applications. The integral oxygen concentrator is intended for the administration of supplemental oxygen. The integral suction pump is intended for airway fluid removal and oral/ pharyngeal hygiene. The integral cough assist option is intended for patients who are additionally unable to cough or clear secretions effectively. www.VentecLife.com · Class 2
All models of VOCSN (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Ventec VOCSN ventilator has 371 reports, with no deaths and 20 injuries. Most describe output problems and unexpected shutdowns, and some describe the device rebooting during use. There are 3 recalls, and the 2021, 2025 and 2026 ones are all still open, covering shutdowns, wrong parts used in servicing and a damaged back-up battery that may stop the backup alarm from sounding.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 366
- Deaths reported
- 0
- Injuries reported
- 21
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 366 adverse event reports that list this exact device (DI 00855573007792), received between Jun 22, 2019 and Sep 29, 2026 — classified as 21 injuries, 345 malfunctions.
The most frequently reported specific device problem is “Output Problem” (234 reports).
3 FDA recall records name this device's identifier, 3 still open.
2 other recalls cover devices cleared under the same FDA submission (K162877); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 8, 2026Class IIOpen, Classified
Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
Z-2525-2026 · Root cause: Vendor change control · Recall details · FDA record ↗
- Feb 3, 2025Class IOpen, Classified
Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.
Z-1196-2025 · Root cause: Component change control · Recall details · FDA record ↗
- May 10, 2021Class IIOpen, Classified
Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.
Z-1902-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Output Problem234
- Unexpected Shutdown88
- Display or Visual Feedback Problem25
- Connection Problem13
- Device Alarm System11
- Inaccurate Delivery4
- Improper Flow or Infusion3
- Patient Device Interaction Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions339
- No Known Impact Or Consequence To Patient8
- Discomfort5
- Low Oxygen Saturation5
- No Patient Involvement5
- Lethargy2
- Airway Obstruction1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 366 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Use of Device Problem
Patient problems: Discomfort
MAUDE report 50941194 ↗MalfunctionJun 30, 2026Display or Visual Feedback Problem
IT WAS REPORTED TO VENTEC THAT THE LCD TOUCHSCREEN ON THE DEVICE WAS UNRESPONSIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25673945 ↗MalfunctionJun 4, 2026Output Problem
IT WAS REPORTED TO VENTEC THAT THE DEVICE WAS "BARELY BLOWING ANY AIR" HOWEVER NO FURTHER DESCRIPTION WAS PROVIDED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25402338 ↗MalfunctionJun 1, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25352339 ↗MalfunctionMay 27, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) LEVELS. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25311236 ↗MalfunctionMay 27, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25311120 ↗MalfunctionMay 26, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE'S EVALUATION, VENTEC OBSERVED THAT IT WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25285003 ↗MalfunctionMay 21, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT MEASURED LOWER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET AND WAS DISPLAYING A PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25250797 ↗MalfunctionMay 13, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25163286 ↗MalfunctionMay 12, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25156370 ↗MalfunctionMay 7, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE'S EVALUATION, VENTEC OBSERVED THAT IT WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25119708 ↗MalfunctionMay 1, 2026Output Problem
AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE'S EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25059311 ↗MalfunctionApr 30, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE'S EVALUATION, VENTEC OBSERVED THAT IT WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25053352 ↗MalfunctionApr 30, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25038869 ↗MalfunctionApr 29, 2026Display or Visual Feedback Problem
IT WAS REPORTED TO VENTEC THAT THE DEVICE HAD A "PROBLEM WITH THE SCREEN". VENTEC FOLLOWED-UP WITH THE INITIAL REPORTER TO ASK IF THE DEVICE'S LCD TOUCHSCREEN WAS RESPONSIVE TO TOUCHES, AND THEY RESPONDED WITH "WE WERE UNABLE TO DETERMINE THAT, AS THE SCREEN WAS FLICKERING." THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25038447 ↗MalfunctionApr 27, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT THE VENTILATOR MEASURED LOWER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET AND WAS DISPLAYING A PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25009335 ↗MalfunctionApr 16, 2026Unexpected Shutdown
IT WAS REPORTED TO VENTEC THAT THE DEVICE POWERED OFF UNEXPECTEDLY DURING USE, THEN BEGAN TO CONTINUOUSLY REBOOT BY ITSELF. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT; HOWEVER, THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE. THE PATIENT WAS PLACED ON THEIR BACKUP VENTILATOR FOR SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24906140 ↗MalfunctionApr 15, 2026Unexpected Shutdown
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS REBOOTING BY ITSELF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24891848 ↗MalfunctionApr 9, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT HAD NO VENTILATION OUTPUT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24837880 ↗MalfunctionApr 3, 2026Unexpected Shutdown
IT WAS REPORTED TO VENTEC THAT THE DEVICE POWERED OFF UNEXPECTEDLY DURING USE, THEN REBOOTED BY ITSELF. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT; HOWEVER, THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781992 ↗MalfunctionMar 23, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS DISPLAYING THE PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24671474 ↗MalfunctionMar 23, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24671203 ↗MalfunctionMar 20, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS DISPLAYING THE PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24655159 ↗MalfunctionMar 19, 2026Device Alarm System
IT WAS REPORTED TO VENTEC THAT THE DEVICE DID NOT PROVIDE A PATIENT CIRCUIT DISCONNECT ALARM WHEN THE PATIENT WAS REMOVED FROM THE VENTILATOR. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. THE PATIENT HAD GONE INTO CARDIAC ARREST AND PERSONNEL HAD REMOVED THE VENTILATOR IN ORDER TO CARE FOR THE PATIENT. THERE WAS NO ALLEGATION THAT THE DEVICE USE HAD CONTRIBUTED TO THE PATIENT'S CARDIAC ARREST, ONLY THAT PERSONNEL HAD EXPECTED TO HEAR THE ALARM WHEN THE PATIENT WAS DISCONNECTED FROM THE VENTILATOR. THE PATIENT SURVIVED THE REPORTED EVENT. VENTEC MADE MULTIPLE ATTEMPTS TO CONTACT THE INITIAL REPORTER IN ORDER TO OBTAIN PATIENT INFORMATION, BUT NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24645977 ↗MalfunctionMar 18, 2026Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT HAD NO VENTILATION OUTPUT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24635616 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K162877. One submission often covers many models, so check the product and lot details in each record.
- Z-1942-2026 · Mar 25, 2026 · Open, ClassifiedVOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
- Z-1070-2024 · Jan 9, 2024 · Open, ClassifiedVOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides c
Other Filter, Bacterial, Breathing-Circuit devices
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall570 reports
- VOCSN V+Pro Emergency, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit558 reports
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit2 recalls2 deaths325 reports
- VOCSN V+C+S+N+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit185 reports
- GIBECK IPN043751Teleflex Incorporated · Filter, bacterial, breathing-circuit2 recalls7 reports
- VOCSN VOCSN, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 death96 reports
- VOCSN V+C+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall33 reports
- VOCSN V+O+C+S+N+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall32 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page