Skip to content

HUDSON RCI IPN042652Recalls & reported problems

Teleflex Incorporated · BACTERIAL/VIRAL FILTER · Class 2

All models of HUDSON RCI (150)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04026704399208), received between Feb 16, 2021 and Feb 19, 2021 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Complete Blockage” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Complete Blockage1
  • Disconnection1

Patient problems

  • Airway Obstruction1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 19, 2021Complete Blockage; Complete Blockage

    CUSTOMER REPORTED THE PLASTIC PIECE THAT CONNECTS TO THE PATIENT ET TUBE AND HOOKS TO CIRCUIT WAS BLOCKED BY AN UNCUT PIECE OF PLASTIC. "PATIENT INDUCED UNDER GENERAL ANESTHESIA, LMA APPLIED. UNABLE TO HAVE RETURN OF ETCO2. CHANGED LMA TO ETT AND INTUBATED PATIENT, GRADE 1 VIEW. NO ETCO2, HIGH PEAK AIRWAY PRESSURES. PT HAD HISTORY OF ASTHMA, SO SUSPECTED BRONCHOSPASM (OCCLUSION OF THE AIRWAY OF SOME SORT). RECONFIRMED BY DL ETT WAS IN TRACHEA, MASK VENTILATION ALBUTEROL GIVEN. CALLED FOR HELP. EPINEPHRINE GIVEN. ETT SUCTIONED WITH NO OBSTRUCTION NOTED. STILL NO ETCO2 AND HIGH PEAK AIRWAY PRESSURES. CHANGED PATIENT FROM VENTILATOR TO AMBU BAG AND IMMEDIATE FOGGING OF ETT AND BILATERAL BREATH SOUNDS SO OCCLUSION WAS IN THE ANESTHESIA VENTILATOR CIRCUIT. REMOVED THE VIRAL FILTER FROM THE CIRCUIT AND IMMEDIATE ETCO2 WITH NORMAL PEAK AIRWAY PRESSURES." THE CUSTOMER REPORTED UPON INSPECTION THERE WAS "AN ENTIRE PLASTIC PIECE" THAT OCCLUDED THE FILTER AND MADE A "COMPLETE OBSTRUCTION TO THE PATIENT'S AIRWAY". PATIENT CONDITION UNKNOWN AT TIME OF REPORT.

    Patient problems: Airway Obstruction

    MAUDE report 11348093 ↗
  • MalfunctionFeb 16, 2021Disconnection; Disconnection

    IT WAS REPORTED THE "FILTER POPS OFF VENTILATOR DURING USE ON PATIENTS. INCIDENTS OCCURRED ON TRILOGY, V-60, AND LTV VENTS". NO PATIENT INJURTY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11333764 ↗

FDA clearance and classification

Cleared through 510(k) K961914 on Aug 14, 1996 (decision: Substantially Equivalent). Submitted by Hudson Respiratory Care, Inc. as “Bacteria/Viral Filter”. The FDA review took 89 days. FDA record ↗

The FDA regulates product code CAH (filter, bacterial, breathing-circuit) under 21 CFR 868.5260 as a Class II device. 21 CFR 868.5260 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HUDSON RCI IPN042652

Has the HUDSON RCI IPN042652 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04026704399208), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the HUDSON RCI IPN042652?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Feb 16, 2021 and Feb 19, 2021. In those reports, the most common specific device problems are “Complete Blockage” (1) and “Disconnection” (1); the most common patient problem is “Airway Obstruction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HUDSON RCI IPN042652?

None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HUDSON RCI IPN042652 FDA approved?

It was cleared through 510(k) K961914 on Aug 14, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the HUDSON RCI IPN042652, and is it still sold?

The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand HUDSON RCI, model IPN042652). Its record shows it as in commercial distribution.