LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001256)Recalls & reported problems
Physio-Control, Inc. · 15AACBABBABBBAAAAAAAAAABLP15 MONITOR/DEFIB · Class 3
All models of LIFEPAK 15 monitor/defibrillator (93)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Physio-Control LIFEPAK 15 is a monitor and defibrillator that delivers shocks to restart the heart. Its 377 reports mostly describe sudden shutdowns, failures to power up, or failures to deliver a shock. The reports include no injuries and 3 deaths. There are 2 open recalls: one for missed inspections after servicing, and one for an out-of-tolerance tool.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 376
- Deaths reported
- 3
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 376 adverse event reports that list this exact device (DI 00883873911631), received between Mar 23, 2018 and Jul 30, 2026 — classified as 3 deaths, 373 malfunctions.
The most frequently reported specific device problem is “Unexpected Shutdown” (162 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 22, 2025Class IIOpen, Classified
Due to required inspections not being performed on products/units that have gone through servicing.
Z-2497-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Aug 30, 2024Class IIOpen, Classified
Due to an out of tolerance tool being used on monitor/defibrillator systems.
Z-0230-2025 · Root cause: Vendor change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Unexpected Shutdown162
- Electrical /Electronic Property Problem48
- Failure to Power Up33
- Failure to Deliver Energy26
- Failure to Deliver Shock/Stimulation23
- Inappropriate or Unexpected Reset16
- Key or Button Unresponsive/not Working14
- Device Sensing Problem12
Patient problems
- No Clinical Signs, Symptoms or Conditions204
- No Patient Involvement97
- No Known Impact Or Consequence To Patient74
- Cardiac Arrest3
- No Consequences Or Impact To Patient3
- Death2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 25 most recent of 376 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2026Failure to Charge
The customer contacted Stryker to report a non-critical issue with their device. Upon evaluation of the customer¿s device, Stryker observed that the device was unable to charge. In this state, defibrillation therapy may be delayed or prevented from being delivered to the patient if needed. There was no patient involvement reported with the event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50080387 ↗MalfunctionJul 29, 2026Unexpected Shutdown
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50066241 ↗MalfunctionJul 20, 2026Unexpected Shutdown
A CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON INSPECTION, IT WAS OBSERVED BY THE SERVICE TECHNICIAN THAT THE DEVICE EXPERIENCED UNEXPECTED SHUTDOWN. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25879543 ↗MalfunctionMay 28, 2026Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE HAS A POTENTIAL THERAPY BOARD FAILURE. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25312722 ↗MalfunctionJan 21, 2026Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UPEXPECTEDLY POWERED OFF. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24134124 ↗MalfunctionJan 20, 2026Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE UNEXPECTEDLY LOST POWER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24126716 ↗MalfunctionJan 14, 2026Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24066939 ↗MalfunctionJan 14, 2026Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD UNEXPECTEDLY LOST POWER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24066789 ↗MalfunctionJan 5, 2026Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE EXPERIENCED AN UNEXPECTED LOSS OF POWER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23961107 ↗MalfunctionNov 26, 2025Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE RANDOMLY TURNS OFF. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23655670 ↗MalfunctionNov 4, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S MAIN KEYPAD WAS PEELING OFF. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF IT WAS NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23467938 ↗MalfunctionSep 2, 2025Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT STAY POWERED ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22931664 ↗MalfunctionAug 28, 2025Mechanical Problem; Mechanical Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION STRYKER OBSERVED OBSERVED THAT THE THERAPY TRANSFER RELAY PLUG P24 CABLE TO BE PINCHED IN FRONT CASE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22918978 ↗MalfunctionAug 26, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD NO DISPLAY AND LEDS WERE ILLUMINATED DURING INTERVENTION ON A PATIENT. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22899599 ↗MalfunctionAug 26, 2025Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22899597 ↗MalfunctionAug 20, 2025Failure to Deliver Energy; Inappropriate/Inadequate Shock/Stimulation; Failure to Deliver Energy; Inappropriate/Inadequate Shock/Stimulation
STRYKER WAS PERFORMING PREVENTATIVE MAINTENANCE ON A STRYKER-OWNED LOANER DEVICE AND OBSERVED THAT THE DEVICE DOES NOT DELIVER DEFIBRILLATION ENERGY WHEN SHOCKING. IT ALSO DISPLAYS AN "ABNORMAL ENERGY DELIVERY" MESSAGE. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22847390 ↗MalfunctionAug 5, 2025Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE UNEXPECTEDLY SHUT DOWN. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22718744 ↗MalfunctionJul 8, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S MAIN KEYPAD WAS NO LONGER RESPONDING. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22435762 ↗MalfunctionJul 2, 2025Device Sensing Problem; Device Sensing Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DID NOT RECOGNIZE THE CONNECTED DEFIBRILLATION ELECTRODES. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22387664 ↗MalfunctionApr 1, 2025Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS PERIODICALLY SHUTTING OFF DURING MONITORING. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21744361 ↗MalfunctionMar 20, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT WHEN ATTEMPTING TO POWER ON THEIR DEVICE IT WOULD NOT COMPLETE A BOOT CYCLE. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21652611 ↗MalfunctionMar 20, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT SOME OF THE BUTTONS ON THE RIGHT SIDE (MAIN KEYPAD) OF THEIR DEVICE WERE NOT FUNCTIONING. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21652235 ↗MalfunctionMar 11, 2025Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS SHUTTING OFF UNEXPECTEDLY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21574559 ↗MalfunctionFeb 28, 2025Failure to Charge; Failure to Charge
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD AN ISSUE WITH CHARGING FOR DEFIBRILLATION. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21491509 ↗MalfunctionJan 28, 2025Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21257429 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page