GIBECK IPN043768Recalls & reported problems
Teleflex Incorporated · HUMID-VENT FILTER COMPACT S · Class 2
All models of GIBECK (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04026704348022), received between May 7, 2021 and Feb 20, 2025 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Disconnection” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Disconnection4
- Blocked Connection1
- Fluid/Blood Leak1
- Free or Unrestricted Flow1
- Gas/Air Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Capsular Contracture1
- Insufficient Information1
- Low Oxygen Saturation1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryFeb 20, 2025Blocked Connection; Blocked Connection
IT WAS REPORTED THAT "WE HAD A BACTERIAL FILTER FAIL ON ONE OF OUR ANESTHETIC MACHINES CAUSING THE PATIENT TO DESATURATE AND DRUGS WERE GIVEN TO THE PATIENT TO NO EFFECT. ON FURTHER INVESTIGATION IT WAS DISCOVERED THAT PART OF THE PAPER FILTER WAS MOVING WITH THE PATIENT RESPIRATIONS. THE FILTER WAS NOT WET. THERE MAY ALSO HAVE BEEN A LEAK AT THE BOTTOM OF THE CO2 SAMPLING PORT. ONCE THE FILTER WAS REPLACED THE PATIENT'S SATS' RETURNED TO NORMAL LEVELS AND THE OPERATION WAS COMPLETED WITHOUT FURTHER ISSUES". THE PATIENT STATUS IS REPORTED AS "FINE". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION.
Patient problems: Low Oxygen Saturation
MAUDE report 21423458 ↗MalfunctionMay 11, 2023Disconnection; Disconnection
IT WAS REPORTED THAT PRIOR TO USE WHILE ASSEMBLING THE TUBING, THE FILTER DOES NOT STAY ATTACHED TO THE TUBING AND DISCONNECTS FROM THE WHISPER SWIVEL VALVE EASILY, EVEN IF THE TUBING IS NOT MOVED. ADDITIONAL INFORMATION RECEIVED STATES THAT THE FILTERS ARE COMPATIBLE WITH THE SWIVEL VALVE. NO PATIENT INVOLVEMENT.
Patient problems: Capsular Contracture
MAUDE report 16913347 ↗MalfunctionMay 11, 2023Disconnection; Disconnection
IT WAS REPORTED THAT PRIOR TO USE WHILE ASSEMBLING THE TUBING, THE FILTER DOES NOT STAY ATTACHED TO THE TUBING AND DISCONNECTS FROM THE WHISPER SWIVEL VALVE EASILY, EVEN IF THE TUBING IS NOT MOVED. ADDITIONAL INFORMATION RECEIVED STATES THAT THE FILTERS ARE COMPATIBLE WITH THE SWIVEL VALVE. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16913299 ↗MalfunctionMay 11, 2023Disconnection; Disconnection
IT WAS REPORTED THAT PRIOR TO USE WHILE ASSEMBLING THE TUBING, THE FILTER DOES NOT STAY ATTACHED TO THE TUBING AND DISCONNECTS FROM THE WHISPER SWIVEL VALVE EASILY, EVEN IF THE TUBING IS NOT MOVED. ADDITIONAL INFORMATION RECEIVED STATES THAT THE FILTERS ARE COMPATIBLE WITH THE SWIVEL VALVE. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16913179 ↗MalfunctionMay 11, 2023Disconnection; Disconnection
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE FILTER DOES NOT STAY ATTACHED TO THE TUBING AND DISCONNECTS FROM THE WHISPER SWIVEL VALVE EASILY, EVEN IF THE TUBING IS NOT MOVED. THIS CAUSES A BREAK IN THE BREATHING CIRCUIT AND CUTS OFF THE DELIVERY OF OXYGEN TO A PATIENT WHO IS CONTINUOUSLY VENTILATOR DEPENDENT. ADDITIONAL INFORMATION RECEIVED STATES THAT THE FILTERS ARE COMPATIBLE WITH THE SWIVEL VALVE. THERE WAS NO PATIENT HARM, INJURY, OR DESATURATION AS THE ISSUE WAS DISCOVERED PRIOR TO USE WHEN ASSEMBLING THE TUBING. AS A RESULT, AN "ALTERNATIVE" FILTER WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16913003 ↗MalfunctionApr 18, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS "WATER LEAKING IN THE HME FILTER WHILE ATTACHED TO THE VENTILATOR AND PATIENT. OCCURED DURING VENTILATION TREATMENT. THE FILTER WAS REPLACED." AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. THE COMPLAINT FILE WILL BE UPDATED IF FURTHER INFORMATION IS RECEIVED.
Patient problems: Insufficient Information
MAUDE report 16765572 ↗MalfunctionMay 7, 2021Free or Unrestricted Flow; Free or Unrestricted Flow
IT WAS REPORTED THAT "DURING THE PRE-TEST, ALERT LEAKAGE". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11794508 ↗MalfunctionMay 7, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT "DURING THE PRE-TEST, ALERT LEAKAGE". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11794353 ↗
FDA clearance and classification
Cleared through 510(k) K964382 on Apr 15, 1997 (decision: Substantially Equivalent). Submitted by Gibeck, Inc. as “Humid-Vent Filter Compact Angled (18402)/Humid-Vent Filter Compact Straight(19402)”. The FDA review took 162 days. FDA record ↗
The FDA regulates product code CAH (filter, bacterial, breathing-circuit) under 21 CFR 868.5260 as a Class II device. 21 CFR 868.5260 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the GIBECK IPN043768
Has the GIBECK IPN043768 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04026704348022), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the GIBECK IPN043768?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between May 7, 2021 and Feb 20, 2025. In those reports, the most common specific device problems are “Disconnection” (4), “Blocked Connection” (1) and “Fluid/Blood Leak” (1); the most common patient problems are “Capsular Contracture” (1) and “Low Oxygen Saturation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the GIBECK IPN043768?
None of the reports are classified as deaths, 1 as an injury and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the GIBECK IPN043768 FDA approved?
It was cleared through 510(k) K964382 on Apr 15, 1997. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the GIBECK IPN043768, and is it still sold?
The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand GIBECK, model IPN043768). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Filter, bacterial, breathing-circuit devices
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit2 recalls570 reports
- VOCSN V+Pro Emergency, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall557 reports
- VOCSN V+O+C+S+N+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit4 recalls371 reports
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit3 recalls2 deaths330 reports
- VOCSN V+C+S+N+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall185 reports
- VOCSN VOCSN, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall1 death97 reports
- VOCSN V+C+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit2 recalls34 reports
- VOCSN V+O+C+S+N+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit2 recalls32 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page