LIFEPAK 15 monitor/defibrillator recallZ-0230-2025 · Physio-Control, Inc. · initiated Aug 30, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Physio-Control, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to an out of tolerance tool being used on monitor/defibrillator systems.
FDA root cause category: Vendor change control
What the company is doing
On 09/30/2024, the firm mailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them that Stryker discovered that an out of tolerance service tool was used on affected LP15 devices in which there is a potential for Noninvasive Blood Pressure (NIBP), System or Auxiliary Power Connectors to be loose. There is a possibility that water or dust may accumulate in the ingress and can lead to permanent or intermittent failures. Customers are instructed to: 1. Immediately check internal inventory to locate the product listed on the attached business reply form. 2. To confirm receipt of this Medical Device Correction Notice and understanding of the provided information, email the enclosed Business Reply Form (BRF) by October 31st, 2024 date, to [email protected]. 3. Upon receipt of the completed business reply form, Stryker will contact customers to arrange for corrective action on the device. 4. There are no additional actions that users can and should take. Please continue to follow the Operating Instructions and perform the daily checks and user tests. These tests can be found in the chapter Maintaining the Equipment subsection General Maintenance and Testing . 5. In the interim, Stryker recommends that you continue to use your LP15 device. 6. Maintain awareness of this communication internally until all required actions have been completed within your facility. Please ensure this letter is kept with the affected device until the correction has been completed. 7. Inform Stryker if any of the subject devices have been distributed to other organizations by contacting Stryker at: a. Device Tracking: [email protected] b. Website: https://www.strykeremergencycare.com/service-support-overview/device-registration/ If you have any questions or concerns, please contact Customer Service +1 800 442 1142 or by email [email protected].
Recall details
- Recall number
- Z-0230-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 30, 2024
- Posted by the FDA
- Oct 25, 2024
- Recalling firm
- Physio-Control, Inc.
- Quantity
- 105 systems
- Product code
- MKJ
- FDA submissions
- P160026
- Distribution
- U.S.: CT, GA, MA, NY, OH, WA, and WV. O.U.S.: N/A
Product description
LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99577-001955 o 99577-001956 o 99577-001957 o 99577-001958 o 99577-001962 LIFEPAK 15 intended use monitor/defibrillator.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99577-001955 o 99577-001956 o 99577-001957 o 99577-001958 o 99577-001962 UDI-DI codes: o 00883873808214 o 00883873988022 o 00883873979426 o 00883873979358 o 00883873911693 o 00883873979198 o 00883873911648 o 00883873884386 o 00883873871713 o 00883873871706 o 00883873871690 o 00883873871683 o 00883873911631 o 00883873871645 Serial Numbers: o 45435833 o 46470647 o 46686048 o 46686573 o 46686708 o 46687360 o 46906073 o 49037777 o 49704552 o 38166647 o 38282988 o 38290136 o 38291331 o 38802782 o 39541663 o 39541666 o 39752193 o 40259639 o 41446867 o 41447020 o 41453331 o 42325059 o 43574409 o 43984215 o 45040678 o 45308104 o 45455023 o 46017745 o 46572054 o 46576300 o 48349664 o 48605645 o 48605663 o 49001744 o 49025185 o 49070762 o 49087040 o 49087053 o 49224619 o 49374174 o 49422314 o 49619921 o 38200440 o 38805564 o 38805583 o 39382735 o 41446693 o 41446882 o 41446975 o 41446977 o 41447027 o 41453239 o 41893891 o 42324176 o 42610924 o 43181935 o 43182268 o 43182389 o 43462227 o 43465372 o 43487806 o 43610792 o 43791346 o 43841374 o 43860850 o 43862247 o 43869747 o 44169862 o 44356379 o 44360003 o 44601246 o 45590467 o 45907317 o 46364496 o 46691551 o 47447195 o 47515791 o 47539203 o 47539310 o 47785570 o 47794250 o 48524850 o 48583111 o 48585709 o 48600317 o 48605407 o 48605467 o 48704914 o 48749443 o 49022830 o 49031342 o 49036187 o 49056308 o 49057928 o 49123367 o 49126651 o 49127012 o 49132399 o 49132740 o 49134638 o 49511402 o 49513814
Devices named in this recall
14 devices in the FDA device registry whose identifier appears in the recall's code information.
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)3 recalls7 deaths755 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)2 recalls5 deaths540 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)2 recalls7 deaths402 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)2 recalls3 deaths376 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)2 recalls4 deaths273 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)3 recalls189 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)2 recalls1 death110 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)2 recalls1 death100 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)2 recalls2 deaths50 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall1 death26 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall16 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall4 reports
- LIFEPAK 15 monitor/defibrillator 97577Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall2 reports
- LIFEPAK® 15 monitor/defibrillator 99577Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall0 reports
Product lines
Other devices under the same FDA submission
26 devices cleared under P160026 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 death228 reports
- LIFEPAK 1000 defibrillator 99425Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 death188 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99402Physio-Control, Inc. · Automated external defibrillators (non-wearable)4 deaths173 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)4 deaths129 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)124 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)5 deaths101 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)91 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)3 deaths78 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99402Physio-Control, Inc. · Automated external defibrillators (non-wearable)77 reports
- LIFEPAK 1000 defibrillator 99425Physio-Control, Inc. · Automated external defibrillators (non-wearable)76 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall3 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99402Physio-Control, Inc. · Automated external defibrillators (non-wearable)38 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page