ShockPulse Lithotripsy Generator SPL-GRecalls & reported problems
Cybersonics Inc · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00855279005054), received between Nov 4, 2020 and Sep 5, 2023 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Output Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Output Problem4
- Defective Component2
- Material Twisted/Bent2
- Communication or Transmission Problem1
- Electrical /Electronic Property Problem1
Patient problems
- No Patient Involvement4
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2023Communication or Transmission Problem; Communication or Transmission Problem
A USER FACILITY REPORTED TO OLYMPUS WHEN THE PEDAL ON SHOCKPULSE-SE LITHOTRIPSY SYSTEM IS PRESSED, AN ERROR MESSAGE POPS UP AND THE UNIT WILL NOT WORK. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17683276 ↗MalfunctionJun 30, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
A USER FACILITY REPORTED THAT THE SHOCKPULSE-SE LITHOTRIPSY SYSTEM STARTS WITH AN ERROR. THE EVENT OCCURRED DURING PREPARATION FOR USE. DURING A STANDARD SERVICE INSPECTION OF THE CUSTOMER RETURNED DEVICE, DAMAGED BOARDS WERE OBSERVED. THIS REPORT IS TO CAPTURE THE REPORTABLE MALFUNCTION FOUND AT INSPECTION. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14879965 ↗MalfunctionJan 28, 2021Output Problem; Output Problem
IT WAS REPORTED THAT DURING AN UNSPECIFIED PROCEDURE THE DEVICE WAS FOUND NOT PRODUCING NORMAL ENERGY , INTERFERENCE SAYING ELECTRICAL SHORT MESSAGE, NO FURTHER INFORMATION PROVIDED. NO PATIENT IMPACT OR HARM REPORTED. NO USER INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11246061 ↗MalfunctionJan 21, 2021Output Problem; Output Problem
IT WAS REPORTED THAT THE DEVICE WAS FOUND WITH NO OUTPUT FROM THE GENERATOR. THE USER WAS REPORTED TO BE NOT USING A FOOTSWITCH AND ONLY THE BUTTONS ON THE TRANSDUCER. THE HIGH POWER INDICATOR ON THE FRONT PANEL ILLUMINATES WHEN THE BUTTON ON THE TRANSDUCER IS PUSHED HOWEVER NO OUTPUT PRODUCED. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY REPORTED ON THIS EVENT. NO USER INJURY REPORTED.
Patient problems: No Patient Involvement
MAUDE report 11210891 ↗MalfunctionJan 20, 2021Defective Component; Material Twisted/Bent; Defective Component; Material Twisted/Bent
IT WAS REPORTED THAT DURING PREPARATION FOR USE THE DEVICE WAS FOUND WITH BROKEN TRANSDUCER RECEPTACLE. BENT PINS WERE FOUND INSIDE THE TRANSDUCER. THERE WAS NO PATIENT INVOLVEMENT ON THIS EVENT REPORTED . NO USER INJURY REPORTED.
Patient problems: No Patient Involvement
MAUDE report 11204764 ↗MalfunctionJan 13, 2021Defective Component; Defective Component
IT WAS REPORTED THAT DURING AN ASSET RETURN INSPECTION THE DEVICE WAS OBSERVED WITH FAULTY TRANSDUCER RECEPTACLE. THERE WAS NO PATIENT INVOLVEMENT ON THIS EVENT REPORTED. NO USER INJURY REPORTED.
Patient problems: No Patient Involvement
MAUDE report 11163830 ↗MalfunctionNov 11, 2020Output Problem; Output Problem
UPDATED INFORMATION ON THE EVENT REPORTED. THE EVENT OCCURRED DURING A THERAPEUTIC PCNL(PERCUTANEOUS NEPHROLITHOTOMY ) PROCEDURE. NO OTHER DEVICES WERE INVOLVED IN THE EVENT OR REPLACED DURING THE COURSE OF THE PROCEDURE. THE PROCEDURE WAS DELAYED 5 MINUTES NO INFORMATION WAS PROVIDED AS TO WHETHER OR NOT THE PATIENT WAS UNDER GENERAL ANESTHESIA. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. NO PATIENT INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10824958 ↗MalfunctionNov 4, 2020Material Twisted/Bent; Output Problem; Material Twisted/Bent; Output Problem
IT WAS REPORTED THAT DURING A THERAPEUTIC PROCEDURE THE DEVICE WAS FOUND WITH CONNECTION ISSUES. THE PINS INSIDE THE CONNECTOR WERE FOUND BROKEN. THERE WERE NO FURTHER DETAILS PROVIDED REGARDING THE EVENT. NO PATIENT HARM OR INJURY WAS REPORTED. NO USER INJURY REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10784665 ↗
Other Lithotriptor, Ultrasonic devices
- ShockPulse SPL-TGyrus ACMI, LLC · Lithotriptor, ultrasonic279 reports
- ShockPulse SPL-SGyrus ACMI, LLC · Lithotriptor, ultrasonic2 recalls159 reports
- SHOCKPULSE SPL-SRGyrus ACMI, LLC · Lithotriptor, Ultrasonic2 recalls97 reports
- ShockPulse SPL-GGyrus ACMI, LLC · Lithotriptor, ultrasonic2 recalls10 reports
- Lithotripsy Transducer ShockPulse-SE SPL-TCybersonics Inc · Lithotriptor, Ultrasonic1 recall24 reports
- ShockPulse SPL-FSGyrus ACMI, LLC · Lithotriptor, ultrasonic20 reports
- Lithotripsy System ShockPulse-SE SPL-SCybersonics Inc · Lithotriptor, Ultrasonic7 reports
- Lithotripsy Probe, Sterile Single Use ShockPulse -SE, 11.3 Fr (3.76mm) X 396mm SPL-PD376Cybersonics Inc · Lithotriptor, Ultrasonic6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.