Xper Flex Cardio Physiomonitoring System FC 2020 (REF 453564243601)Recalls & reported problems
Philips Ds North America LLC · Class 2
All models of Xper Flex Cardio Physiomonitoring System (19)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884838083516), received between May 31, 2019 and Mar 31, 2021 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Mar 29, 2021Class IIOpen, Classified
Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is connected o the Flex Cardio device; display of certain data from the FC2010 becomes frozen, i.e., waveforms cease sweeping and updating and the ECG, IBP, and respiration numeric values cease to update;e ECG, IBP, and respiration waveforms become flat lines and no audible alarms are produced for HR and IBP, upon start up, an unexpected non physiological ECG waveform, erratic heart rate numeric value, and non-physiological display of any other active waveforms may appear on the Boom Monitor
Z-1517-2021 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Alarm System3
- Incorrect, Inadequate or Imprecise Result or Readings2
- Operating System Becomes Nonfunctional1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 31, 2021Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
THE CUSTOMER REPORTED THAT THE VITALS WERE FREEZING DURING CASES. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11597388 ↗MalfunctionNov 29, 2020Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT THE NIBP (NON-INVASIVE BLOOD PRESSURE) WILL GIVE THE SAME READING UNTIL MANUALLY GETTING ONE. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10916214 ↗MalfunctionMar 20, 2020Device Alarm System
THE CUSTOMER REPORTED ALARM ISSUE ON ALL BEDSIDES. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9860647 ↗MalfunctionMar 19, 2020Device Alarm System
THE CUSTOMER REPORTED ALARM ISSUE ON ALL BEDSIDES. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9856239 ↗MalfunctionMar 19, 2020Device Alarm System
THE CUSTOMER REPORTED ALARM ISSUE ON ALL BED SIDES. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9856129 ↗MalfunctionMay 31, 2019Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED ONGOING NIBP (NON-INVASIVE BLOOD PRESSURE) ISSUES, WHERE THE DEVICE WOULD GIVE THE SAME READINGS AT LEAST TWO TIMES IN A ROW. THERE WAS NO REPORTED PATIENT IMPACT/INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8658584 ↗
FDA clearance and classification
The FDA regulates product code MWI (monitor,physiological,patient(without arrhythmia detection or alarms)) under 21 CFR 870.2300 as a Class II device. 21 CFR 870.2300 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Xper Flex Cardio Physiomonitoring System FC 2020
Has the Xper Flex Cardio Physiomonitoring System FC 2020 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1517-2021 (Class II), began on Mar 29, 2021: Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Xper Flex Cardio Physiomonitoring System FC 2020?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between May 31, 2019 and Mar 31, 2021. In those reports, the most common specific device problems are “Device Alarm System” (3), “Incorrect, Inadequate or Imprecise Result or Readings” (2) and “Operating System Becomes Nonfunctional” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Xper Flex Cardio Physiomonitoring System FC 2020?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Xper Flex Cardio Physiomonitoring System FC 2020 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Xper Flex Cardio Physiomonitoring System FC 2020, and is it still sold?
The FDA's device database lists Philips Ds North America LLC as the company responsible for it (brand Xper Flex Cardio Physiomonitoring System, model FC 2020). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices
- Rad-97 9738Masimo Corporation · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)1 death222 reports
- Xper Flex Cardio Physiomonitoring System Xper Flex Cardio FC2010 REV CPhilips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls1 death42 reports
- Xper Flex Cardio Physiomonitoring System FC 2010Philips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 deaths38 reports
- Welch Allyn, Inc. 901058Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls5 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls3 reports
- McKesson Cardiology Hemo 14.0Change Healthcare Canada Company · Monitor,Physiological,Patient(Without Arrhythmia Detection Or Alarms)2 recalls2 deaths2 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 reports
- Change Healthcare Cardiology Hemo 14.2Change Healthcare Canada Company · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)2 recalls1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page