Protego S 65 379969Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Protego S 65 defibrillator lead has 388 reports in the FDA database, with 150 injuries and no deaths. Many reports describe electrical noise being read as false signals, which can lead to inappropriate shocks, and some describe fractured leads or high electrical resistance. Some patients felt shocks from the lead, and some leads were switched off (capped). No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
393
Deaths reported
0
Injuries reported
153
Recalls naming it
0

Summary

The FDA's MAUDE database contains 393 adverse event reports that list this exact device (DI 04035479121488), received between Oct 6, 2017 and Sep 16, 2026 — classified as 153 injuries, 240 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (218 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 4 reportsApr 2024: 3 reportsMay 2024: 2 reportsJun 2024: 4 reportsJul 2024: 8 reportsAug 2024: 3 reportsSep 2024: 7 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 3 reportsJan 2025: 5 reportsFeb 2025: 4 reportsMar 2025: 4 reportsApr 2025: 2 reportsMay 2025: 4 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 2 reportsOct 2025: 4 reportsNov 2025: 3 reportsDec 2025: 2 reportsJan 2026: 5 reportsFeb 2026: 5 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Ambient Noise Problem218
  • Inappropriate/Inadequate Shock/Stimulation87
  • Fracture76
  • Adverse Event Without Identified Device or Use Problem55
  • High impedance45
  • Over-Sensing32
  • Impedance Problem14
  • Low impedance13

Patient problems

  • No Clinical Signs, Symptoms or Conditions199
  • Shock from Patient Lead(s)88
  • No Known Impact Or Consequence To Patient56
  • Unspecified Infection49
  • Twiddlers Syndrome4
  • Pocket Erosion2
  • Sepsis2
  • Ventricular Fibrillation2

Recent adverse event reports

The 25 most recent of 393 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2026Ambient Noise Problem

    This lead was explanted due to noise. No adverse patient events were reported. Should additional information become available, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50792396 ↗
  • MalfunctionSep 16, 2026Ambient Noise Problem

    This lead shows noise and remains implanted. No adverse patient events were reported. Should additional information become available, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50789933 ↗
  • InjurySep 8, 2026High impedance

    This lead was capped due to increased RV lead impedance showing >3000 ohms. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50671257 ↗
  • InjurySep 8, 2026Fracture; Low impedance

    This lead was explanted due to insulation failure and impedance <200ohms. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50669905 ↗
  • MalfunctionSep 3, 2026Ambient Noise Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50590289 ↗
  • InjuryAug 31, 2026Fracture

    This lead was capped due to signs of progressing fracture. No adverse patient events were reported. Should additional information become available, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50527648 ↗
  • InjuryAug 20, 2026High impedance; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    This RV lead was capped due to noise resulting in inappropriate shock and pace/sense impedance >3000 ohms. Should additional information be received this file will be updated.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50348020 ↗
  • InjuryAug 7, 2026High impedance

    Lead was capped due to Pace/sense impedance >3000 ohms. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50180370 ↗
  • MalfunctionJun 26, 2026Ambient Noise Problem; Device Sensing Problem

    DOUBLE-COUNTED EVENTS SEEN ON THE RV CHANNEL DURING SENSING IN A VF EPISODE AND HIGH SHORT RV INTERVAL COUNT. IMPEDANCE IS UNREMARKABLE AND THE RV CAPTURE CONTROL TEST CONSISTENTLY FAILS SO NO DATA WAS AVAILABLE REGARDING PACING THRESHOLD. FULL LEAD EVALUATION WITH POSTURAL TESTING AND ISOMETRICS WAS RECOMMENDED. LEAD REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25643769 ↗
  • InjuryMay 11, 2026Ambient Noise Problem

    LEAD WAS CAPPED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25137048 ↗
  • InjuryApr 27, 2026Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    NOISE SEEN IN MULTIPLE IEGMS. PATIENT RECEIVED 4 INAPPROPRIATE SHOCKS. LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 25010266 ↗
  • MalfunctionMar 11, 2026Ambient Noise Problem

    BRIEF NOISE SEEN ON FF IN ONE IEGM. NO IMPEDANCE OR RV CHANNEL CHANGES. DEVICE REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24572108 ↗
  • InjuryMar 2, 2026Ambient Noise Problem

    NOISE SEEN ON THIS RV LEAD. NO ADVERSE PATIENT EVENTS WERE REPORTED. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24483445 ↗
  • InjuryFeb 25, 2026Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    RV LEAD NOISE LED TO INAPPROPRIATE VF DETECTION AND INAPPROPRIATE SHOCKS. PATIENT ENDED UP IN EMERGENCY ROOM. LEAD REMAINS IMPLANTED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24449976 ↗
  • InjuryFeb 20, 2026Impedance Problem; Unstable Capture Threshold

    RV LEAD CAPPED DUE TO THRESHOLD AND IMPEDANCE OUT OF RANGE POST AUTOMOBILE ACCIDENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24403467 ↗
  • InjuryFeb 16, 2026Pacing Problem; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE CAUSING INHIBITION OF PACING. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24351778 ↗
  • InjuryFeb 10, 2026Fracture

    THIS RV LEAD WAS CAPPED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24307236 ↗
  • InjuryFeb 10, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24303451 ↗
  • InjuryJan 26, 2026Adverse Event Without Identified Device or Use Problem

    THIS DEVICE WAS EXPLANTED DUE TO BACTEREMIA INFECTION. PATIENT WAS ADMITTED ON (B)(6) 2026, IS STILL HOSPITALIZED, AND ALSO HAS LVAD AND WOUND VAC. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Bacteremia

    MAUDE report 24185390 ↗
  • InjuryJan 20, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24125300 ↗
  • InjuryJan 20, 2026Ambient Noise Problem; Impedance Problem; Unstable Capture Threshold

    THIS LEAD SHOWS NOISE PLUS IMPEDANCE AND THRESHOLD ISSUES. LEAD REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24119887 ↗
  • MalfunctionJan 12, 2026High impedance; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO SHOCK IMPEDANCE >150 OHMS AND NOISE ON RV CHANNEL. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24043112 ↗
  • MalfunctionJan 9, 2026Intermittent Capture; Fracture; Over-Sensing; Pacing Problem

    THIS RV LEAD WAS CAPPED DUE TO FRACTURE, OVERSENSING, INTERMITTENT LOSS OF CAPTURE, AND PACING INHIBITION. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24030791 ↗
  • InjuryDec 15, 2025Fracture; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Fracture; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO FRACTURE CREATING NOISE AND CAUSING OVERSENSING, WHICH LED TO INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 23804899 ↗
  • InjuryDec 3, 2025Break; Over-Sensing; Ambient Noise Problem; Break; Over-Sensing; Ambient Noise Problem

    T-WAVE OVERSENSING COUPLED WITH SOME EVENTS ON RECORDING THAT APPEAR TO BE LEAD NOISE WERE REPORTED. ADVISED TO BRING PATIENT IN FOR FULL FOLLOW UP WITH ISOMETRIC TESTING. OTHER TRENDS APPEAR NORMAL. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23704659 ↗

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page