da Vinci Energy 378848Recalls & reported problems
Intuitive Surgical, Inc. · E-100 Generator · Class 2
All models of da Vinci Surgical System (9) · da Vinci Energy (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00886874116982), received between Apr 26, 2024 and Jun 25, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Energy Output Problem” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 22, 2022Class IICompleted
Due to preventative maintenance (PM) not being performed on E-100 generators which is a bipolar electrosugical unit designed to be used with the da Vinci Xi and X systems.
Z-1484-2022 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Energy Output Problem2
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 25, 2024Energy Output Problem; Energy Output Problem
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE E-100 GENERATOR DID NOT WORK. THE LED ON POWER BUTTON WAS ON, BUT IT WAS NOT WHILE THE INSTRUMENT WAS CONNECTED. THE CUSTOMER STATED THAT THEY USED THE ERBE GENERATOR WITH THE VESSEL SEALER INSTRUMENT WITHOUT ANY PROBLEM. AN INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SERVICE ENGINEER (TSE) VIEWED THE SYSTEM EVENT LOGS BUT DID NOT NOTICE ANYTHING RELEVANT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE E-100 WAS NOT WORKING AT ALL. IT WAS A NEW DA VINCI SYSTEM INSTALLED. THE CUSTOMER USED THE ERBE GENERATOR AND WERE ABLE TO USE THE VESSEL SEALER WITHOUT ANY PROBLEMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19610574 ↗MalfunctionMay 9, 2024Output Problem; Output Problem
IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE E-100 GENERATOR FAILED TO FIRE INTERMITTENTLY. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING INFORMATION: THE CUSTOMER WAS UNABLE TO PROVIDE ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19281326 ↗InjuryApr 26, 2024Energy Output Problem; Energy Output Problem
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED GASTRIC BYPASS (ROUX-EN-Y) SURGICAL PROCEDURE, THE CUSTOMER REPORTED AN ISSUE WITH THE VESSEL SEALER EXTEND (VSE). THE VSE FAILED AND RESULTED IN AN ABORT TO OPEN. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) REVIEWED THE LOGS BUT FOUND NO RELATED ISSUE. THE CUSTOMER HAD NO ADDITIONAL DETAILS ABOUT THE ISSUE AS FAR AS THE ARM IT WAS INSTALLED IN, WHETHER THE VSE WAS USED IN THE E100/IESU, OR HOW EXACTLY THE VSE FAILED BUT THE CUSTOMER HAS INDICATED SOMETHING WAS WRONG AND WAS NOT GOING TO PERFORM SURGERY WITHOUT THE SYSTEM BEING CHECKED. THE PROCEDURE WAS CONVERTED TO OPEN SURGERY.
Patient problems: Insufficient Information
MAUDE report 19191815 ↗
Other System, surgical, computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page