Protego SD 60/16 399408Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Protego SD permanent defibrillator lead sits under the skin and wires the heart to a defibrillator. The 43 reports mostly describe noise picked up on the lead that led to shocks delivered when none were needed, plus cracks or breaks in the lead itself and insulation damage, sometimes requiring the lead to be removed or capped. Reports mention 16 injuries and no deaths, with 27 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
0
Injuries reported
16
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 04035479136093), received between Jul 19, 2018 and Mar 6, 2026 — classified as 16 injuries, 27 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Ambient Noise Problem22
  • Fracture12
  • Inappropriate/Inadequate Shock/Stimulation10
  • Adverse Event Without Identified Device or Use Problem4
  • Over-Sensing4
  • Impedance Problem3
  • Defibrillation/Stimulation Problem1
  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Shock from Patient Lead(s)11
  • No Known Impact Or Consequence To Patient9
  • Unspecified Infection4
  • Ventricular Fibrillation2
  • Fall1
  • Ventricular Tachycardia1

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 6, 2026Fracture

    THIS RV LEAD WAS EXPLANTED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24537794 ↗
  • InjuryJan 22, 2026Ambient Noise Problem

    THIS RV LEAD SHOWS NOISE WHICH LED TO INAPPROPRIATE THERAPIES. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24151440 ↗
  • InjuryJan 12, 2026Fracture; Inappropriate/Inadequate Shock/Stimulation

    THIS LEAD WAS EXPLANTED DUE TO FRACTURE RESULTING IN INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24037087 ↗
  • MalfunctionDec 11, 2025Fracture; Fracture

    THIS LEAD WAS EXPLANTED DUE TO LEAD INSULATION BREACH AND FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23779969 ↗
  • MalfunctionMay 2, 2025Ambient Noise Problem; Ambient Noise Problem

    NOISE SEEN ON RV LEAD IN HMSC RECORDINGS. NOISE WAS NOT ABLE TO BE REPRODUCED IN OFFICE. EVALUATION OF LEAD IS NEEDED AND XRAY WAS ORDERED. DEVICE REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21950045 ↗
  • InjuryJan 17, 2025Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS RV LEAD WAS CAPPED DUE TO NOISE RESULTING IN INAPPROPRIATE SHOCK. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 21178826 ↗
  • MalfunctionNov 20, 2024Fracture; Fracture

    THIS LEAD WAS EXPLANTED DUE TO SUSPECTED FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20733945 ↗
  • MalfunctionAug 22, 2024Over-Sensing; Ambient Noise Problem; Over-Sensing; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO INTERMITTENT NOISE NOTED ON RV PACE SENSE. PATIENT REPORTED A FALL WITH POSSIBLE CLAVICLE FRACTURE/DISLOCATION ABOUT SIX WEEKS PRIOR TO OVERSENSING. NEW RV PACE SENSE LEAD ADDED TO LEFT SIDE AND TUNNELED TO EXISTING RIGHT SIDED SYSTEM AND DOWNGRADE TO CRT-P. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Fall

    MAUDE report 20039170 ↗
  • InjuryMay 2, 2024Fracture; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Fracture; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO FRACTURE AND NOISE CAUSING INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 19228445 ↗
  • MalfunctionApr 30, 2024Impedance Problem; Impedance Problem

    THIS LEAD WAS CAPPED DUE TO IMPEDANCE OUT OF RANGE. PATIENT DOWNGRADED TO A CRTP SYSTEM. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19217402 ↗
  • InjuryApr 15, 2024Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem; Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE RESULTING IN INAPPROPRIATE THERAPY. THERE WAS NOISE WITH MOVING ARM ABOVE HEAD. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 19111758 ↗
  • MalfunctionApr 1, 2024Ambient Noise Problem; Ambient Noise Problem

    THIS LEAD WAS CAPPED DUE TO NOISE DETECTING VF. THRESHOLD AND IMPEDANCE NORMAL. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19015212 ↗
  • MalfunctionFeb 15, 2024Fracture; High impedance; Fracture; High impedance

    THIS LEAD WAS CAPPED DUE TO HIGH IMPEDANCE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18710679 ↗
  • MalfunctionDec 11, 2023Ambient Noise Problem; Ambient Noise Problem

    LEAD WAS EXPLANTED DUE TO NOISE AFTER CAR ACCIDENT WITH SEATBELT ON. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18307394 ↗
  • InjuryNov 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 18069762 ↗
  • MalfunctionJun 21, 2023Ambient Noise Problem; Ambient Noise Problem

    THIS LEAD WAS EXPLANTED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17169861 ↗
  • MalfunctionJun 6, 2023Impedance Problem; Impedance Problem

    HMSC WEBSITE SHOWS OUT OF RANGE SHOCK IMPEDANCE SINCE (B)(6) 2021. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17075207 ↗
  • MalfunctionJul 8, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS RV LEAD WAS CAPPED AND REPLACED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14955951 ↗
  • MalfunctionMay 24, 2022Ambient Noise Problem; Ambient Noise Problem

    INTERMITTENT NOISE WAS REPORTED ON THE RV CHANNEL IN RECORDINGS TRANSMITTED TO HOME MONITORING. NO INAPPROPRIATE THERAPIES WERE REPORTED. LEAD TO BE EVALUATED FURTHER AND REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14492383 ↗
  • MalfunctionMar 19, 2022Over-Sensing; Ambient Noise Problem; Over-Sensing; Ambient Noise Problem

    THIS LEAD WAS EXPLANTED DUE TO NOISE AND OVERSENSING. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13826251 ↗
  • InjuryFeb 17, 2022Fracture; Inappropriate/Inadequate Shock/Stimulation; Fracture; Inappropriate/Inadequate Shock/Stimulation

    THIS RV LEAD WAS EXPLANTED AND REPLACED DUE TO A FRACTURE. THE PATIENT RECEIVED INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 13549492 ↗
  • InjuryFeb 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LEAD EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: Unspecified Infection

    MAUDE report 13440388 ↗
  • MalfunctionJun 2, 2021Fracture; Fracture

    LEAD CAPPED DUE TO RV LEAD FRACTURE. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11925360 ↗
  • MalfunctionMay 25, 2021Fracture; Ambient Noise Problem; Fracture; Ambient Noise Problem

    LEAD CAPPED DUE TO FRACTURE CAUSING NOISE. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11881997 ↗
  • InjuryApr 9, 2021Failure to Deliver Shock/Stimulation; Low impedance; Failure to Deliver Shock/Stimulation; Low impedance

    LEAD REMAINS IMPLANTED. REPRESENTATIVE REPORTED AN ABORTED SHOCK DURING A VF EPISODE DUE TO SHOCK IMPEDANCE LESS THAN 20 OHMS. LEAD TRENDS APPEARED STABLE. FULL LEAD EVALUATION IN OFFICE RECOMMENDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: Ventricular Fibrillation

    MAUDE report 11642550 ↗

Other Permanent defibrillator electrodes devices