LUCEA LED® ARDLCA219000ARecalls & reported problems
Maquet Sas · LCA 100 · Class 2
All models of LUCEA LED (33)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03700712400613), received between Sep 4, 2023 and Jul 26, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Nov 8, 2023Class IIOpen, Classified
A potential for a light system to fall in the operating room.
Z-0759-2024 · Root cause: Equipment maintenance · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ON 25TH JULY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 100. IT WAS STATED THE CIRCULAR SPRING ARM COVERS WERE MISSING. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19839237 ↗MalfunctionDec 1, 2023Peeled/Delaminated; Detachment of Device or Device Component; Peeled/Delaminated; Detachment of Device or Device Component
ON 23RD NOVEMBER, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR SURGICAL LIGHTS - LUCEA 100. BASED ON VIDEO EVIDENCE THE CONNECTION BETWEEN SPRING ARM AND FORK WAS LOOSE WITH RISK OF DETACHMENT, ALSO THE PAINT WAS PEELING FROM THE FORK. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18244408 ↗MalfunctionSep 4, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ON 7TH AUGUST 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 100. BASED ON PHOTOGRAPHIC EVIDENCE THE CAP AT THE SIDE OF THE SPRING ARM WAS MISSING. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17675436 ↗
FDA clearance and classification
Cleared through Special 510(k) K113679 on Jan 12, 2012 (decision: Substantially Equivalent). Submitted by Maquet S.A.S. as “Lucea Led Surgical Light System”. The FDA review took 29 days. FDA record ↗
The FDA regulates product code FSY (light, surgical, ceiling mounted) under 21 CFR 878.4580 as a Class II device. 21 CFR 878.4580 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the LUCEA LED® ARDLCA219000A
Has the LUCEA LED® ARDLCA219000A been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0759-2024 (Class II), began on Nov 8, 2023: A potential for a light system to fall in the operating room. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the LUCEA LED® ARDLCA219000A?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Sep 4, 2023 and Jul 26, 2024. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (3) and “Peeled/Delaminated” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the LUCEA LED® ARDLCA219000A?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the LUCEA LED® ARDLCA219000A FDA approved?
It was cleared through Special 510(k) K113679 on Jan 12, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the LUCEA LED® ARDLCA219000A MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the LUCEA LED® ARDLCA219000A, and is it still sold?
The FDA's device database lists Maquet Sas as the company responsible for it (brand LUCEA LED®, model ARDLCA219000A). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Light, Surgical, Ceiling Mounted devices
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls6 reports
- POWERLED TM ARD568474114AMaquet Sas · Light, surgical, ceiling mounted2 recalls3 reports
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls2 reports
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls2 reports
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page