Skip to content

Xper Flex Cardio Physiomonitoring System FC 2010 (REF 453564621791)Recalls & reported problems

Philips Ds North America LLC · XPER FLEX CARDIO REV-C-EXCHANGE · Class 2

Quick summary

This Philips Xper Flex Cardio monitor is a patient physiological monitor used during cardiac procedures. The 50 reports mostly describe inaccurate or missing readings, software that freezes or fails to start, ECG sweep speed mismatches between screens, and alarms that stay silent. No deaths were reported; 3 injuries were. One recall, started in March 2018 and since closed, involved inaccurate ventricular pressure readings on the live display.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 00884838084926), received between Jan 20, 2019 and Oct 21, 2024 — classified as 3 injuries, 47 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (10 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 14, 2018Terminated

    The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue.

    Z-1867-2018 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings10
  • Computer Software Problem8
  • Incorrect Measurement8
  • Application Program Freezes or Fails to Launch6
  • No Audible Alarm4
  • Insufficient Device Problem Information2
  • Unable to Obtain Readings2
  • Application Network Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions45
  • Insufficient Information4
  • No Known Impact Or Consequence To Patient3
  • Arrhythmia1

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2024No Audible Alarm; No Audible Alarm

    THE CUSTOMER REPORTED THAT THE FC2010 WAS NOT ALARMING. THE VIDEO OUTPUT IS SHOWING ALARMS, BUT NO SOUND IS COMING OUT. THE DEVICE WAS IN USE ON PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20494841 ↗
  • MalfunctionOct 10, 2024No Audible Alarm; No Audible Alarm

    PHILIPS RECEIVED A COMPLAINT ON THE XPER FLEX CARDIO 2010 REV-C-EXCHANGE INDICATING THERE WAS NO AUDIO. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20419306 ↗
  • MalfunctionOct 3, 2023Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT THE SWEEP SPEED IS CHANGING INTERMITTENTLY AND THERE IS A DIFFERENCE BETWEEN THE 2 MONITORS. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17859949 ↗
  • MalfunctionJul 25, 2023Incorrect Measurement; Incorrect Measurement

    THE CUSTOMER REPORTED THAT THE SWEEP SPEED IS VERY FAST. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17390216 ↗
  • MalfunctionJul 6, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    PHILIPS RECEIVED A COMPLAINT ON THE XPER FLEX CARDIO 2010 REV-C-EXCHANGE INDICATING THAT THE SWEEP SPEED ON THE BOOM MONITOR WAS DISPLAYING FASTER THAN ON THE HOST. THE CUSTOMER STATES THAT A DEVICE RESTART RESOLVES THE ISSUE. THERE WAS NO REPORT OF A DEATH OR SERIOUS INJURY, NO REPORT OF MISINTERPRETATION OR MISDIAGNOSIS, NOR WAS THERE A REPORT OF ANY ADVERSE IMPACT TO ANY USER OR PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17269036 ↗
  • MalfunctionNov 10, 2022Thermal Decomposition of Device; Sparking

    IT WAS REPORTED THAT THE POWER SUPPLY HAD AN ELECTRICAL BURN AND PRODUCED SPARKS BETWEEN THE POWER SUPPLY AND POWER CABLE. WHEN PHILIPS WENT ONSITE, FOUND THAT THE CUSTOMER HAD ALREADY ISOLATED THE CABLE AT BOTH ENDS AND REMOVED THE POWER SUPPLY FROM THE THEATRE. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME THE ISSUE OCCURRED. THERE AS NO ADVERSE EVENT OR USER/PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15775446 ↗
  • MalfunctionOct 11, 2022No Audible Prompt/Feedback; No Audible Prompt/Feedback

    THE CUSTOMER REPORTED THAT ARTIFACTS ARE APPEARING IN ECG AND AUDIO ALARM IS NOT COMING. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED. A PHILIPS FIELD SERVICE ENGINEER (FSE) WENT TO THE CUSTOMER SITE. THE FSE EVALUATED THE DEVICE AND DETERMINED THAT THE AUDIO ERROR WAS NOT PRESENT AT THE TIME OF INSPECTION, AND THAT THE AUDIO VOLUME WAS VERY LOW. THE FSE RESEATED THE DEVICE SPEAKERS AND CONFIRMED THAT THE DEVICE AUDIO IS NOW FUNCTIONING AS EXPECTED. NO FURTHER INVESTIGATION OR ACTION IS WARRANTED AT THIS TIME.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15578090 ↗
  • MalfunctionSep 12, 2022Incorrect Measurement; Incorrect Measurement

    THE CUSTOMER REPORTED HAVING INTERMITTENT MONITORING ISSUES IN ALL ROOMS, INTERMITTENT OFF OR ECG NOT MATCHING THE BOOM. IT WAS CONFIRMED THERE AS NO ISSUE WITH THE MONITORING OFF, THE ISSUE WAS THE ECG SWEEP SPEED. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15406322 ↗
  • MalfunctionSep 7, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT IT LOOKS LIKE THE SWEEP SPEED CHANGES ON THE MONITOR. THE DEVICE WAS IN USE AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15375439 ↗
  • MalfunctionSep 7, 2022Incorrect Measurement; Incorrect Measurement

