Lithotripsy Transducer ShockPulse-SE SPL-TRecalls & reported problems
Cybersonics Inc · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Cybersonics ShockPulse lithotripsy transducer is used to break up kidney stones. The 24 reports mainly describe the transducer breaking, leaking, failing to start or produce output, and showing errors; there were no deaths or injuries. There is 1 open recall, started in October 2025, because the device may fail to start or stop shortly after starting, while the handpiece may heat up during use.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00855279005023), received between Jan 21, 2021 and Nov 8, 2024 — classified as 24 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 8, 2025Class IIOpen, Classified
The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.
Z-0452-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break11
- Communication or Transmission Problem2
- Energy Output Problem2
- Power Problem2
- Connection Problem1
- Defective Component1
- Defective Device1
- Electrical /Electronic Property Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Patient Involvement2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 8, 2024Energy Output Problem; Energy Output Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED NO OUTPUT, WHICH FAILED THE PROBE TEST. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20642650 ↗MalfunctionJun 21, 2024Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE SUBJECT DEVICE EXPERIENCED AN ERROR CODE DISPLAYED WHEN PLUGGED INTO THE TRANSDUCER DURING SET UP. THE ISSUE WAS OBSERVED DURING PREPARATION FOR USE FOR A LITHOTRIPSY PROCEDURE TO BREAK DOWN KIDNEY STONES. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19580232 ↗MalfunctionMay 2, 2024Break; Break
IT WAS REPORTED THE SHOCK PULSE LITHOTRIPSY TRANSDUCER HAD A BROKEN TRANSMITTER. THE ISSUE OCCURRED DURING AN UNKNOWN PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19233168 ↗MalfunctionApr 26, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THE LITHOTRIPSY TRANSDUCER SUCTION ADJUSTMENT KNOB WAS STUCK. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19195869 ↗MalfunctionApr 23, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THERE WAS FLUID LEAKING FROM A HOLE ON THE SUCTION CONTROL RING OF THE LITHOTRIPSY TRANSDUCER. THE EVENT OCCURRED DURING A THERAPEUTIC PERCUTANEOUS NEPHROLITHOTRIPSY. THE PROCEDURE WAS DELAYED IN THE MIDDLE OF THE PROCEDURE TO OBTAIN ANOTHER SIMILAR DEVICE IN ORDER TO CONTINUE TO COMPLETION OF THE PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19167718 ↗MalfunctionOct 26, 2023Break; Break
IT WAS REPORTED TO OLYMPUS THAT THE SHOCKPULSE LITHOTRIPSY TRANSDUCER DURING A THERAPEUTIC PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE THERE WAS A PHYSICAL DEFECT OF THE CONNECTING PIECE, AND THE END WAS CRACKED. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE DEVICE WAS EVALUATED, AND IT WAS FOUND THAT THE DEVICE SUCTION PORT END WAS BROKEN. THIS MDR (MEDICAL DEVICE REPORT) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18010896 ↗MalfunctionSep 26, 2023Leak/Splash; Leak/Splash
THE CUSTOMER REPORTED THAT HIS OLYMPUS SHOCKPULSE LITHOTRIPSY TRANSDUCER HAD A LEAK UPON OPENING THE PACKAGE. THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17817463 ↗MalfunctionSep 26, 2023Power Problem; Power Problem
THE CUSTOMER REPORTED THAT HIS OLYMPUS SHOCKPULSE LITHOTRIPSY TRANSDUCER ULTRASOUND WAS NOT WORKING DURING PROCEDURE PREPARATION. ACCORDING TO THE INITIAL REPORTER ANOTHER LASER WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WAS NO PATIENT HARM REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17817362 ↗MalfunctionApr 13, 2023Energy Output Problem; Energy Output Problem
IT WAS REPORTED THE DEVICE WAS NOT WORKING, THE ERROR LIGHT ON THE DEVICE WAS HIGHLIGHTED. THE ISSUE WAS FOUND DURING A PERCUTANEOUS NEPHRECTOMY PROCEDURE (THERAPEUTIC). THE DEVICE WAS REPLACED AND THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS A DELAY OF APPROXIMATELY TWENTY (20) MINUTES IN THE PROCEDURE, HOWEVER, PER THE REPORT, THERE WAS NO PATIENT HARM OR ADVERSE EVENT OCCURRED DUE TO THE DELAY. NO HARM WAS REPORTED, NO PATIENT INJURY, NO USER INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16729792 ↗MalfunctionNov 1, 2022Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED TO OLYMPUS WHILE PREPARING FOR USE, A TRANSDUCER ERROR CODE AND OTHER ERROR MESSAGES DISPLAYED WHEN THE SHOCKPULSE LITHOTRIPSY TRANSDUCER WAS PLUGGED IN. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15709725 ↗MalfunctionOct 19, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
AN OLYMPUS REP REPORTED ON BEHALF OF THE CUSTOMER, THE SHOCKPULSE LITHOTRIPSY TRANSDUCER DISPLAYED AN ERROR CODE. BOTH THE ERROR LIGHT AND PROBE CHECK LIGHT WERE LIT. THE EVENT OCCURRED IN THE MIDDLE OF A THERAPEUTIC LEFT COMPLEX PERCUTANEOUS NEPHROLITHOTOMY. THE PROCEDURE WAS DELAYED WHILE THE USER FACILITY TRIED TROUBLESHOOTING WITHOUT SUCCESS; AND HAD TO BE COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15633929 ↗MalfunctionAug 10, 2022Break; Break
