Linox Smart SD 65/18 359067Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Linox Smart defibrillator lead has 192 reports, 67 injuries and 1 death. Many reports describe electrical noise on the lead, sometimes leading to inappropriate shocks, and some describe high shock impedance or fractures, with some systems removed because of infection. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
193
Deaths reported
1
Injuries reported
68
Recalls naming it
0

Summary

The FDA's MAUDE database contains 193 adverse event reports that list this exact device (DI 04035479100407), received between Jul 13, 2016 and Sep 29, 2026 — classified as 1 death, 68 injuries, 124 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (74 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Ambient Noise Problem74
  • High impedance41
  • Inappropriate/Inadequate Shock/Stimulation36
  • Adverse Event Without Identified Device or Use Problem31
  • Over-Sensing27
  • Fracture22
  • High Capture Threshold10
  • Break9

Patient problems

  • No Clinical Signs, Symptoms or Conditions69
  • No Known Impact Or Consequence To Patient57
  • Shock from Patient Lead(s)35
  • Unspecified Infection27
  • Ventricular Fibrillation2
  • Cardiomyopathy1
  • Hematoma1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 193 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50926760 ↗
  • InjurySep 25, 2026Unstable Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905338 ↗
  • InjuryApr 21, 2026High impedance

    LEAD WAS EXPLANTED DUE TO SHOCK IMPEDANCE GREATER THAN 150. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24947637 ↗
  • MalfunctionDec 30, 2025Ambient Noise Problem; Device Sensing Problem; Ambient Noise Problem; Device Sensing Problem

    IN OFFICE FOLLOW-UP SHOWS NOISE ON RV CHANNEL AFTER PACE DURING THRESHOLD TESTING. HMSC DATA SHOWS PRESENCE OF SHORT INTERVAL COUNTS FOR THIS DEVICE. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23926347 ↗
  • MalfunctionSep 3, 2025Ambient Noise Problem; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22958563 ↗
  • MalfunctionJul 30, 2025Fracture; Ambient Noise Problem; Fracture; Ambient Noise Problem

    LEAD WAS CAPPED DUE TO FRACTURE WITH NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22671798 ↗
  • MalfunctionJul 24, 2025High impedance; Ambient Noise Problem; High impedance; Ambient Noise Problem

    OUT OF RANGE SHOCK IMPEDANCE (GREATER THAN 150 OHMS) AND NOISE ON THE RV CHANNEL WERE SEEN IN RECORDINGS TRANSMITTED TO HOME MONITORING. LEAD TO BE EVALUATED FURTHER AND REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22593963 ↗
  • MalfunctionMar 11, 2025High impedance; High impedance

    HIGH SHOCK IMPEDANCES OF GREATER THAN 150 OHMS WAS REPORTED ON THIS RV LEAD. RECENTLY, SHOCK IMPEDANCE HAS FLUCTUATED INTERMITTENTLY TO IN-RANGE MEASUREMENTS ALSO. LEAD TO BE EVALUATED FURTHER AND REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21574635 ↗
  • InjuryNov 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT HAD INFECTION. THE DEVICE WAS EXPLANTED AND THE LEADS WERE CAPPED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 20592093 ↗
  • MalfunctionJul 15, 2024Ambient Noise Problem; Ambient Noise Problem

    THE PACE/SENSE PORTION OF THIS LEAD WAS CAPPED DUE TO NOISE. SHOCK PORTION REMAINS ACTIVE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19747148 ↗
  • MalfunctionJun 5, 2024Fracture; Fracture

    LEAD WAS EXPLANTED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19468747 ↗
  • MalfunctionApr 23, 2024Ambient Noise Problem; Ambient Noise Problem

    NOISE WAS SEEN ON RECORDINGS FOR THIS LEAD IN THE NEARFIELD AND FARFIELD SIGNALS. NO ADVERSE PATIENT EVENTS WERE REPORTED. LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19162731 ↗
  • MalfunctionFeb 26, 2024Fracture; Ambient Noise Problem; Fracture; Ambient Noise Problem

    PACE/SENSE PORTION OF LEAD WAS CAPPED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18778812 ↗
  • MalfunctionNov 6, 2023Under-Sensing; Under-Sensing

    THIS RV LEAD WAS EXPLANTED DUE TO LOW R WAVE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18073758 ↗
  • MalfunctionAug 11, 2023Ambient Noise Problem; Ambient Noise Problem

    REPRODUCIBLE NOISE WAS REPORTED ON THE ATRIAL AND FAR-FIELD CHANNELS DURING TESTING IN OFFICE. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17515331 ↗
  • MalfunctionJun 21, 2023Fracture; Ambient Noise Problem; Fracture; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE AND POSSIBLE FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17169852 ↗
  • MalfunctionJun 14, 2023High impedance; High Capture Threshold; High impedance; High Capture Threshold

    THIS LEAD WAS CAPPED DUE TO INCREASED IMPEDANCE (2134 OHMS) AND THRESHOLD ([email protected]). NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17131478 ↗
  • MalfunctionApr 3, 2023Ambient Noise Problem; Ambient Noise Problem

    NOISE CAN BE SEEN ON HMSC CAUSING INAPPROPRIATE VF DETECTIONS. NO INAPPROPRIATE THERAPY HAS BEEN DELIVERED. LEAD TRENDS ARE STABLE AND WITHIN NORMAL RANGE. THE PATIENT WILL BE BROUGHT IN FOR LEAD EVALUATION. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16665955 ↗
  • InjuryDec 30, 2022Fracture; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Fracture; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS LEAD SHOWS NOISE AND DELIVERED INAPPROPRIATE SHOCKS. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 16078242 ↗
  • MalfunctionNov 23, 2022Fracture; Ambient Noise Problem; Fracture; Ambient Noise Problem

    LEAD WAS CAPPED DUE TO FRACTURE CAUSING NOISE ON LEAD. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15852050 ↗
  • InjuryAug 23, 2022High impedance; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; High impedance; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    LEAD SHOWED IMPEDANCE GREATER THAN 3,000 AND NOISE RESULTING IN INAPPROPRIATE SHOCKS. LEAD WAS EXTRACTED VIA LASER WITH NO COMPLICATIONS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 15279754 ↗
  • InjuryAug 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15217036 ↗
  • MalfunctionAug 2, 2022Ambient Noise Problem; Ambient Noise Problem

    THIS RV LEAD WAS EXPLANTED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15147045 ↗
  • InjuryJul 28, 2022Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS LEAD WAS EXPLANTED DUE TO NOISE AND INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 15126790 ↗
  • MalfunctionJul 26, 2022High impedance; Loose or Intermittent Connection; High impedance; Loose or Intermittent Connection

    ON (B)(6) 2022 THE INFORMATION ABOUT THE LEAD IDENTITY WAS RECEIVED. THIS LEAD WAS CAPPED, LEFT IN THE PATIENT AND REPLACED BY A NEW ONE APPROX. 117 MONTHS AFTER THE IMPLANTATION. AFTER AN IMPLANTATION PERIOD OF APPROXIMATELY 20 DAYS IT WAS REPORTED THAT THE ICD WAS EXPLANTED DUE TO HIGH SHOCK IMPEDANCE. A CONNECTION ISSUE WAS OBSERVED BETWEEN LEAD AND ICD DURING THE EXPLANT PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15103971 ↗