LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001955)Recalls & reported problems

Physio-Control, Inc. · 15AABBABBABBBAAAAAAAAAAFLP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This LIFEPAK 15 monitor and defibrillator from Physio-Control delivers shocks to restore a normal heartbeat. The FDA database holds 540 reports for it, many describing units that shut off unexpectedly or fail to power up, with 2 injuries and 5 deaths. Two open recalls name this device, over a manufacturing tool that was out of specification and battery pins that can loosen and cause power problems.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
540
Deaths reported
5
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 540 adverse event reports that list this exact device (DI 00883873871713), received between Mar 27, 2018 and Jun 29, 2026 — classified as 5 deaths, 2 injuries, 533 malfunctions.

The most frequently reported specific device problem is “Unexpected Shutdown” (171 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 8 reportsJan 2024: 2 reportsFeb 2024: 19 reportsMar 2024: 7 reportsApr 2024: 15 reportsMay 2024: 9 reportsJun 2024: 13 reportsJul 2024: 11 reportsAug 2024: 9 reportsSep 2024: 12 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2024Class IIOpen, Classified

    Due to an out of tolerance tool being used on monitor/defibrillator systems.

    Z-0230-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

  2. Sep 29, 2023Class IIOpen, Classified

    Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.

    Z-0301-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unexpected Shutdown171
  • Failure to Power Up87
  • Defibrillation/Stimulation Problem78
  • Electrical /Electronic Property Problem51
  • Key or Button Unresponsive/not Working30
  • Failure to Charge26
  • Device Sensing Problem20
  • Failure to Deliver Shock/Stimulation17

Patient problems

  • No Clinical Signs, Symptoms or Conditions408
  • No Patient Involvement86
  • No Known Impact Or Consequence To Patient46
  • Cardiac Arrest5
  • Death2
  • No Consequences Or Impact To Patient2
  • Anemia1
  • Asystole1

Recent adverse event reports

The 25 most recent of 540 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 29, 2026Connection Problem

    DURING REGULAR SERVICE STRYKER OBSERVED THE THERAPY CABLE CONNECTION WAS BROKEN. IN THIS STATE DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25649508 ↗
  • MalfunctionMar 3, 2026Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, STRYKER OBSERVED THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY.THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24495951 ↗
  • MalfunctionFeb 24, 2026Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER DURING INTERVENTION ON A PATIENT. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427519 ↗
  • MalfunctionOct 8, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE MAIN KEYPAD WAS NOT FULLY RESPONSIVE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20394160 ↗
  • MalfunctionOct 2, 2024Power Problem; Power Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE SHUT OFF AND WILL NOT STAY ON WITH BATTERIES OR AUXILIARY POWER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20355160 ↗
  • MalfunctionSep 26, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE IS UNEXPECTEDLY POWERING OFF. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20311936 ↗
  • MalfunctionSep 25, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    STRYKER WAS PERFORMING PREVENTATIVE MAINTENANCE ON THE CUSTOMER'S DEVICE AND OBSERVED THAT THE DEVICE' MAIN KEYPAD WAS UNRESPONSIVE. AS A RESULT, DEFIBRILLATION THERAPY MAY HAVE BEEN DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20302545 ↗
  • MalfunctionSep 24, 2024Unexpected Shutdown; Unexpected Shutdown

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD POWERED OFF UNEXPECTEDLY DURING TRANSPORT. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297682 ↗
  • MalfunctionSep 24, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT POWER ON WITH AC OR DC POWER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20291796 ↗
  • MalfunctionSep 24, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD UNEXPECTEDLY LOST POWER. AS A RESULT, DEFIBRILLATION WOULD NOT BE AVAILABLE, IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20290363 ↗
  • MalfunctionSep 24, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20290122 ↗
  • MalfunctionSep 6, 2024Failure to Power Up; Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. AS A RESULT, DEFIBRILLATION WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20156625 ↗
  • MalfunctionSep 6, 2024Fracture; Fracture

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE BATTERY PINS WERE PUSHED BACK INTO THE REAR CASE. IN THIS STATE, THE DEVICE MAY POWER OFF OR NOT POWER ON, WHICH MAY RESULT IN DEFIBRILLATION THERAPY BEING DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20154829 ↗
  • MalfunctionSep 5, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE UNEXPECTEDLY LOST POWER. N THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20143117 ↗
  • MalfunctionSep 4, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20133341 ↗
  • MalfunctionSep 3, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT BUTTONS ON THEIR DEVICE ARE UNRESPONSIVE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20123348 ↗
  • MalfunctionSep 3, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS NOT COMPLETING ITS BOOTUP CYCLE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20123185 ↗
  • MalfunctionAug 30, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON INSPECTION BY STRYKER IT WAS DISCOVERED THE DEVICE WOULD POWER ON BUT INTERMITTENTLY WOULD NOT COMPLETE THE BOOT UP CYCLE. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20104893 ↗
  • MalfunctionAug 27, 2024Failure to Power Up; Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. AS A RESULT, DEFIBRILLATION WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20076206 ↗
  • MalfunctionAug 22, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEY WERE UNABLE TO SELECT THE PADDLES LEAD. AS A RESULT, DEFIBRILLATION THERAPY MAY BE DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20037127 ↗
  • MalfunctionAug 22, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    DURING PREVENTIVE MAINTENANCE OF A DEVICE, STRYKER OBSERVED THAT THE DEVICE DID NOT COMPLETE THE BOOT-UP CYCLE. AS A RESULT, DEFIBRILLATION WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20035576 ↗
  • MalfunctionAug 21, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20015426 ↗
  • MalfunctionAug 20, 2024Device Sensing Problem; Device Sensing Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE THERAPY CONNECTION ON THEIR DEVICE WAS NOT WORKING. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20000708 ↗
  • MalfunctionAug 16, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE IS INTERMITTENTLY POWERING OFF UNEXPECTEDLY. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19998359 ↗
  • MalfunctionAug 13, 2024Mechanical Problem; Mechanical Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S BATTERY PIN WAS PUSHED IN. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19977940 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →