Skip to content

LIFEPAK 15 monitor/defibrillator 97577 (REF 97577-000101)Recalls & reported problems

Physio-Control, Inc. · 15AACBABBABBBAAAAAAAAAAFLP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00883873808214), received between May 12, 2022 and Jan 18, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

16 other recalls cover devices cleared under the same FDA submission (P160026); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2024Class IIOpen, Classified

    Due to an out of tolerance tool being used on monitor/defibrillator systems.

    Z-0230-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Power Up1
  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 18, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE POWERED OFF WHILE MONITORING A PATIENT. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18533067 ↗
  • MalfunctionMay 12, 2022Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS INTERMITTENTLY NOT POWERING ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14375937 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160026. One submission often covers many models, so check the product and lot details in each record.

  • Z-2065-2026 · Apr 13, 2026 · Open, Classified
    LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001
  • Z-2497-2025 · Jul 22, 2025 · Open, Classified
    LIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 99577-000046 / 883873988022 V15-2-000099 / 99577-000093 / 883873979020 V
  • Z-2498-2025 · Jul 22, 2025 · Open, Classified
    LIFEPAK 1000 Part / Catalog Numbers / UDI-DI code: 320371500023 / 99425-000025 / 885074698953 320371500229 / 99425-000025 / 885074698953 331706400001 / 99425-000025 / 885074698953
  • Z-1232-2025 · Feb 3, 2025 · Open, Classified
    LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.
  • Z-1184-2025 · Jan 21, 2025 · Open, Classified
    LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sen
  • Z-0567-2024 · Nov 6, 2023 · Open, Classified
    AC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098
  • Z-0300-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001373, Part Number V15-7-000002
  • Z-0299-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001255, Part Number V15-2-001602
  • Z-0301-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001955, Part Number V15-5-000584 and V15-7-000020
  • Z-0302-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001956, Part Number V15-7-000021
  • Z-0298-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-000061, Part Number V15-2-000067
  • Z-0304-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001958, Part Number: V15-7-000023
  • Z-0303-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001957, Part Number V15-5-000586 and V15-7-000022
  • Z-0305-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001960, Part Number: V15-7-000026
  • Z-0296-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-000046, Part Number V15-2-000051

…and 1 more.

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources and updates