Skip to content

GIBECK IPN043768Recalls & reported problems

Teleflex Incorporated · HUMID-VENT FILTER COMPACT S · Class 2

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04026704348022), received between May 7, 2021 and Feb 20, 2025 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Disconnection” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Disconnection4
  • Blocked Connection1
  • Fluid/Blood Leak1
  • Free or Unrestricted Flow1
  • Gas/Air Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Capsular Contracture1
  • Insufficient Information1
  • Low Oxygen Saturation1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 20, 2025Blocked Connection; Blocked Connection

    IT WAS REPORTED THAT "WE HAD A BACTERIAL FILTER FAIL ON ONE OF OUR ANESTHETIC MACHINES CAUSING THE PATIENT TO DESATURATE AND DRUGS WERE GIVEN TO THE PATIENT TO NO EFFECT. ON FURTHER INVESTIGATION IT WAS DISCOVERED THAT PART OF THE PAPER FILTER WAS MOVING WITH THE PATIENT RESPIRATIONS. THE FILTER WAS NOT WET. THERE MAY ALSO HAVE BEEN A LEAK AT THE BOTTOM OF THE CO2 SAMPLING PORT. ONCE THE FILTER WAS REPLACED THE PATIENT'S SATS' RETURNED TO NORMAL LEVELS AND THE OPERATION WAS COMPLETED WITHOUT FURTHER ISSUES". THE PATIENT STATUS IS REPORTED AS "FINE". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION.

    Patient problems: Low Oxygen Saturation

    MAUDE report 21423458 ↗
  • MalfunctionMay 11, 2023Disconnection; Disconnection

    IT WAS REPORTED THAT PRIOR TO USE WHILE ASSEMBLING THE TUBING, THE FILTER DOES NOT STAY ATTACHED TO THE TUBING AND DISCONNECTS FROM THE WHISPER SWIVEL VALVE EASILY, EVEN IF THE TUBING IS NOT MOVED. ADDITIONAL INFORMATION RECEIVED STATES THAT THE FILTERS ARE COMPATIBLE WITH THE SWIVEL VALVE. NO PATIENT INVOLVEMENT.

    Patient problems: Capsular Contracture

    MAUDE report 16913347 ↗
  • MalfunctionMay 11, 2023Disconnection; Disconnection

    IT WAS REPORTED THAT PRIOR TO USE WHILE ASSEMBLING THE TUBING, THE FILTER DOES NOT STAY ATTACHED TO THE TUBING AND DISCONNECTS FROM THE WHISPER SWIVEL VALVE EASILY, EVEN IF THE TUBING IS NOT MOVED. ADDITIONAL INFORMATION RECEIVED STATES THAT THE FILTERS ARE COMPATIBLE WITH THE SWIVEL VALVE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16913299 ↗
  • MalfunctionMay 11, 2023Disconnection; Disconnection

    IT WAS REPORTED THAT PRIOR TO USE WHILE ASSEMBLING THE TUBING, THE FILTER DOES NOT STAY ATTACHED TO THE TUBING AND DISCONNECTS FROM THE WHISPER SWIVEL VALVE EASILY, EVEN IF THE TUBING IS NOT MOVED. ADDITIONAL INFORMATION RECEIVED STATES THAT THE FILTERS ARE COMPATIBLE WITH THE SWIVEL VALVE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16913179 ↗
  • MalfunctionMay 11, 2023Disconnection; Disconnection

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE FILTER DOES NOT STAY ATTACHED TO THE TUBING AND DISCONNECTS FROM THE WHISPER SWIVEL VALVE EASILY, EVEN IF THE TUBING IS NOT MOVED. THIS CAUSES A BREAK IN THE BREATHING CIRCUIT AND CUTS OFF THE DELIVERY OF OXYGEN TO A PATIENT WHO IS CONTINUOUSLY VENTILATOR DEPENDENT. ADDITIONAL INFORMATION RECEIVED STATES THAT THE FILTERS ARE COMPATIBLE WITH THE SWIVEL VALVE. THERE WAS NO PATIENT HARM, INJURY, OR DESATURATION AS THE ISSUE WAS DISCOVERED PRIOR TO USE WHEN ASSEMBLING THE TUBING. AS A RESULT, AN "ALTERNATIVE" FILTER WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16913003 ↗
  • MalfunctionApr 18, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THERE WAS "WATER LEAKING IN THE HME FILTER WHILE ATTACHED TO THE VENTILATOR AND PATIENT. OCCURED DURING VENTILATION TREATMENT. THE FILTER WAS REPLACED." AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. THE COMPLAINT FILE WILL BE UPDATED IF FURTHER INFORMATION IS RECEIVED.

    Patient problems: Insufficient Information

    MAUDE report 16765572 ↗
  • MalfunctionMay 7, 2021Free or Unrestricted Flow; Free or Unrestricted Flow

    IT WAS REPORTED THAT "DURING THE PRE-TEST, ALERT LEAKAGE". NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11794508 ↗
  • MalfunctionMay 7, 2021Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT "DURING THE PRE-TEST, ALERT LEAKAGE". NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11794353 ↗

Other Filter, bacterial, breathing-circuit devices

All Filter, bacterial, breathing-circuit devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.