Protego S 60 379970Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Protego S 60 defibrillator lead is a wire that connects a heart defibrillator to the heart. Many of the 113 reports describe electrical noise on the lead, which can lead to inappropriate shocks, and some describe the lead fracturing or showing abnormal electrical resistance. Several leads were removed or capped. The reports include 45 injuries and no deaths, and 19 list an infection. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
113
Deaths reported
0
Injuries reported
45
Recalls naming it
0

Summary

The FDA's MAUDE database contains 113 adverse event reports that list this exact device (DI 04035479121495), received between Dec 9, 2016 and Sep 11, 2026 — classified as 45 injuries, 68 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (53 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Ambient Noise Problem53
  • Adverse Event Without Identified Device or Use Problem21
  • Fracture21
  • Inappropriate/Inadequate Shock/Stimulation20
  • High impedance15
  • High Capture Threshold5
  • Break4
  • Device Dislodged or Dislocated4

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • No Known Impact Or Consequence To Patient22
  • Shock from Patient Lead(s)20
  • Unspecified Infection19
  • Pocket Erosion2
  • Cardiopulmonary Arrest1
  • Failure of Implant1
  • Myocarditis1

Recent adverse event reports

The 25 most recent of 113 reports, in the reporter's own words as published by the FDA.

  • InjurySep 11, 2026Ambient Noise Problem

    This lead was capped due to Noise. No adverse patient events were reported. Should additional information be received this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50733569 ↗
  • InjuryAug 12, 2026Fracture

    This lead was capped due to suspected lead fracture. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50230169 ↗
  • InjuryJun 24, 2026Ambient Noise Problem

    THIS RV LEAD WAS CAPPED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25610663 ↗
  • InjuryApr 13, 2026Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS RV LEAD HAD NOISE LEADING TO INAPPROPRIATE SHOCKS. LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24867727 ↗
  • InjuryMar 5, 2026Impedance Problem

    THIS RV LEAD WAS EXPLANTED DUE TO STRONG, RAPID DEVIATIONS OF LEAD IMPEDANCE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS REPORT WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24515658 ↗
  • MalfunctionJan 21, 2026Ambient Noise Problem

    NOISE CAN BE SEEN ON THE RV CHANNEL LEADING TO AN INAPPROPRIATE NST DETECTION. THE SHORT RV INTERVAL COUNTER BEGAN INCREMENTING ON (B)(6) 2026. NO NOISE COULD BE REPRODUCED IN PERSON ON (B)(6) 2026 AFTER PERFORMING POSTURAL AND ISOMETRIC TESTING. ALL LEAD TRENDS ARE STABLE AND WITHIN NORMAL RANGE. LEAD REMAINS IMPLANTED, NO ADVERSE EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24137063 ↗
  • InjuryJan 5, 2026Fracture; Inappropriate/Inadequate Shock/Stimulation

    THIS RV LEAD WAS EXPLANTED DUE TO FRACTURE WITH INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 23967253 ↗
  • MalfunctionDec 3, 2025Ambient Noise Problem; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23711558 ↗
  • MalfunctionOct 15, 2025Fracture; Fracture

    THIS RV LEAD WAS CAPPED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23306219 ↗
  • MalfunctionSep 10, 2025Failure to Capture; Impedance Problem; Failure to Capture; Impedance Problem

    HMSC REPORTED EPISODES SHOWING LOSS OF CAPTURE. SHOCK IMPEDANCE TREND INCREASING. ADVISED FURTHER LEAD EVALUATION. LEAD CURRENTLY REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23021446 ↗
  • MalfunctionAug 12, 2025Ambient Noise Problem; Ambient Noise Problem

    NOISE WAS REPORTED ON THE RV CHANNEL. TACHY THERAPIES HAVE BEEN DISABLED AND PHYSICIAN WILL BE CONSULTED FOR NEXT STEPS. DEVICE REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22782062 ↗
  • MalfunctionAug 5, 2025Fracture; Fracture

    THIS LEAD WAS CAPPED DUE TO FRACTURED LEAD. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22723326 ↗
  • MalfunctionJul 18, 2025Ambient Noise Problem; Ambient Noise Problem

    HM REVEALED NOISE POTENTIALS PRE AND POST QRS. LEAD REMAINS IMPLANTED AND PATIENT TO BE BROUGHT IN FOR IN OFFICE F/U. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22541238 ↗
  • InjuryJul 9, 2025Fracture; Ambient Noise Problem; Fracture; Ambient Noise Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. LEAD FRACTURE AND NOISE WAS ALSO NOTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22459085 ↗
  • MalfunctionJun 27, 2025Ambient Noise Problem; Ambient Noise Problem

    THIS RV LEAD WAS EXPLANTED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22346906 ↗
  • MalfunctionApr 17, 2025Ambient Noise Problem; Ambient Noise Problem

    THIS LEAD WAS EXPLANTED DUE TO NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21856076 ↗
  • InjuryFeb 6, 2025Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS RV LEAD WAS CAPPED DUE TO NOISE LEADING TO INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 21327356 ↗
  • MalfunctionJan 24, 2025High impedance; Over-Sensing; High Capture Threshold; High impedance; Over-Sensing; High Capture Threshold

    RV LEAD WAS CAPPED AND REPLACED DUE TO HIGH PACING IMPEDANCE (>200 OHMS). OVERSENSING AND HIGH THRESHOLD ALSO NOTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21229345 ↗
  • InjuryNov 8, 2024Fracture; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Fracture; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    LEAD HAD SIGNS OF FRACTURE. ONE INAPPROPRIATE SHOCK DELIVERED DUE TO NOISE. LEAD WAS EXPLANTED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 20640518 ↗
  • InjuryOct 3, 2024Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE AND INAPPROPRIATE SHOCK. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 20363839 ↗
  • MalfunctionSep 18, 2024Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated

    LEAD WAS CAPPED DUE TO DISLODGEMENT AND FRACTURE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20252430 ↗
  • InjuryJul 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS LEAD WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 19761849 ↗
  • MalfunctionApr 18, 2024Ambient Noise Problem; Device Sensing Problem; Ambient Noise Problem; Device Sensing Problem

    LEAD WAS EXPLANTED DUE TO NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19140226 ↗
  • InjuryApr 3, 2024Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE AND INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 19030978 ↗
  • MalfunctionMar 28, 2024Ambient Noise Problem; Ambient Noise Problem

    EXTRA SYSTOLES NOTED ON NEAR FIELD IEGM THAT IS REPRODUCIBLE, SHOWING NOISE. LEAD REMAINS IMPLANTED. MD STATED LEAD WILL NEED TO BE CHANGED OUT IN NEAR FUTURE WITH GENERATOR CHANGE OUT. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18997786 ↗