PARADIGM® REAL-time MMT-722NAP (UDI 00885074419824)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-722NAP PRDGM INSULIN PL EN PR · Class 3
All models of PARADIGM REAL-time (95)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00885074419824), received between Jul 6, 2017 and Nov 13, 2019 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (3 reports).
14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message3
- Date/Time-Related Software Problem1
- Device Difficult to Program or Calibrate1
- Device Operates Differently Than Expected1
- Inappropriate or Unexpected Reset1
- Patient Device Interaction Problem1
- Power Problem1
Patient problems
- No Consequences Or Impact To Patient5
- Hyperglycemia1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 13, 2019Device Difficult to Program or Calibrate; Power Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Power Problem; Patient Device Interaction Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP BUTTONS WERE HARDER TO PRESS AND REJECTING THE NEW BATTERIES. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9313903 ↗MalfunctionOct 22, 2018Date/Time-Related Software Problem; Date/Time-Related Software Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD RECEIVED BUTTON ERROR AND TIME CLOCK ANOMALY. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 207 MG/DL. THE CUSTOMER REPORTED THAT THEY HAD BUTTON ERROR. THE CUSTOMER REPORTED THAT THE TIME CLOCK WAS NOT ADVANCING. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO BACKUP PLAN PER HEALTH CARE PROFESSIONAL INSTRUCTIONS. THE CUSTOMER WAS ADVISED THE PUMP WILL NEED TO BE REPLACED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7990943 ↗MalfunctionMay 1, 2018Device Displays Incorrect Message
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. CUSTOMER STATED THAT DID THE REWIND AGAIN AND GOT THE COMPROMISED FORCE SENSOR SYSTEM ERROR AGAIN. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7475035 ↗InjuryNov 1, 2017Device Operates Differently Than Expected
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY HAD A HIGH BLOOD GLUCOSE VALUE OF 522 MG/DL AND COMPROMISED FORCE SENSOR SYSTEM. THE INSULIN PUMP IS EXPECTED TO BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 6995249 ↗MalfunctionAug 3, 2017Device Displays Incorrect Message
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED MOTOR ERROR SEVERAL TIMES. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 127 MG/DL. THE DRIVE SUPPORT CAP WAS PROTRUDED. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACKUP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6765999 ↗MalfunctionJul 6, 2017Device Displays Incorrect Message; Inappropriate or Unexpected Reset
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THE ALARM. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS 146 MG/DL. THE CUSTOMER WAS ASKED TO OBSERVE IF THE TIME CLOCK ADVANCES AND IT DOES. CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP HAD AN UNEXPECTED RESTART SHORTLY AFTER INSERTING A NEW BATTERY. THE ALARM RECURRED DURING TROUBLESHOOTING. THE CUSTOMER WAS ADVISED TO MONITOR THE INSULIN PUMP FOR FURTHER ISSUES. THE PRODUCT WAS NOT RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6689166 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.
- Z-1750-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
- Z-1749-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Insulin Pump (MMT-522, MMT-722)
- Z-1256-2025 · Jan 31, 2025 · Open, ClassifiedParadigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
- Z-1587-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-522
- Z-1593-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723K
- Z-1592-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523K
- Z-1589-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-522K
- Z-1594-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723
- Z-1591-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523
- Z-1588-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722
- Z-1590-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722K
- Z-2442-2015 · Jul 22, 2015 · TerminatedGuardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
- Z-2179-2014 · Jul 21, 2014 · TerminatedMedtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Sensor, glucose, invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page