Brilliance iCT Brilliance iCT SPRecalls & reported problems
Philips Medical Systems Nederland B.V. · Class 2
All models of Brilliance iCT (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884838059481), received between May 8, 2018 and May 10, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Defective Component” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K250648); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 1, 2021Class IIOpen, Classified
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Z-1990-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Defective Component1
- No Audible Alarm1
- Unintended System Motion1
Patient problems
- Bone Fracture(s)1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 10, 2023Unintended System Motion; Unintended System Motion
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THE CUSTOMER REPORTED THAT WHILE LOWERING THE CT COUCH, THE COUCH DESCENDED RAPIDLY AND WHEN IT REACHED THE LOWEST POINT THERE WAS A LOUD NOISE. THE SYSTEM WAS IN CLINICAL USE WITH PATIENT ON THE TABLE, HOWEVER THE PATIENT WAS NOT INJURED. THE FIELD SERVICE ENGINEER (FSE) EVALUATED THE SYSTEM AND DETERMINED THE BALL SCREW HAD FAILED. HE REPLACED THE BALL BEARING AND BALL SCREW AT THE SITE. THERE IS POTENTIAL OF SERIOUS INJURY IF THIS ISSUE WERE TO RECUR, THEREFORE THIS EVENT IS CONSIDERED REPORTABLE. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16905771 ↗InjuryNov 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THE ISSUE REPORTED WAS THAT DURING ROUTINE PREVENTIVE MAINTENANCE (PM) WORK ON A BRILLIANCE ICT, THE PHILIPS FIELD SERVICE ENGINEER (FSE) SUSTAINED A COMPOUND FRACTURE OF THE LEFT PROXIMAL INDEX FINGER WHICH REQUIRED SURGERY. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. THIS EVENT IS CURRENTLY UNDER INVESTIGATION.
Patient problems: Bone Fracture(s)
MAUDE report 12851856 ↗MalfunctionMay 8, 2018No Audible Alarm; Defective Component
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. THE CUSTOMER REPORTED THE INTERCOM SYSTEM WAS FAULTY. A PHILIPS FIELD SERVICE ENGINEER (FSE) REPORTED THE OPERATOR WAS NOT ABLE TO HEAR THE PATIENT AT ALL TIMES. IF THE OPERATOR IS UNABLE TO HEAR THE PATIENT DUE TO A FAILED COMPONENT, THERE IS POTENTIAL FOR INJURY TO THE PATIENT. THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. THIS EVENT IS CURRENTLY UNDER INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7495055 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K250648. One submission often covers many models, so check the product and lot details in each record.
- Z-3040-2026 · Aug 7, 2026 · Open, ClassifiedPhilips Brilliance. Models: Brilliance iCT, Brilliance iCT Upgrades, Brilliance iCT SP. Software 4.1.10.x Product Codes: 728306, 728305, 728311.
- Z-2650-2026 · Jun 11, 2026 · Open, ClassifiedBrilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 Upgrades
- Z-1315-2026 · Jan 7, 2026 · Open, ClassifiedBrilliance iCT; Product Code (REF): 728306;
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls6 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page