THE CUSTOMER REPORTED THAT, DURING PREPARATION FOR USE, THE SUCTION CONNECTOR ON THE SUBJECT DEVICE WAS FOUND TO BE DAMAGED. THE DEVICE WAS SENT TO AN OLYMPUS SERVICE CENTER FOR EVALUATION. DURING INSPECTION AND TESTING, SERVICE FOUND THE TIP OF THE SUCTION PORT WAS BROKEN AND DAMAGED. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION (BROKEN SUCTION PORT). THERE WAS NO EFFECT ON THE PATIENT DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15200081 ↗MalfunctionJun 10, 2022Break; Break
A USER FACILITY RETURNED THE OLYMPUS ASSET SHOCKPULSE LITHOTRIPSY TRANSDUCER, TO THE OLYMPUS SERVICE CENTER. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THE SUCTION PORT WAS BROKEN OFF. THIS REPORT IS BEING SUBMITTED FOR THE EVENT FOUND DURING EVALUATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14653361 ↗MalfunctionMay 27, 2022Break; Break
THE OLYMPUS ASSET DEVICE, SHOCKPULSE LITHOTRIPSY TRANSDUCER, WAS SENT TO THE OLYMPUS SERVICE CENTER. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS DISCOVERED THE SUCTION PORT HAD BROKEN OFF. THERE WAS NOT PATIENT INVOLVEMENT. THE MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION OF THE SUCTION PORT BROKEN OFF FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14522785 ↗MalfunctionFeb 3, 2022Defective Device; Output Problem; Defective Device; Output Problem
THE CUSTOMER REPORTED TRANSDUCER ERROR CODE . DEFECTIVE - ERROR ON THE SCREEN. THE ISSUE FOUND DURING AN UNKNOWN EVENT. THERE IS NO PATIENT INVOLVEMENT ASSOCIATED ON THIS REPORTED EVENT. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13447297 ↗MalfunctionJan 28, 2022Break; Break
IT WAS REPORTED THAT "THE DEVICE BLEW OFF THE TIP DURING THE ONGOING SURGICAL APPLICATION. TIP FOR SUCTION DEVICE INTRAOPERATIVELY BROKEN OFF." THERE WAS NO PATIENT INJURY OR INFECTION REPORTED. NO USER INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13387746 ↗MalfunctionSep 17, 2021Break; Defective Component; Break; Defective Component
DURING AN ASSET RETURN INSPECTION THE DEVICE WAS FOUND WITH DAMAGED SUCTION PORT. THERE IS NO PATIENT INVOLVEMENT ASSOCIATED ON THIS REPORTED EVENT. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12492572 ↗MalfunctionJul 22, 2021Break; Break
IT WAS REPORTED THAT THE DEVICE SUCTION PORT WAS BROKEN. THE ISSUE FOUND DURING AN ASSET RETURN INSPECTION. THERE WAS NO PATIENT INVOLVEMENT REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12215308 ↗MalfunctionJul 9, 2021Break; Break
IT WAS REPORTED THAT THE PLASTIC TIP OF THE DEVICE IS BROKEN. THE ISSUE FOUND DURING PREPARATION FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12147249 ↗MalfunctionJun 16, 2021Break; Break
THIS IS AN ASSET RETURN WHICH WAS FOUND TO HAVE A BROKEN SUCTION PORT DURING ESTIMATION. THERE WAS NO DEVICE ISSUE OR PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12008955 ↗MalfunctionApr 21, 2021Connection Problem; Connection Problem
IT WAS REPORTED, THE SHOCKPULSE LITHOTRIPSY TRANSDUCER WAS NOT RECOGNIZED BY THE GENERATOR. ACCORDING TO THE INITIAL REPORTER, THE DEVICE WAS BROKEN ON THE END AND WAS NOT FUNCTIONING PROPERLY. THERE WAS NO PATIENT HARM OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11701419 ↗MalfunctionJan 29, 2021Fracture; Fracture
AS REPORTED, THE SECTION OF THE SHOCKPULSE LITHOTRIPSY TRANSDUCER THAT ATTACHES TO THE MONITOR WAS NOTED TO BE BROKEN DURING DEVICE MAINTENANCE. THERE WAS NOT REPORTED PATIENT INVOLVEMENT FROM THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 11252123 ↗MalfunctionJan 28, 2021Break; Break
THE SERVICE CENTER WAS INFORMED DURING PREPARATION FOR USE, THERE WAS A PROBLEM WITH THE PROBE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 11247405 ↗MalfunctionJan 21, 2021Power Problem; Power Problem
AS REPORTED, THE POWER CORD ON THE SHOCKPULSE LITHOTRIPSY TRANSDUCER HAS A BROKEN CONSOLE CONNECTOR. THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE FROM THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11210626 ↗
Other Lithotriptor, Ultrasonic devices
- ShockPulse SPL-TGyrus ACMI, LLC · Lithotriptor, ultrasonic279 reports
- ShockPulse SPL-SGyrus ACMI, LLC · Lithotriptor, ultrasonic2 recalls159 reports
- SHOCKPULSE SPL-SRGyrus ACMI, LLC · Lithotriptor, Ultrasonic2 recalls97 reports
- ShockPulse SPL-GGyrus ACMI, LLC · Lithotriptor, ultrasonic2 recalls10 reports
- ShockPulse SPL-FSGyrus ACMI, LLC · Lithotriptor, ultrasonic20 reports
- ShockPulse Lithotripsy Generator SPL-GCybersonics Inc · Lithotriptor, Ultrasonic8 reports
- Lithotripsy System ShockPulse-SE SPL-SCybersonics Inc · Lithotriptor, Ultrasonic7 reports
- Lithotripsy Probe, Sterile Single Use ShockPulse -SE, 11.3 Fr (3.76mm) X 396mm SPL-PD376Cybersonics Inc · Lithotriptor, Ultrasonic6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page