Protego TD 65/16 359076Recalls & reported problems
BIOTRONIK SE & Co. KG · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 04035479100490), received between Jun 20, 2018 and Jun 8, 2026 — classified as 4 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture4
- High impedance4
- Ambient Noise Problem3
- Break3
- Inappropriate/Inadequate Shock/Stimulation2
- Adverse Event Without Identified Device or Use Problem1
- Failure to Capture1
- High Capture Threshold1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient5
- Shock from Patient Lead(s)2
- Endocarditis1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryJun 8, 2026Adverse Event Without Identified Device or Use Problem
SYSTEM WAS EXPLANTED DUE TO ENDOCARDITIS. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.
Patient problems: Endocarditis
MAUDE report 25437274 ↗InjuryMar 11, 2026Break
LEAD WAS CAPPED DUE TO BREAK. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24573314 ↗MalfunctionSep 4, 2025Fracture; High impedance; Ambient Noise Problem; Fracture; High impedance; Ambient Noise Problem
THIS LEAD WAS CAPPED DUE TO FRACTURE, IMPEDANCE >3000 AND NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22965886 ↗MalfunctionAug 15, 2024High impedance; High impedance
LEAD WAS CAPPED DUE TO IMPEDANCE >3000. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19996534 ↗MalfunctionJul 14, 2023Failure to Capture; High impedance; Failure to Capture; High impedance
LEAD WAS CAPPED DUE TO OUT OF RANGE PACING IMPEDANCE (GREATER THAN 3000 OHMS) AND NON-CAPTURE AT MAXIMUM OUTPUT DURING INTERROGATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17326918 ↗InjuryOct 20, 2022Break; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Break; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem
LEAD WAS CAPPED DUE TO NOISE CAUSING VF DETECTION AND INAPPROPRIATE SHOCKS. LONG TERM IMPEDANCE GRAPH ALSO SHOWS INSULATION ISSUES. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Patient problems: Shock from Patient Lead(s)
MAUDE report 15642962 ↗MalfunctionDec 22, 2020Fracture; High impedance; No Pacing; Fracture; High impedance; No Pacing
THIS RV LEAD WAS CAPPED DUE TO A FRACTURE. IMPEDANCES WERE INCREASING AND THERE WAS NO PACING PRESENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11055667 ↗MalfunctionJun 15, 2020Fracture; Fracture
LEAD WAS EXPLANTED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10155361 ↗InjuryJun 11, 2020Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem
PATIENT CAME TO ER AFTER RECEIVING INAPPROPRIATE SHOCKS DUE TO NOISE ON THE RV LEAD. DETECTION WAS SWITCHED TO OFF.
Patient problems: Shock from Patient Lead(s)
MAUDE report 10144733 ↗MalfunctionSep 25, 2019Break; Break
THIS LEAD WAS EXPLANTED AND REPLACED DUE TO LEAD FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9118518 ↗MalfunctionAug 26, 2019Fracture; Fracture
THIS LEAD WAS EXPLANTED AND REPLACED DUE TO FRACTURE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8932774 ↗MalfunctionOct 23, 2018Positioning Failure; Positioning Failure
DURING THE IMPLANT OF THIS RV LEAD IT GOT STUCK, AND IF THE PHYSICIAN WOULD HAVE TAKEN IT OUT, IT WOULD HAVE PULLED OUT THE LV LEAD WHICH WOULD NOT PROVIDE BACKUP PACING. THE PHYSICIAN DECIDED TO CAP THIS LEAD AND IMPLANT AN ACTIVE RV LEAD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7995511 ↗MalfunctionJun 20, 2018High Capture Threshold; High Capture Threshold
THIS LEAD WAS CAPPED DUE TO HIGH THRESHOLDS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7619800 ↗
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page