Xper Flex Cardio Physiomonitoring System FC 2010 (REF 453564483321)Recalls & reported problems
Philips Ds North America LLC · Class 2
All models of Xper Flex Cardio Physiomonitoring System (19)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00884838083523), received between Nov 19, 2018 and Mar 31, 2025 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (5 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Mar 14, 2018Class IITerminated
The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue.
Z-1867-2018 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings5
- Application Program Freezes or Fails to Launch4
- Therapeutic or Diagnostic Output Failure2
- Appropriate Device Problem Term/Code Not Available1
- Date/Time-Related Software Problem1
- Defective Device1
- Electrical /Electronic Property Problem1
- Imprecise or Erratic Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient6
- No Known Impact Or Consequence To Patient4
- Insufficient Information1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 31, 2025Imprecise or Erratic Results; Imprecise or Erratic Results
PHILIPS RECEIVED A COMPLAINT ON THE XPER FLEX CARDIO 2010 REV-B-EXCHANGE INDICATING THAT THE ECG DOES NOT HAVE A SMOOTH CURVE AND IS PIXELATED AND SQUARED OFF. THE DEVICE WAS IN USE ON PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21730474 ↗MalfunctionApr 12, 2024Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT WAS REPORTED THAT THE GROUND FLOOR ELECTROCARDIOGRAM (ECG) SIGNAL WAS FROZEN DURING THE EXAMINATION. HOWEVER, THE PRESSURE CONTINUED TO RUN CORRECTLY. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19096252 ↗MalfunctionMay 22, 2023Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT WAS REPORTED THAT THE ELECTROCARDIOGRAM (ECG) IMAGE ON THE FLEXVISION MONITOR FREEZES. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16981586 ↗MalfunctionDec 15, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
THE CUSTOMER REPORTED THAT VITALS FROM THE FC2010 DEVICE ARE FREEZING UNTIL REBOOT. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13010489 ↗MalfunctionOct 18, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT THE ECG WAVEFORM IS STUCK. THIS HAS BEEN INTERPRETED TO MEAN THAT THE ECG WAVEFORM IS FROZEN, STALE OR NOT UPDATING AS EXPECTED. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12650875 ↗MalfunctionOct 13, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER INITIALLY REPORTED THAT THE MACHINE IS HANGING AND THE ECG LEADS ARE NOT FUNCTIONING PROPERLY. ONSITE SUPPORT WAS DISPATCHED TO FURTHER EVALUATE THE DEVICE. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12626531 ↗MalfunctionApr 22, 2021Electrical /Electronic Property Problem; Date/Time-Related Software Problem
THE CUSTOMER REPORTED THAT THE HOST AND FLEX CARDIO BOOM SCREENS HAVE DIFFERENT TIMES AND ALSO ARE NOT ALWAYS IN SYNC. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11711735 ↗MalfunctionApr 22, 2021Therapeutic or Diagnostic Output Failure
THE CUSTOMER REPORTED A PROBLEM WITH THE XIM FLEX SWEEP SPEED. THE DEVICE WAS IN USE MONITORING A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT INVOLVING A PATIENT WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 11711435 ↗MalfunctionApr 2, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
THE CUSTOMER REPORTED THAT THE WAVEFORMS ON THE BOOM MONITOR FREEZE. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11609551 ↗MalfunctionAug 24, 2020Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER STATED THAT THERE IS A PROBLEM WITH MEASUREMENT OF ECG OR BLOOD PRESSURE INTERMITTENTLY. THE SCANNING SPEED (SWEEP SPEED) AS WELL AS THE HEART RATE CALCULATED AS A FUNCTION OF THE ECG WERE INCORRECT. THERE WAS NO REPORTED PATIENT IMPACT / INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10442019 ↗MalfunctionAug 14, 2020Appropriate Device Problem Term/Code Not Available
THE CUSTOMER REPORTED "SWEEP SPEED CHANGES". THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10408514 ↗MalfunctionAug 14, 2020Intermittent Continuity
THE CUSTOMER REPORTED THAT THE LIVE PORTION OF THE SYSTEM PAUSES FOR A SECOND, THEN THE SWEEP SPEED CHANGES FROM 25 TO 100. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10408418 ↗MalfunctionAug 14, 2020Therapeutic or Diagnostic Output Failure
