Skip to content

LUCEA LED® ARDLCA219001ARecalls & reported problems

Maquet Sas · LCA DUO 100 · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03700712400651), received between May 7, 2022 and Dec 13, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 8, 2023Open, Classified

    A potential for a light system to fall in the operating room.

    Z-0759-2024 · Root cause: Equipment maintenance · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component6
  • Crack4
  • Overheating of Device1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA LED®. IT WAS STATED AND CONFIRMED BY PHOTOGRAPHIC EVIDENCE THAT THE COVER WAS CRACKED RESULTING IN MISSING PARTICLES. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20923125 ↗
  • MalfunctionOct 14, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    ON 8TH OCTOBER, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 100. IT WAS STATED THE COVERS OF THE DOMES WERE BROKEN DUE TO MISUSE AND THERE WAS A RISK OF THEM OR THEIR PARTS FALLING OFF. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE HEADLIGHT COVERS WERE BROKEN IN SEVERAL PLACES AND SOME PARTS WERE MISSING. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20438040 ↗
  • MalfunctionMay 29, 2024Peeled/Delaminated; Detachment of Device or Device Component; Peeled/Delaminated; Detachment of Device or Device Component

    ON 23TH MAY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 100. IT WAS STATED THE ACRYLIC FROM THE DOME FELL. IT WAS DISCOVERED BEFORE THE SURGERY AND FIXED WITH ADHESIVE TAPE. ACCORDING TO THE GETINGE TECHNICIAN, THE EQUPIMENT IS NOT RELEASED FOR USE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19413545 ↗
  • MalfunctionNov 2, 2023Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR SURGICAL LIGHTS ¿ LUCEA 100/100. AS IT WAS STATED AND CONFIRMED WITH THE PHOTOGRAPHIC EVIDENCE, THE HEADLIGHT COVERS ON BOTH DOMES WERE CRACKED WITH MISSING PARTICLES. THE ISSUE WAS DISCOVERED DURING DAILY CHECK UP. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18055426 ↗
  • MalfunctionJun 21, 2023Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    MANUFACTURER'S REFERENCE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17171713 ↗
  • MalfunctionFeb 2, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    MANUFACTURER'S REFERENCE NUMBER (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16283003 ↗
  • MalfunctionMay 7, 2022Overheating of Device; Overheating of Device

    ON 28TH OF APRIL, 2022 GETINGE BECAME AWARE OF AN ISSUE WITH LUCEA 100 - SURGICAL LIGHT. AS IT WAS STATED, CONNECTION BETWEEN SPRING ARM AND FORK HAS OVERHEATED. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS OVERHEATING MAY CAUSE FIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14315221 ↗

Other Light, Surgical, Ceiling Mounted devices

All Light, Surgical, Ceiling Mounted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page