CAPSURE® SENSE 457453Recalls & reported problems
Medtronic, Inc. · LEAD 457453 CAPSURE SENSE PACING US · Class 3 · Implant
All models of CAPSURE SENSE (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Capsure Sense lead has 196 reports, 79 injuries and 2 deaths. Many reports describe the lead sensing heartbeats poorly or needing higher electrical output to pace, and some describe infections that led to the system being replaced. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 198
- Deaths reported
- 2
- Injuries reported
- 79
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 198 adverse event reports that list this exact device (DI 00885074200958), received between Apr 6, 2017 and Sep 23, 2026 — classified as 2 deaths, 79 injuries, 117 malfunctions.
The most frequently reported specific device problem is “Under-Sensing” (91 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Under-Sensing91
- Adverse Event Without Identified Device or Use Problem32
- High Capture Threshold30
- High impedance19
- Over-Sensing13
- Pacing Problem13
- Failure to Capture12
- Break5
Patient problems
- No Clinical Signs, Symptoms or Conditions82
- No Known Impact Or Consequence To Patient64
- Unspecified Infection23
- No Consequences Or Impact To Patient7
- Bacteremia3
- Bradycardia3
- Dyspnea3
- Erosion3
Recent adverse event reports
The 25 most recent of 198 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Pacing Problem; Under-Sensing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50871297 ↗MalfunctionSep 18, 2026Under-Sensing
It was reported that the Right Ventricular (RV) lead exhibited undersensing during detected arrhythmia episodes and suspected during detected ventricular sensing episodes. It was also reported that the Right Atrial (RA) lead exhibited undersensing. The leads remain in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50821007 ↗InjuryAug 19, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the pacing system was explanted and replaced due to infection. No further patient complications have been reported as a result of this event.
Patient problems: Unspecified Infection
MAUDE report 50324191 ↗MalfunctionMay 22, 2026High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED RISING CAPTURE THRESHOLDS AND DIMINISHED R WAVE SENSING. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH CAPTURE THRESHOLDS. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25255434 ↗MalfunctionMay 15, 2026Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING ON THE CURRENT ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) ELECTROGRAM (EGM), RESULTING IN TERMINATION OF THE AT/AF DETECTED EVENT IN PROGRESS AND UNNECESSARY PACING AT TIMES. THE LAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25185381 ↗MalfunctionMay 15, 2026Pacing Problem; Under-Sensing; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLA TION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENTS IN PROGRESS AND CAUSING INAPPROPRIATE ATRIAL PACING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25182693 ↗MalfunctionApr 13, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24855674 ↗MalfunctionMar 31, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING ON A STORED EPISODE, RESULTING IN TERMINATION OF ATRIAL TACHY CARDIA/ATRIAL FIBRILLATION (AT/AF) DETECTED EVENTS IN PROGRESS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24746209 ↗MalfunctionMar 25, 2026High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24685786 ↗InjuryMar 17, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED A POSSIBLE DEVICE POCKET INFECTION. THE PATIENT EXPERIENCED SWELLING AND HAD WOUND DEHISCENCE. THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS PACEMAKER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Medical device site infection; Swelling/ Edema; Wound Dehiscence
MAUDE report 24619935 ↗MalfunctionMar 13, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. LOW P WAVES WERE NOTED. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24588504 ↗MalfunctionFeb 2, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA (AT) / ATRIAL FIBRILLATION (AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24233888 ↗InjuryJan 23, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED DUE TO INFECTION. VEGETATION WAS SUSPECTED. THE PATIENT EXPERIENCED DEEP VEIN THROMBOSIS (DVT) AND MEDICATION WAS REQUIRED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Medical device site infection; Thrombosis/Thrombus; Unspecified Infection
MAUDE report 24169059 ↗MalfunctionJan 5, 2026Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA (AT) / ATRIAL FIBRILLATION (AF) EVENTS RESULTING IN UNNECESSARY PACING AT TIMES AND EARLY EPISODE TERMINATION. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23967078 ↗InjuryDec 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BACTEREMIA AND EROSION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS REMOVED AND REPLACED WITH A LEADLESS IPG. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia; Erosion
MAUDE report 23906072 ↗InjuryNov 24, 2025High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICAT IONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23632714 ↗MalfunctionOct 31, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACH YCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23433843 ↗MalfunctionAug 18, 2025High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS). THE RA LEAD EXHIBITED A HIGH THRESHOLD. THE RV AND RA LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22833702 ↗MalfunctionAug 18, 2025Signal Artifact/Noise; Over-Sensing; Signal Artifact/Noise; Over-Sensing
IT WAS FURTHER REPORTED THAT THE PATIENT WAS HAVING SURGERY AT THE TIME AND THE NOISE WAS DETERMINED TO BE CAUTERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22832567 ↗MalfunctionJul 28, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22631826 ↗InjuryJul 16, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT THERE WAS "SOMETHING WRONG" WITH THEIR RIGHT ATRIAL (RA) LEAD. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22518520 ↗InjuryJun 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND LEFT VENTRICULAR (LV) LEAD WERE EXPLANTED. THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD WERE PARTIALLY EXPLANTED AS THEY WERE CUT AT THE PROXIMAL END DUE TO TEARS IN THE SUPERIOR VENA CAVA (SVC) JUNCTION AND ABANDONING LOWER LEFT EXTRACTION. IT WAS REPORTED THAT PATIENT¿S CHEST WAS CRACKED FROM TWO TEARS NEAR THE SVC JUNCTION DURING LOWER LEFT EXTRACTION. IT WAS NOTED THAT IT WAS DUE TO THE EXTRACTION PROCESS AND NOT THE DEVICE AND LEADS. AN IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS IMPLANTED. THE PATIENT WAS STABLE IN THE INTENSIVE CARE UNIT (ICU). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection; Vascular Dissection
MAUDE report 22324270 ↗InjuryJun 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED WITHOUT REPLACEMENT DUE TO INFECTION. ADDITIONALLY, VEGETATION WAS OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Medical device site infection; Unspecified Infection
MAUDE report 22197916 ↗MalfunctionJun 3, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED POTENTIAL UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22129912 ↗InjuryApr 25, 2025Break; Break
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED WITHOUT REPLACEMENT DUE TO INFECTION. THE RIGHT VENTRICULAR (RV) LEAD AND THE RIGHT ATRIAL (RA) LEAD WERE PARTIALLY EXPLANTED AND CAPPED DUE TO THE SUBCLAVIAN VEIN BEING OCCLUDED AND THE POCKET WAS HEAVILY CALCIFIED AND THE LEADS BROKE ON EXTRACTION ATTEMPT. THE LEADS WERE CAPPED. THE ORGANISM WAS IDENTIFIED AS STAPHYLOCOCCUS AUREUS AND BACTEREMIA. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia; Bacterial Infection; Obstruction/Occlusion
MAUDE report 21907410 ↗
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page