Linox Smart S DX 65/17 365501Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Linox Smart S defibrillator lead is a wire implanted in the heart that carries shock and pacing signals. Many of the 289 reports describe electrical noise, high resistance readings, and wire fractures, and some describe unexpected shocks. Some leads were switched off or removed after these problems. The reports include no deaths and 85 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
290
Deaths reported
0
Injuries reported
86
Recalls naming it
0

Summary

The FDA's MAUDE database contains 290 adverse event reports that list this exact device (DI 04035479110024), received between Aug 12, 2018 and Sep 28, 2026 — classified as 86 injuries, 204 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (129 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 4 reportsMay 2024: 3 reportsJun 2024: 4 reportsJul 2024: 3 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 5 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 3 reportsMay 2025: 3 reportsJun 2025: 3 reportsJul 2025: 2 reportsAug 2025: 5 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 4 reportsFeb 2026: 4 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Ambient Noise Problem129
  • High impedance62
  • Inappropriate/Inadequate Shock/Stimulation40
  • Fracture37
  • Adverse Event Without Identified Device or Use Problem33
  • Over-Sensing25
  • Device Sensing Problem17
  • High Capture Threshold16

Patient problems

  • No Clinical Signs, Symptoms or Conditions160
  • No Known Impact Or Consequence To Patient51
  • Shock from Patient Lead(s)40
  • Unspecified Infection32
  • Twiddlers Syndrome3
  • Ventricular Fibrillation3
  • Fall1
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 290 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Impedance Problem

    This lead was explanted due to OOR shock impedance. No adverse patient events were reported. Should additional information be received this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916445 ↗
  • InjurySep 16, 2026Ambient Noise Problem

    This lead was capped due to noise. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50788086 ↗
  • InjuryAug 6, 2026Fracture; High impedance

    This lead was explanted due to increased impedance above normal range consistent with possible lead fracture. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50164028 ↗
  • InjuryAug 4, 2026Insufficient Device Problem Information

    This lead was explanted due to chronic lead malfunction. Should additional information become available, this file will be updated.

    Patient problems: Insufficient Information

    MAUDE report 50134817 ↗
  • InjuryMay 26, 2026Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25290108 ↗
  • InjuryApr 15, 2026Failure to Capture; High impedance; High Capture Threshold

    LEAD WAS CAPPED DUE TO HIGH THRESHOLD (FAILURE TO CAPTURE AT MAX OUTPUT) AND HIGH PACING IMPEDANCE OF 2,700+ OHMS. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24891061 ↗
  • InjuryApr 9, 2026Fracture

    THIS LEAD WAS CAPPED DUE TO FRACTURED LEAD. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24829866 ↗
  • InjuryApr 8, 2026Fracture

    THIS RV LEAD WAS EXPLANTED DUE TO FRACTURE. NO OTHER ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24822183 ↗
  • MalfunctionFeb 25, 2026High impedance

    HOME MONITORING TREND FOR SHOCK IMPEDANCE OUT OF RANGE GREATER THAN 150 OHMS INTERMITTENTLY SINCE FEBRUARY 6. THE LAST TWO DAYS HAS RETURNED TO NORMAL. PATIENT WAS CHECKED IN OFFICE THIS MORNING AND IMPEDANCE WAS NORMAL AND NO NOISE SEEN. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24450114 ↗
  • MalfunctionFeb 12, 2026Ambient Noise Problem

    HMSC SHOWED RV LEAD MONITORING RECORDING SHOWING NOISE ON RV AND FF CHANNEL. ALL LEAD TRENDS APPEAR UNREMARKABLE. SHORT INTERVAL COUNTER INCREASING SINCE (B)(6) 2026. ADVISED FURTHER LEAD EVALUATION. LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24327390 ↗
  • InjuryFeb 5, 2026Fracture; Ambient Noise Problem; Impedance Problem

    HMSC REPORTED SHOCK IMPEDANCE OUT OF RANGE ON 5-FEB-2026, THE DAY AFTER A GENERATOR CHANGEOUT. RECORDINGS SHOWS NOISE ON ATRIAL CHANNEL OF LEAD. LEAD TO BE EVALUATED FURTHER AND CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24274000 ↗
  • InjuryFeb 3, 2026High impedance; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    HMSC REPORTED NOISE RESULTING IN INAPPROPRIATE SHOCKS AND TEMPORARY INCREASE IN PACING IMPEDANCE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED. LEAD REMAINS IMPLANTED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24248127 ↗
  • InjuryJan 29, 2026Over-Sensing

