Plexa ProMRI DF-1 S 65Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Plexa ProMRI DF-1 S65 is an implanted permanent defibrillator lead, the wire that carries shock therapy to the heart. The 65 reports mostly describe leads that picked up stray signals or background noise, shifted position after implantation, or showed rising electrical resistance, and several mention infection that led to the whole system being taken out. Reports include 2 deaths and 21 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
65
Deaths reported
2
Injuries reported
21
Recalls naming it
0

Summary

The FDA's MAUDE database contains 65 adverse event reports that list this exact device (DI 04035479150969), received between May 3, 2019 and Jun 30, 2026 — classified as 2 deaths, 21 injuries, 42 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem16
  • Over-Sensing12
  • Device Dislodged or Dislocated9
  • Ambient Noise Problem6
  • Impedance Problem6
  • High impedance5
  • Break3
  • Fracture3

Patient problems

  • No Clinical Signs, Symptoms or Conditions38
  • Unspecified Infection13
  • No Known Impact Or Consequence To Patient6
  • Shock from Patient Lead(s)2
  • Cardiac Perforation1
  • Cardiac Tamponade1
  • Chest Pain1
  • Death1

Recent adverse event reports

The 25 most recent of 65 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 30, 2026Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25667269 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24867756 ↗
  • InjuryMar 10, 2026Device Dislodged or Dislocated

    THIS LEAD WAS REPOSITIONED DUE TO DISLODGEMENT AND REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24564159 ↗
  • MalfunctionFeb 23, 2026Impedance Problem

    HMSC REPORTED SHOCK IMPEDANCE GRADUALLY INCREASING OVER THE PAST YEAR. ALL OTHER LEAD TRENDS WERE UNREMARKABLE. NO NOISE EVIDENT. ADVISED FURTHER LEAD EVALUATION. LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24420977 ↗
  • InjuryFeb 3, 2026Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24249119 ↗
  • MalfunctionNov 5, 2025Fracture; Fracture

    THE P/S PORTION OF THIS LEAD WAS CAPPED DUE TO FRACTURE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23478441 ↗
  • InjuryNov 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 23472100 ↗
  • InjuryAug 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 22844226 ↗
  • MalfunctionJul 21, 2025High impedance; High impedance

    THIS LEAD WAS CAPPED DUE TO CHRONIC HIGH IMPEDANCE, >3000 OHMS. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22561347 ↗
  • MalfunctionJul 15, 2025Break; Impedance Problem; Break; Impedance Problem

    LEAD WAS EXPLANTED AND REPLACED BECAUSE OF OOR B COIL (RV) IMPEDANCE. UPON POCKET OPENING, THE RV COIL (DF1) SEGMENT WAS COMPLETELY SEPARATED FROM THE LEAD AT EXACTLY THE LEAD YOKE EXIT. THE LEAD SEGMENT APPEARED TO BE CUT. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22515585 ↗
  • InjuryJan 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 21224652 ↗
  • MalfunctionJan 13, 2025Device Sensing Problem; Device Sensing Problem

    THE PACING PORTION OF THIS LEAD WAS CAPPED DUE TO NO SENSING WITH FINE VF. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21145493 ↗
  • MalfunctionJan 9, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21113835 ↗
  • InjuryJun 20, 2024Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    THIS RV LEAD WAS EXPLANTED AFTER FAILED ATTEMPT TO REPOSITION LEAD DUE TO HELIX WOULD NOT RETRACT. PATIENT HAD HAD A SYNCOPAL EPISODE AND VF WITH FAILED ICD SHOCK, AND DURING (B)(6) 2024 EP VISIT FOR CARDIOVERSION, DFT RESULTED IN THREE FAILED SHOCKS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Syncope/Fainting; Ventricular Fibrillation

    MAUDE report 19577102 ↗
  • MalfunctionMay 20, 2024Fracture; Ambient Noise Problem; Fracture; Ambient Noise Problem

    LEAD WAS EXPLANTED DUE TO NOISE POSSIBLY FROM FRACTURE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19354772 ↗
  • MalfunctionApr 25, 2024Ambient Noise Problem; Ambient Noise Problem

    RV LEAD IS SHOWING NOISE VIA HOME MONITORING. ALL LEAD TRENDS APPEAR STABLE. LEAD TO BE EVALUATED FURTHER. LEAD CURRENTLY REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19185416 ↗
  • MalfunctionMar 18, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18921680 ↗
  • InjuryMar 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ICD AND RV LEAD WERE EXPLANTED DUE TO POCKET INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 18866754 ↗
  • MalfunctionOct 17, 2023Ambient Noise Problem; Ambient Noise Problem

    NOISE SEEN ON FAR FIELD SIGNAL. DISCUSSED POSSIBILITY OF EXTERNAL EMI OR LEAD INSULATION ISSUES. PATIENT WILL BE BROUGHT IN FOR FURTHER EVALUATION AND DISCUSSION WITH EP. DEVICE REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17953902 ↗
  • MalfunctionJun 28, 2023Ambient Noise Problem; Ambient Noise Problem

    THIS LEAD SHOWS NOISE. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17218062 ↗
  • MalfunctionJun 12, 2023High impedance; High impedance

    LEAD SHOWED HIGH SHOCK IMPEDANCE (>150 OHMS) UPON DISCHARGE CHECK. PATIENT RETURNED TO HOSPITAL ON (B)(6) 2023 AND THE SET SCREW WAS TIGHTENED WHICH RESOLVED THE IMPEDANCE ISSUE. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17114021 ↗
  • InjuryNov 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SYSTEM EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15886470 ↗
  • MalfunctionOct 18, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT FOUND BY X-RAY. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15620955 ↗
  • InjurySep 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 15417264 ↗
  • MalfunctionJun 22, 2022Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THIS LEAD WAS EXPLANTED DUE TO OVERSENSING APPROX. 14 MONTHS AFTER THE IMPLANTATION. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14783258 ↗

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →