Linox Smart S 60 375012Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Linox Smart S 60 lead is a permanent defibrillator electrode implanted under the skin to deliver pacing and shock therapy. Most reports describe noise on the lead's signals, high impedance, broken or cracked lead wires, and shocks given when none were needed; some leads were capped or removed and replaced. There are 53 reports, including 1 death and 22 injuries.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
53
Deaths reported
1
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 53 adverse event reports that list this exact device (DI 04035479116194), received between Jan 3, 2019 and Jan 29, 2024 — classified as 1 death, 22 injuries, 30 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Ambient Noise Problem20
  • High impedance12
  • Adverse Event Without Identified Device or Use Problem11
  • Inappropriate/Inadequate Shock/Stimulation10
  • Fracture7
  • Under-Sensing4
  • Impedance Problem3
  • Low impedance3

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • No Known Impact Or Consequence To Patient15
  • Shock from Patient Lead(s)11
  • Unspecified Infection9
  • Pocket Erosion2
  • Chest Pain1
  • Endocarditis1
  • Ventricular Fibrillation1

Recent adverse event reports

The 25 most recent of 53 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2024Fracture; Fracture

    THIS LEAD WAS EXPLANTED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18595120 ↗
  • MalfunctionJul 25, 2023Ambient Noise Problem; Ambient Noise Problem

    THIS RV LEAD WAS CAPPED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17389781 ↗
  • MalfunctionNov 30, 2022Ambient Noise Problem; Ambient Noise Problem

    RV NOISE SEEN ON HOMEMONITORING IEGMS SHOWING AFTER T WAVES. NOISE SEEMS TO BE OCCURRING MORE OFTEN. COULD NOT BE REPRODUCED IN OFFICE. WILL BE DISCUSSED FURTHER WITH PHYSICIAN ON NEXT STEPS. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15886060 ↗
  • MalfunctionJun 8, 2022Ambient Noise Problem; Ambient Noise Problem

    HMSC REPORTED NOISE RESULTING IN A VF DETECTION (SHOCK ABORTED). LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14632213 ↗
  • MalfunctionApr 14, 2022Failure to Capture; Fracture; Impedance Problem; Failure to Capture; Fracture; Impedance Problem

    PACING PORTION OF THE LEAD WAS CAPPED DUE TO MALFUNCTION. FRACTURE CAUSED OUT OF RANGE IMPEDANCE AND LOSS OF CAPTURE. ONLY IS-1 PORTION OF LEAD WAS CAPPED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14116131 ↗
  • MalfunctionMar 10, 2022Pacing Problem; Ambient Noise Problem; Device Sensing Problem; Pacing Problem; Ambient Noise Problem; Device Sensing Problem

    NOISE ON RV CHANNEL REPRODUCEABLE WITH ISOMETRICS. CAPTURE AND SENSING APPEAR APPROPRIATE HOWEVER PATIENT HAS HAD EPISODES OF POTENTIAL PACING INHIBITION AND EPISODES OF LOW SENSING. FULL LEAD EVALUATION RECOMMENDED. LEAD REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13726766 ↗
  • MalfunctionFeb 14, 2022Break; Break

    DURING A NORMAL DEVICE CHANGE OUT, A CUT WAS FOUND IN THE INSULATION OF THIS LEAD. THE LEAD WAS CAPPED AND REPLACED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13524806 ↗
  • MalfunctionFeb 14, 2022High impedance; High impedance

    HIGH SHOCK IMPEDANCES OF GREATER THAN 150 OHMS WERE REPORTED ON THIS LEAD. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13524624 ↗
  • MalfunctionJan 12, 2022Ambient Noise Problem; Ambient Noise Problem

    THIS RV LEAD WAS EXPLANTED DUE TO NOISE. NOISE WAS ABLE TO BE RECREATED WITH POCKET MANIPULATION. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13239968 ↗
  • InjuryJan 4, 2022Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    OVERSENSING AND NOISE ON THE RV LEAD, LED TO 7 INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 13154801 ↗
  • InjuryNov 16, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LEAD EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: Unspecified Infection

    MAUDE report 12824692 ↗
  • InjurySep 10, 2021Ambient Noise Problem; Ambient Noise Problem

    LEAD EXPLANTED IN 2017 DUE TO NOISE CAUSING INAPPROPRIATE SHOCKS. LEAD WAS RETURNED TO BIOTRONIK FOR ANALYSIS ON 16-AUG-2021. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 12454518 ↗
  • InjurySep 7, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LEAD EXPLANTED DUE TO ENDOCARDITIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: Endocarditis

    MAUDE report 12436373 ↗
  • InjuryJul 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LEAD EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: Unspecified Infection

    MAUDE report 12099174 ↗
  • MalfunctionJun 30, 2021Fracture; Low impedance; Ambient Noise Problem; Fracture; Low impedance; Ambient Noise Problem

    IT WAS REPORTED THAT THIS LEAD WAS CAPPED DUE TO FRACTURE CAUSING NOISE ON THE RV CHANNEL. A DECREASE IN SHOCK IMPEDANCE WAS ALSO NOTICED AT THE BEGINNING OF JUNE 2021. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12096344 ↗
  • InjuryJun 4, 2021Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    RV LEAD WAS CAPPED DUE TO NOISE. INAPPROPRIATE SHOCK DELIVERED AND PACING INHIBITED DUE TO PACER DEPENDENCY. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 11942700 ↗
  • DeathMay 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LEAD CAPPED DUE TO INFECTION WITH EROSION. DEVICE AND RA LEAD EXTRACTED. LV AND RV CAPPED. PATIENT EXPIRED ON THE SAME DAY AS THE PROCEDURE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: Pocket Erosion; Unspecified Infection

    MAUDE report 11892981 ↗
  • MalfunctionApr 21, 2021High impedance; High impedance

    PER CLINICIAN, SHOCK IMPEDANCE INCREASING IN THE PAST FEW DAYS. ADVISED TO EVALUATE LEAD FURTHER. LEAD CURRENTLY REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11700944 ↗
  • InjuryMar 10, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 11455887 ↗
  • InjuryFeb 22, 2021Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    THIS LEAD WAS CAPPED AND REPLACED DUE TO 20 INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 11361235 ↗
  • InjuryFeb 18, 2021Ambient Noise Problem; Ambient Noise Problem

    THIS RV LEAD WAS EXPLANTED AND REPLACED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Chest Pain; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11345948 ↗
  • InjuryJan 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 11112652 ↗
  • InjuryDec 15, 2020Fracture; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Fracture; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS LEAD WAS CAPPED AND REPLACED DUE TO MULTIPLE SHOCKS FROM NOISE. PHYSICIAN SUSPECTED FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 11015299 ↗
  • MalfunctionDec 4, 2020High impedance; High impedance

    INTERMITTENT OUT OF RANGE SHOCK IMPEDANCE WAS REPORTED ON THIS LEAD (GREATER THAN 150 OHMS), BEGINNING IN (B)(6) OF 2020. OUT OF RANGE MEASUREMENTS COULD NOT BE REPRODUCED WITH POSTURAL TESTING AND POCKET MANIPULATION IN OFFICE. LEAD REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10945395 ↗
  • MalfunctionNov 24, 2020Over-Sensing; Under-Sensing; Low impedance; Over-Sensing; Under-Sensing; Low impedance

    DEVICE WAS CAPPED DUE TO UNDERSENSING, OVERSENSING, AND LOW IMPEDANCE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10891736 ↗

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →