SharkCore DSC-25-01Recalls & reported problems
Covidien LP · Fine Needle Biopsy System · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00855208005094), received between Jan 31, 2019 and Jul 13, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Difficult to Insert” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Insert1
- Retraction Problem1
Patient problems
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO LITERATURE SOURCE OF STUDY PERFORMED BETWEEN (B)(6) 2018 AND (B)(6) 2019, ONE PATIENT HAD INTRA-PROCEDURAL PERITONEAL BLEEDING WITH A 4 G/DL HEMOGLOBIN LEVEL DROP AFTER THE EUS-FNB (ENDOSCOPIC ULTRASOUND FINE-NEEDLE BIOPSY PROCEDURE). THE PATIENT WAS A RENAL CANCER SAMPLED METASTASIS SAMPLED WITH A 25-GAUGE FORK-TIP NEEDLE. THE PATIENT WAS HOSPITALIZED BUT NO BLOOD TRANSFUSIONS WERE REQUIRED.
Patient problems: No Code Available
MAUDE report 10264918 ↗MalfunctionJan 31, 2019Difficult to Insert; Retraction Problem; Difficult to Insert; Retraction Problem
ACCORDING TO THE REPORTER, THE FIRST PASS WAS DONE CORRECTLY AND THE NEEDLE WAS EXTRACTED FROM THE SHEATH TO REMOVE THE SAMPLE, BUT, IT WAS NO LONGER POSSIBLE TO REINSERT THE NEEDLE. THE NEEDLE WAS NOT FOLDED, BUT, IT BLOCKED THE HEIGHT OF THE HAND PIECE AND THE SAFETY SHEATH DIS NOT LOCK IN PLACE. THE INCIDENT OCCURRED ON THE FIRST NUMBER PASS AND THEY CHANGED THE NEEDLE TO CORRECT THE PROBLEM. THERE WAS NO PATIENT INJURY. A REPEAT PROCEDURE WAS NECESSARY BUT THERE WAS NO EXTENDED HOSPITALIZATION OR MEDICAL OR SURGICAL INTERVENTION NEEDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8294730 ↗
Other Biopsy needle devices
- MAGNUM MN1410Bard Peripheral Vascular, Inc. · Biopsy needle1 recall118 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4021Olympus Medical Systems Corp. · Biopsy needle1 recall94 reports
- EchoTip G31520Cook Ireland Ltd · Biopsy needle1 recall84 reports
- Single Use Aspiration Needle NA-201SX-4021Olympus Medical Systems Corp. · Biopsy needle1 death126 reports
- EchoTip ProCore G34785Cook Ireland Ltd · Biopsy needle97 reports
- MAGNUM MN1820Bard Peripheral Vascular, Inc. · Biopsy needle1 recall37 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4022Olympus Medical Systems Corp. · Biopsy needle1 recall29 reports
- Interject M00518250Boston Scientific Corporation · Biopsy needle70 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.