Plexa S 60Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Quick summary

This Biotronik Plexa S 60 is a permanently implanted lead (electrode) for a heart defibrillator. The 33 reports mostly describe the lead picking up electrical noise, failing to sense the heartbeat properly, or needing to be capped or taken out — several because of infection. Reports mention 18 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
18
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 04035479133139), received between Aug 6, 2019 and Apr 27, 2026 — classified as 18 injuries, 15 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem14
  • Ambient Noise Problem7
  • Device Dislodged or Dislocated7
  • High Capture Threshold3
  • Under-Sensing3
  • High impedance2
  • Break1
  • Device Sensing Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Unspecified Infection10
  • No Known Impact Or Consequence To Patient4
  • Cardiac Perforation2
  • Endocarditis1
  • Implant Pain1
  • Perforation1
  • Shock from Patient Lead(s)1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 27, 2026High impedance; Ambient Noise Problem

    OUT OF RANGE SHOCK IMPEDANCE REPORTED >150 OHMS. NOISE ALSO SEEN ON FAR FIELD SIGNAL. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25010196 ↗
  • InjuryMar 4, 2026Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24510050 ↗
  • InjuryNov 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO ENDOCARDITIS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Endocarditis

    MAUDE report 23658352 ↗
  • MalfunctionMay 23, 2025Ambient Noise Problem; Ambient Noise Problem

    LEAD WAS CAPPED DUE TO NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22077796 ↗
  • MalfunctionDec 4, 2024Under-Sensing; Under-Sensing

    THIS RV LEAD WAS CAPPED DUE TO UNDERSENSING SINCE (B)(6) 2019 IMPLANT. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20847179 ↗
  • InjurySep 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED, DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 20141680 ↗
  • InjurySep 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 17814638 ↗
  • MalfunctionSep 12, 2023Ambient Noise Problem; Ambient Noise Problem

    RV LEAD CAPPED DUE TO NOISE FOLLOWING DFT TEST. (DFT TESTED DUE TO LOW R WAVES ). NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17731534 ↗
  • MalfunctionFeb 6, 2023Ambient Noise Problem; Ambient Noise Problem

    THIS LEAD WAS EXPLANTED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16304167 ↗
  • MalfunctionMar 9, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LEAD CAPPED DUE TO PATIENT EXPERIENCING POCKET PAIN. THE PATIENT WAS DOWNGRADE TO CRTP. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: Implant Pain

    MAUDE report 13711921 ↗
  • MalfunctionDec 30, 2021Device Sensing Problem; High Capture Threshold; Device Sensing Problem; High Capture Threshold

    HM REPORTED RV SENSING AMPLITUDE BELOW LIMIT (LESS THAN 2 MV). SENSING AMPLITUDE HAS BEEN DECREASING SINCE IMPLANT. THERE WAS A SUDDEN INCREASE IN PACING THRESHOLD EARLY (B)(6). THE LEAD WAS REPOSITIONED ON (B)(6) 2021. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13122181 ↗
  • InjuryMay 17, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RV LEAD WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 11838311 ↗
  • MalfunctionMay 10, 2021Under-Sensing; Under-Sensing

    RV LEAD WAS EXPLANTED AND REPLACED DUE TO LOW RV SENSING DETECTED THROUGH DEVICE INTERROGATION. RV SENSING WAS LESS THAN 2.0MV. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11802680 ↗
  • MalfunctionMay 3, 2021Low impedance; Ambient Noise Problem; Low impedance; Ambient Noise Problem

    LEAD EXPLANTED DUE TO RV LEAD DEMONSTRATING NOISE AND IMPEDANCE DROP. WHOLE NEW SYSTEM IMPLANTED. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11766719 ↗
  • InjuryMar 10, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 11454269 ↗
  • MalfunctionFeb 8, 2021Mechanical Problem; Under-Sensing; Mechanical Problem; Under-Sensing

    IT WAS REPORTED THAT THIS LEAD HAD A GRADUAL DROP IN SENSING OVER TIME, AND THEN A SUDDEN DROP BELOW 2MV IN (B)(6) 2020. THE LEAD WAS CAPPED AFTER THE SCREW WAS NOT ABLE TO BE RETRACTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11297991 ↗
  • InjuryJan 6, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS ADMITTED TO THE HOSPITAL WITH CHEST PAIN 12 DAYS AFTER THE INITIAL IMPLANT. THIS LEAD WAS EXPLANTED AND REPLACED DUE TO AN RV APEX PERFORATION. THE REVISION PROCEDURE WENT WELL AND THE PATIENT WAS DOING WELL THE DAY AFTER REVISION. SHOULD ADDITIONAL INFORMATION BE RECEIVED THE FILE WILL BE UPDATED.

    Patient problems: Cardiac Perforation

    MAUDE report 11127975 ↗
  • MalfunctionDec 3, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    LEAD DISLODGED ONE DAY AFTER IMPLANT. PHYSICIAN ATTEMPTED TO REPOSITION THE LEAD, HOWEVER THEY DISCOVERED TISSUE IN HELIX AND THEREFORE CHOSE TO IMPLANT NEW LEAD. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10938629 ↗
  • InjuryNov 10, 2020Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Device Dislodged or Dislocated; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Device Dislodged or Dislocated

    LEAD WAS EXPLANTED DUE TO DISLODGEMENT, INAPPROPRIATE SHOCKS, AND OVERSENSING. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 10819969 ↗
  • InjuryOct 22, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED AND REPLACED DUE TO DISLODGEMENT AND POSSIBLE MICRO PERFORATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Perforation

    MAUDE report 10720814 ↗
  • MalfunctionSep 25, 2020Device Dislodged or Dislocated; High Capture Threshold; Device Dislodged or Dislocated; High Capture Threshold

    DEVICE INTERROGATION ON (B)(6) 2020 SHOWED HIGH LEAD THRESHOLDS. LEAD WAS FOUND TO BE DISLODGED, AND PATIENT UNDERWENT REVISION ON (B)(6) 2020. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10590809 ↗
  • InjurySep 3, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    LEAD WAS REVISED DUE TO DISLODGEMENT. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10491400 ↗
  • MalfunctionSep 2, 2020Break; Ambient Noise Problem; Break; Ambient Noise Problem

    NOISE ON LEAD, OUTER INSULATION BREACH. LEAD WAS EXPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10484021 ↗
  • InjuryAug 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED BECAUSE THE PATIENT SEEMED TO BE ALLERGIC TO THE EPOXY AND RESINS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Skin Inflammation

    MAUDE report 10393590 ↗
  • MalfunctionJun 15, 2020High impedance; High Capture Threshold; High impedance; High Capture Threshold

    LEAD WAS EXPLANTED AND REPLACED DUE TO HIGH THRESHOLDS AND IMPEDANCES. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10155374 ↗

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →