Comprehensive Shoulder System recallZ-0979-2018 · Zimmer Biomet, Inc. · initiated Dec 8, 2017
Class II Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer Biomet is conducting a medical device recall for two lots of Comprehensive Humeral stems. The lots were commingled during the manufacturing process and were etched with incorrect lot numbers. The size difference between the two stem sizes is 1.1 inches. It is visually recognizable by the user that the box label does not match the contents of the box.
FDA root cause category: Under Investigation by firm
What the company is doing
On December 13, 2017 an URGENT MEDICAL DEVICE RECALL letter was issued to customers notifying them of the mix-up and requesting that they review the notification, discuss with team members, see if any product is in inventory and if so, quarantine and discontinue use. Affected product should be returned to the recalling firm. Questions or concerns can be directed to customer service at 574-371-3071.
Recall details
- Recall number
- Z-0979-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 8, 2017
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 12
- Product code
- HSD
- FDA submissions
- K060692
- Distribution
- US Nationwide in the states of MN, NJ, CA, TX, ND
Product description
Humeral Stem 83mm(Size 17) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
Code information (models, lots, serial numbers)
PN: 113633, LN: 421700, UDI# - (01) 00880304462632 (17) 271014 (10) 421700
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- COMPREHENSIVE SHOULDER SYSTEM 113633Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 recalls4 reports
Product lines
Other devices under the same FDA submission
45 devices cleared under K060692 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- COMPREHENSIVE SHOULDER SYSTEM 113629Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall2 reports
- COMPREHENSIVE SHOULDER SYSTEM 113631Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall2 reports
- COMPREHENSIVE SHOULDER SYSTEM 113618Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall2 reports
- Comprehensive® Shoulder System 113627Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall2 reports
- COMPREHENSIVE SHOULDER SYSTEM 113635Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 recall1 report
- COMPREHENSIVE SHOULDER SYSTEM 113611Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 recall1 report
- COMPREHENSIVE SHOULDER SYSTEM 113617Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall1 report
- Comprehensive® Shoulder System 113607Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall1 report
- COMPREHENSIVE SHOULDER SYSTEM 113632Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 recall0 reports
- COMPREHENSIVE SHOULDER SYSTEM 113614Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall0 reports
- Comprehensive® Shoulder System 113605Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall0 reports
- Comprehensive® Shoulder System 113606Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 79385).
- Z-0980-2018 · TerminatedHumeral Stem 55mm(Size 13) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page