THE CUSTOMER REPORTED THAT THE LIVE PORTION OF THE SYSTEM PAUSES FOR A SECOND, THEN THE SWEEP SPEED CHANGES FROM 25 TO 100. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10408365 ↗MalfunctionJul 24, 2020Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYS 2 DIFFERENT HEART FREQUENCIES, AND THERE IS A DIFFERENCE BETWEEN THE LIVE IMAGE AND PLAYBACK WORKSTATION. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10321397 ↗MalfunctionJul 1, 2020Operating System Becomes Nonfunctional
THE CUSTOMER REPORTED THAT ONE SINGLE BASELINE IS STUCK ON THE SYSTEM, AND IS NOT GOOD QUALITY. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10217224 ↗MalfunctionNov 19, 2019Defective Device
THE CUSTOMER REPORTED THAT THE TIME IS OFF BY 5 MINUTES WHEN SYNCHING. IT WAS REPORTED BY THE CUSTOMER THAT, THEY ARE NOTICING THE WAVES ARE NOT SYNCHING UP FROM THE BOOM TO THE CONTROL ROOM, AND IT APPEARS TO BE ABOUT 5 MINUTES OF A DIFFERENCE. THERE WAS NO REPORTED PATIENT IMPACT / INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9344928 ↗MalfunctionJul 25, 2019Output Problem
THE CUSTOMER REPORTED THAT, "THE POLIGRAFO CONTINUES TO SEND A DIFFERENT TRACK FROM THE MONITORED TO THE ROOM SCREEN RESPECTING TO WHAT APPEARS ON THE MONITORS IN THE AREA OF THE OUTDOOR SHIELDING ROOM". THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME THE REPORTED ISSUE WAS DISCOVERED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8826426 ↗MalfunctionJul 21, 2019Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED INCORRECT TRACK READINGS. IT WAS REPORTED BY THE CUSTOMER THAT, THE ECG SIGNALS ARE REGULAR, WHILE ON THE MONITOR OF THE POLYGRAPH IN THE CONTROL ROOM, THEY ARE IRREGULAR WITH BREAKS NOT CORRESPONDING TO REALITY (DROPPED QRS). THE DEVICE WAS IN CLINICAL USE.THERE WAS NO REPORTED PATIENT IMPACT / INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8810036 ↗MalfunctionNov 19, 2018Unable to Obtain Readings
THE CUSTOMER REPORTED THE SPO2 IS NOT GIVING A READING, THE SPO2 REMAINS AT 0%. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE REPORTED ISSUE WAS DISCOVERED. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED BY THE CUSTOMER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8085331 ↗
FDA clearance and classification
The FDA regulates product code MWI (monitor,physiological,patient(without arrhythmia detection or alarms)) under 21 CFR 870.2300 as a Class II device. 21 CFR 870.2300 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Xper Flex Cardio Physiomonitoring System FC 2010
Has the Xper Flex Cardio Physiomonitoring System FC 2010 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-1867-2018 (Class II), began on Mar 14, 2018: The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Xper Flex Cardio Physiomonitoring System FC 2010?
The FDA's MAUDE database lists 19 adverse event reports for this device, received between Nov 19, 2018 and Mar 31, 2025. In those reports, the most common specific device problems are “Incorrect, Inadequate or Imprecise Result or Readings” (5), “Application Program Freezes or Fails to Launch” (4) and “Therapeutic or Diagnostic Output Failure” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Xper Flex Cardio Physiomonitoring System FC 2010?
None of the reports are classified as deaths, 0 as injuries and 19 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Xper Flex Cardio Physiomonitoring System FC 2010 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Xper Flex Cardio Physiomonitoring System FC 2010, and is it still sold?
The FDA's device database lists Philips Ds North America LLC as the company responsible for it (brand Xper Flex Cardio Physiomonitoring System, model FC 2010). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices
- Rad-97 9738Masimo Corporation · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)1 death222 reports
- Xper Flex Cardio Physiomonitoring System Xper Flex Cardio FC2010 REV CPhilips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls1 death42 reports
- Xper Flex Cardio Physiomonitoring System FC 2010Philips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 deaths38 reports
- Welch Allyn, Inc. 901058Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls5 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls3 reports
- McKesson Cardiology Hemo 14.0Change Healthcare Canada Company · Monitor,Physiological,Patient(Without Arrhythmia Detection Or Alarms)2 recalls2 deaths2 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 reports
- Change Healthcare Cardiology Hemo 14.2Change Healthcare Canada Company · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)2 recalls1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page