Skip to content

VOCSN V+Pro Emergency, EnglishRecalls & reported problems

Ventec Life Systems, Inc. · V+Pro Emergency, English: The VOCSN Unified Respiratory System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It may be used in invasive and non- invasive applications. The VOCSN is intended for pediatric through adult patients weighing at least 5 kg. It is intended for use in home, hospital, institutional and transport settings, including portable applications. www.VentecLife.com · Class 2

All models of VOCSN (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Ventec VOCSN ventilator provides breathing support. Its 557 reports mostly describe pressure and breath volumes differing from the settings, plus connection problems and unexpected shutdowns. The reports include no deaths and no injuries, and none describe a patient effect. Reports fell from 215 in the earlier 12 months to 48 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
558
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 558 adverse event reports that list this exact device (DI 00855573007945), received between Aug 25, 2021 and Sep 28, 2026 — classified as 558 malfunctions.

The most frequently reported specific device problem is “Output Problem” (503 reports).

5 other recalls cover devices cleared under the same FDA submission (K162877); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 26 reportsDec 2023: 17 reportsJan 2024: 32 reportsFeb 2024: 14 reportsMar 2024: 14 reportsApr 2024: 7 reportsMay 2024: 5 reportsJun 2024: 6 reportsJul 2024: 3 reportsAug 2024: 2 reportsSep 2024: 8 reportsOct 2024: 15 reportsNov 2024: 40 reportsDec 2024: 8 reportsJan 2025: 24 reportsFeb 2025: 1 reportsMar 2025: 34 reportsApr 2025: 28 reportsMay 2025: 9 reportsJun 2025: 22 reportsJul 2025: 25 reportsAug 2025: 17 reportsSep 2025: 7 reportsOct 2025: 0 reportsNov 2025: 15 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 17 reportsApr 2026: 6 reportsMay 2026: 6 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202640
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem503
  • Connection Problem52
  • Unexpected Shutdown24
  • Device Alarm System5
  • Inaccurate Delivery5
  • Display or Visual Feedback Problem4
  • No Display/Image3
  • Improper Flow or Infusion2

Patient problems

  • No Clinical Signs, Symptoms or Conditions558

Recent adverse event reports

The 25 most recent of 558 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Output Problem

    An authorized service provider (ASP) contacted Ventec to report that the device was delivering higher than set PEEP (Positive End Expiratory Pressure).There were no reports of patient involvement associated with the reported issue.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917801 ↗
  • MalfunctionJun 3, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT WAS UNABLE TO MAINTAIN SET PEEP (POSITIVE END EXPIRATORY PRESSURE). THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25381516 ↗
  • MalfunctionMay 29, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS DELIVERING HIGHER THAN SET PEEP (POSITIVE END EXPIRATORY PRESSURE) CAUSING IT TO DISPLAY THE HIGH PEEP ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25340471 ↗
  • MalfunctionMay 27, 2026Device Alarm System

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT THE DEVICE'S ALARMS WERE NOT WORKING AS EXPECTED. THE DEVICE WAS SILENCING ALARMS UPON START UP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25311465 ↗
  • MalfunctionMay 27, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT MEASURED LOWER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25311167 ↗
  • MalfunctionMay 15, 2026Output Problem

    IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25194671 ↗
  • MalfunctionMay 15, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE'S EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25194669 ↗
  • MalfunctionMay 15, 2026Output Problem

    AN AUTHORIZED THIRD PARTY SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS UNABLE TO MAINTAIN PEEP (POSITIVE END EXPIRATORY PRESSURE), RESULTING IN LOWER PEEP THAN SET, AND THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. IN ADDITION, THE DEVICE WAS DISPLAYING A PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25194395 ↗
  • MalfunctionApr 29, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING INACCURATE FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25029837 ↗
  • MalfunctionApr 28, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE HAD NO VENTILATION OUTPUT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023775 ↗
  • MalfunctionApr 28, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS DISPLAYING THE PATIENT CIRCUIT DISCONNECT ALARM AND WAS DELIVERING LOW INSPIRATORY PRESSURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023604 ↗
  • MalfunctionApr 27, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE PROVIDED AN ALARM INDICATING LOW INSPIRATORY PRESSURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25010599 ↗
  • MalfunctionApr 27, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS DELIVERING LOWER THAN SET PEEP (POSITIVE-END EXPIRATORY PRESSURE) AND LOW VTE (EXHALED TIDAL VOLUME). ADDITIONALLY, THE DEVICE DISPLAYED THE PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25010580 ↗
  • MalfunctionApr 27, 2026Unexpected Shutdown

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS REBOOTING BY ITSELF. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25010521 ↗
  • MalfunctionMar 30, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS, THAT IT MEASURED LOWER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET, AND ITS VTE (EXHALED TIDAL VOLUME) LEVELS WERE LOW. ADDITIONALLY, THE DEVICE WAS DISPLAYING A PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24736887 ↗
  • MalfunctionMar 30, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24733210 ↗
  • MalfunctionMar 30, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24716708 ↗
  • MalfunctionMar 30, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24715997 ↗
  • MalfunctionMar 30, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24708949 ↗
  • MalfunctionMar 25, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24694719 ↗
  • MalfunctionMar 25, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24694450 ↗
  • MalfunctionMar 25, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24694422 ↗
  • MalfunctionMar 25, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE'S EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. ADDITIONALLY, THE DEVICE WAS DISPLAYING THE PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24694399 ↗
  • MalfunctionMar 24, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP), CONTACTED VENTEC TO REPORT THAT THE DEVICE'S EXHALED TIDAL VOLUME (VTE) LEVELS WERE NOT WITHIN SPECIFICATION. VENTEC REACHED OUT TO THE ASP FOR CLARIFICATION AND WAS ADVISED THAT THE TECHNICIAN DID NOT DOCUMENT THE VALUES OBSERVED. VENTEC HAS ELECTED TO REPORT THIS OUT OF AN ABUNDANCE OF CAUTION, AS THE DEVICE'S VTE LEVELS MAY HAVE BEEN LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24678221 ↗
  • MalfunctionMar 24, 2026Output Problem

    AN AUTHORIZED SERVICE PROVIDER (ASP) CONTACTED VENTEC TO REPORT THAT THE DEVICE'S EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW AND THAT IT WAS DISPLAYING A PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24678200 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K162877. One submission often covers many models, so check the product and lot details in each record.

  • Z-2525-2026 · May 8, 2026 · Open, Classified
    VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
  • Z-1942-2026 · Mar 25, 2026 · Open, Classified
    VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
  • Z-1196-2025 · Feb 3, 2025 · Open, Classified
    VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002
  • Z-1070-2024 · Jan 9, 2024 · Open, Classified
    VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides c
  • Z-1902-2021 · May 10, 2021 · Open, Classified
    Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900

Other Filter, Bacterial, Breathing-Circuit devices

All Filter, Bacterial, Breathing-Circuit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page