EVO sEEG Electrode NSEEG3410-RFRecalls & reported problems

Neuroone Medical Technologies Corporation · 33.5mm x 10 Contact sEEG - RF Probe · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00856895008597), received between Aug 27, 2026 and Aug 27, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Malposition of Device1

Patient problems

  • Intracranial Hemorrhage1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2026Malposition of Device

    A report was received that on 04-Aug-2026 the implantation of sEEG depth electrodes in one patient went smoothly and all implanted electrodes were recording. A planned post-operative CT scan was performed of the implanted electrodes. During the CT it was noted that there was a bleed. The patient was taken back to the OR where they removed all of the electrodes, performed a craniotomy to remove a clot and used an aneurysm clip to stop the bleed.Post implant, the surgical team, reviewed the trajectories, data and surgical plan. It was confirmed that the implanted Anchor Bolts all matched the pre-planned trajectories, including trajectory 12. The CT scan showed that the tip of the electrode was off trajectory 12 by about 10 to 15 mm within 1 to 2 mm of entering the cranial space. The electrode tip entered the sylvian fissure and hit a branch of the Middle Cerebral Artery resulting in a large bleed.Post-operatively the physician said that he wasn't sure what caused the electrode deviation, and that it could have been caused by a number of things, such as "vascular structure, dura mater deflection, or an incomplete drill hole." The physician commented that he didn't think that it was ne

    Patient problems: Intracranial Hemorrhage

    MAUDE report 50487391 ↗

Other Probe, Radiofrequency Lesion devices