The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00884838083530), received between Jul 10, 2019 and Aug 9, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Mar 14, 2018Terminated
The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue.
Z-1867-2018 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information1
- Unable to Obtain Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 9, 2021Unable to Obtain Readings; Unable to Obtain Readings
THE CUSTOMER REPORTED THAT DURING USE, THE SYSTEM WILL PAUSE, AND THE DISPLAY WILL FREEZE MOMENTARILY, THEN WILL RUN FAST TO CATCH UP TO REAL TIME, MAKING THE WAVEFORM LOOK LIKE ASYSTOLE (FLATLINE). ACCORDING TO THE CUSTOMER, EACH TIME THAT IT HAPPENED, THEY WOULD REBOOT THE DEVICE AND THE WORKSTATION IN ORDER TO GET THE FLEX 2020 TO FUNCTION CORRECTLY. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE REPORTED ISSUE WAS DISCOVERED. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12296104 ↗MalfunctionJul 10, 2019Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT ROOM 3 (IP (B)(4)) WAS POSTING THE SAME NIBP (NON-INVASIVE BLOOD PRESSURE) ON THE SEDATION FLOW SHEET. THERE WAS NO REPORTED PATIENT IMPACT / INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8776306 ↗
Other MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices
- Rad-97 9738Masimo Corporation · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)1 death222 reports
- Xper Flex Cardio Physiomonitoring System Xper Flex Cardio FC2010 REV CPhilips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls1 death42 reports
- Xper Flex Cardio Physiomonitoring System FC 2010Philips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 deaths38 reports
- Welch Allyn, Inc. 901058Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls5 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls3 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 reports
- McKesson Cardiology Hemo 14.0Change Healthcare Canada Company · Monitor,Physiological,Patient(Without Arrhythmia Detection Or Alarms)2 recalls2 deaths2 reports
- Change Healthcare Cardiology Hemo 14.2Change Healthcare Canada Company · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)2 recalls1 death1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.