    THE CUSTOMER REPORTED HAVING INTERMITTENT MONITORING ISSUES IN ALL ROOMS. INTERMITTENT OFF OR ECG NOT MATCHING THE BOOM. THE DEVICE WAS IN USE. THERE WAS NO REPORTED PATIENT IMPACT / INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15371198 ↗
  • MalfunctionAug 9, 2022Incorrect Measurement; Incorrect Measurement

    THE CUSTOMER REPORTED SLOWNESS IN THE DISPLAY OF THE PLOTS (SWEEP SPEED). THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15191662 ↗
  • MalfunctionJun 10, 2022Incorrect Measurement; Incorrect Measurement

    THE CUSTOMER REPORTED THAT THE WAVE TRACING APPEARS TO BE STRETCHED WHEN DISPLAYING ON FLEXVISION. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14652624 ↗
  • MalfunctionApr 4, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    THE CUSTOMER REPORTED THAT THE LIVE SPEED (SWEEP SPEED) ON THE MONITOR IN THE EXAMINATION ROOM IS DIFFERENT FROM THAT IN THE CONTROL ROOM. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13994322 ↗
  • MalfunctionMar 24, 2022Defibrillation/Stimulation Problem; Insufficient Device Problem Information; Defibrillation/Stimulation Problem; Insufficient Device Problem Information

    THE CUSTOMER REPORTED THAT THE PATIENT WAS DECLINING DURING CASE AND WAS DEFIBRILLATED, WHICH IN TURN CAUSED THE EKG TO FLAT LINE AND PRESSURES TO DISAPPEAR. WE ARE IN THE PROCESS OF GATHERING ADDITIONAL INFORMATION REGARDING THIS INCIDENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13892896 ↗
  • InjuryMar 21, 2022Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT DURING A PULMONARY VEIN ISOLATION EXAMINATION OF A PATIENT UNDER GENERAL ANESTHESIA, THE PATIENT HAD TO BE REANIMATED IN THE PROCEDURE, AND THE HEMO MONITORING WAS FROZEN. ACCORDING TO THE CUSTOMER, THERE WAS NO LIVE ECG MONITORING IN EITHER THE EXAMINATION ROOM OR IN THE CONTROL ROOM. REBOOTING THE BOX FIXED THE PROBLEM.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13839941 ↗
  • MalfunctionJan 22, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THE DEVICE SHOWS THE NON-INVASIVE BLOOD PRESSURE READING FROM THE PREVIOUS PATIENT. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE REPORTED ISSUE WAS DISCOVERED. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13329168 ↗
  • MalfunctionDec 9, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    THE CUSTOMER REPORTED THAT HEMOS LOCKED UP. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE REPORTED ISSUE WAS DISCOVERED. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12968015 ↗
  • MalfunctionDec 8, 2021Output Problem; Application Program Freezes or Fails to Launch; Output Problem; Application Program Freezes or Fails to Launch

    THE CUSTOMER REPORTED THAT INTERMITTENTLY WAVEFORMS HANG. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12964367 ↗
  • MalfunctionDec 1, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    THE CUSTOMER REPORTED THAT HEMO'S FROZE. ADDITIONAL INFORMATION RECEIVED FROM REMOTE SUPPORT ENGINEER (RSE), WHO INDICATED THAT, THE HEMO SCREEN FROZE, WAVES STOPPED MOVING, AND VALUES STOPPED CHANGING. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12907854 ↗
  • MalfunctionNov 29, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THE TRACE SPEED (SWEEP SPEED) IS DIFFERENT FROM THE CONTROL ROOM TO THE EXAMINATION ROOM. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12883148 ↗
  • MalfunctionNov 22, 2021Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings; Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THE SWEEP SPEED ON THE BOOM APPEARS HIGHER THAN THE SWEEP SPEED IN THE CONTROL ROOM. THE HEART RATE READING IS NOT ACCURATE, THE ECG IS GETTING DRAGGED ONLY INSIDE THE LAB. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12852481 ↗
  • MalfunctionNov 17, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THE SWEEP SPEED CHANGES ON ITS OWN ON THE BOOM. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12827650 ↗
  • MalfunctionOct 21, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    THE CUSTOMER REPORTED THAT THE IR LAB WAS SEEING PAUSES WITH ECG FOR ONE CASE PERFORMED ON (B)(6), 2021.THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12672673 ↗
  • MalfunctionOct 19, 2021No Audible Alarm; No Audible Alarm

    THE CUSTOMER REPORTED THAT NO SOUND AT ALL SEEMED TO BE COMING OUT OF THE DEVICE. IT IS UNKNOW IF THE DEVICE WAS IN USE AT TIME OF EVENT AND NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12660877 ↗
  • MalfunctionOct 7, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THE SWEEP SPEED GOES FAST THEN SLOW. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12590265 ↗

Other MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.