    THIS LEAD WAS EXPLANTED DUE TO OVERSENSING. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24216947 ↗
  • InjuryJan 26, 2026High Capture Threshold

    THIS LEAD WAS EXPLANTED DUE TO ELEVATED THRESHOLDS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24178255 ↗
  • InjuryJan 14, 2026Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS RV LEAD WAS CAPPED DUE TO NOISE LEADING TO INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24066828 ↗
  • InjuryJan 12, 2026Ambient Noise Problem

    HMSC REPORTED LEAD NOISE ON ATRIAL AND RV CHANNEL RESULTING IN VF DETECTION ON (B)(6) 2025. SHOCK THERAPY ABORTED. LEAD TRENDS APPEAR STABLE. NOISE WAS NOT REPRODUCIBLE DURING OFFICE INTERROGATION ON (B)(6) 2026. FURTHER EVALUATION TO BE DISCUSSED WITH PHYSICIAN. LEAD CURRENTLY REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24038522 ↗
  • MalfunctionDec 30, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    THIS RV LEAD WAS CUT AT YOKE AND CAPPED DUE TO LEAD COMPROMISE AND INAPPROPRIATE THERAPIES. NO FURTHER DETAILS KNOWN. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23927394 ↗
  • MalfunctionDec 9, 2025High impedance; High impedance

    THE SHOCK IMPEDANCE HAS TRENDED >150 OHMS SINCE (B)(6) 2025. OTHER LEAD TRENDS APPEARED STABLE AND WITHIN NORMAL RANGE. THE PATIENT WILL BE BROUGHT IN FOR A LEAD EVALUATION. LEAD REMAINS IMPLANTED. NO ADVERSE EVENTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23761976 ↗
  • MalfunctionNov 25, 2025High impedance; High impedance

    THIS LEAD SHOWS INCREASING SHOCK IMPEDANCE AND REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23652453 ↗
  • MalfunctionNov 20, 2025High impedance; Ambient Noise Problem; High impedance; Ambient Noise Problem

    HMSC REPORTED EPISODES SHOWING NOISE ON ATRIAL CHANNEL AND A SUDDEN INCREASE IN SHOCK IMPEDANCE IN APRIL 2025. ADVISED FURTHER LEAD EVALUATION. LEAD CURRENTLY REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23609699 ↗
  • MalfunctionOct 17, 2025Impedance Problem; Impedance Problem

    THIS LEAD WAS CAPPED DUE TO IMPEDANCE DROPS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23327572 ↗
  • MalfunctionSep 11, 2025Ambient Noise Problem; Ambient Noise Problem

    ATRIAL NOISE REPORTED ON HOME MONITORING EPISODES. THESE SEEM TO HAVE STARTED IN JULY INTERMITTENTLY BUT PATIENT DID HAVE A RECENT IN OFFICE FOLLOW UP IN JUNE WHERE ALL WAS NORMAL. ALL OTHER TRENDS APPEAR STABLE. LEAD REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23025960 ↗
  • MalfunctionAug 29, 2025Break; Break

    WHEN SCREWING THIS LED INTO NEW DEVICE, MD PLUGGED RV COIL IN, TURNED SCREW, AND WHEN HE GENTLY TUGGED THE ENTIRE LEAD TORE (LEAVING PIN IN HEADER AND EXPOSED WIRES ON LEAD). THE LEAD WAS CAPPED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22926846 ↗
  • InjuryAug 22, 2025Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    NOISE SEEN ON RECORDINGS. SHORT INTERVALS GRAPH ALSO POPULATING WITH EVENTS THAT CAN BE INDICATIVE OF LEAD NOISE. LEAD REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Shock from Patient Lead(s)

    MAUDE report 22875224 ↗
  • MalfunctionAug 21, 2025Impedance Problem; Impedance Problem

    THIS RV LEAD WAS EXPLANTED DUE TO IMPEDANCE OUT OF RANGE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22866065 ↗

Other Permanent defibrillator electrodes devices