Protego TD 65/16 359076Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 04035479100490), received between Jun 20, 2018 and Jun 8, 2026 — classified as 4 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4
  • High impedance4
  • Ambient Noise Problem3
  • Break3
  • Inappropriate/Inadequate Shock/Stimulation2
  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Capture1
  • High Capture Threshold1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Known Impact Or Consequence To Patient5
  • Shock from Patient Lead(s)2
  • Endocarditis1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 8, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO ENDOCARDITIS. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Endocarditis

    MAUDE report 25437274 ↗
  • InjuryMar 11, 2026Break

    LEAD WAS CAPPED DUE TO BREAK. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24573314 ↗
  • MalfunctionSep 4, 2025Fracture; High impedance; Ambient Noise Problem; Fracture; High impedance; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO FRACTURE, IMPEDANCE >3000 AND NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22965886 ↗
  • MalfunctionAug 15, 2024High impedance; High impedance

    LEAD WAS CAPPED DUE TO IMPEDANCE >3000. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19996534 ↗
  • MalfunctionJul 14, 2023Failure to Capture; High impedance; Failure to Capture; High impedance

    LEAD WAS CAPPED DUE TO OUT OF RANGE PACING IMPEDANCE (GREATER THAN 3000 OHMS) AND NON-CAPTURE AT MAXIMUM OUTPUT DURING INTERROGATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17326918 ↗
  • InjuryOct 20, 2022Break; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Break; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    LEAD WAS CAPPED DUE TO NOISE CAUSING VF DETECTION AND INAPPROPRIATE SHOCKS. LONG TERM IMPEDANCE GRAPH ALSO SHOWS INSULATION ISSUES. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 15642962 ↗
  • MalfunctionDec 22, 2020Fracture; High impedance; No Pacing; Fracture; High impedance; No Pacing

    THIS RV LEAD WAS CAPPED DUE TO A FRACTURE. IMPEDANCES WERE INCREASING AND THERE WAS NO PACING PRESENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11055667 ↗
  • MalfunctionJun 15, 2020Fracture; Fracture

    LEAD WAS EXPLANTED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10155361 ↗
  • InjuryJun 11, 2020Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    PATIENT CAME TO ER AFTER RECEIVING INAPPROPRIATE SHOCKS DUE TO NOISE ON THE RV LEAD. DETECTION WAS SWITCHED TO OFF.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 10144733 ↗
  • MalfunctionSep 25, 2019Break; Break

    THIS LEAD WAS EXPLANTED AND REPLACED DUE TO LEAD FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9118518 ↗
  • MalfunctionAug 26, 2019Fracture; Fracture

    THIS LEAD WAS EXPLANTED AND REPLACED DUE TO FRACTURE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8932774 ↗
  • MalfunctionOct 23, 2018Positioning Failure; Positioning Failure

    DURING THE IMPLANT OF THIS RV LEAD IT GOT STUCK, AND IF THE PHYSICIAN WOULD HAVE TAKEN IT OUT, IT WOULD HAVE PULLED OUT THE LV LEAD WHICH WOULD NOT PROVIDE BACKUP PACING. THE PHYSICIAN DECIDED TO CAP THIS LEAD AND IMPLANT AN ACTIVE RV LEAD.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7995511 ↗
  • MalfunctionJun 20, 2018High Capture Threshold; High Capture Threshold

    THIS LEAD WAS CAPPED DUE TO HIGH THRESHOLDS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7619800 ↗

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →