Plexa ProMRI DF-1 S DX 65/15Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

All models of Plexa ProMRI (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Plexa ProMRI lead is a permanent defibrillator lead. Its 651 reports mostly describe electrical noise on the lead, with some reports of inappropriate shocks and leads being removed. The reports include 228 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
652
Deaths reported
0
Injuries reported
227
Recalls naming it
0

Summary

The FDA's MAUDE database contains 652 adverse event reports that list this exact device (DI 04035479151027), received between Jun 14, 2018 and Sep 30, 2026 — classified as 227 injuries, 425 malfunctions.

The most frequently reported specific device problem is “Ambient Noise Problem” (283 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 1 reportsJan 2024: 8 reportsFeb 2024: 1 reportsMar 2024: 6 reportsApr 2024: 10 reportsMay 2024: 3 reportsJun 2024: 4 reportsJul 2024: 5 reportsAug 2024: 10 reportsSep 2024: 5 reportsOct 2024: 4 reportsNov 2024: 5 reportsDec 2024: 1 reportsJan 2025: 7 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 3 reportsMay 2025: 10 reportsJun 2025: 4 reportsJul 2025: 5 reportsAug 2025: 3 reportsSep 2025: 4 reportsOct 2025: 4 reportsNov 2025: 9 reportsDec 2025: 3 reportsJan 2026: 2 reportsFeb 2026: 9 reportsMar 2026: 9 reportsApr 2026: 7 reportsMay 2026: 3 reportsJun 2026: 5 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 4 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Ambient Noise Problem283
  • Adverse Event Without Identified Device or Use Problem107
  • Inappropriate/Inadequate Shock/Stimulation91
  • Over-Sensing81
  • Fracture64
  • High impedance64
  • Device Dislodged or Dislocated55
  • High Capture Threshold25

Patient problems

  • No Clinical Signs, Symptoms or Conditions348
  • No Known Impact Or Consequence To Patient92
  • Shock from Patient Lead(s)90
  • Unspecified Infection90
  • Twiddlers Syndrome14
  • Ventricular Fibrillation6
  • Cardiac Perforation4
  • Pocket Erosion4

Recent adverse event reports

The 25 most recent of 652 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Ambient Noise Problem; Device Sensing Problem; Unstable Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50955015 ↗
  • InjurySep 29, 2026Fracture; Ambient Noise Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940688 ↗
  • MalfunctionSep 29, 2026Ambient Noise Problem

    HMSC reported recordings showing noise which resulted in VF detections. RV short interval counter increased >249. All other lead trends appear stable. Lead currently remains implanted. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932652 ↗
  • InjurySep 25, 2026Ambient Noise Problem

    Lead was explanted due to Noise. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898907 ↗
  • InjuryAug 13, 2026Ambient Noise Problem; Impedance Problem

    This lead was explanted due to noise and impedance out of range. No adverse patient events were reported. Should additional information be received this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50235091 ↗
  • MalfunctionJul 30, 2026Ambient Noise Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50084179 ↗
  • InjuryJul 7, 2026Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    THIS LEAD WAS EXPLANTED DUE TO NOISE RESULTING IN INAPPROPRIATE SHOCKS. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 25739111 ↗
  • MalfunctionJun 29, 2026Ambient Noise Problem

    ATRIAL MONITORING RECORDING FROM (B)(6) 2026 SHOWS HIGH FREQUENCY NOISE ON THE RA AND FF CHANNELS. RV CHANNEL IS CLEAN. SUSPECTED EMI. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25652974 ↗
  • InjuryJun 10, 2026Impedance Problem; Unstable Capture Threshold

    LEAD WAS CAPPED DUE TO IMPEDANCE AND THRESHOLD OUT OF RANGE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25461207 ↗
  • InjuryJun 9, 2026Break; Ambient Noise Problem; Impedance Problem

    THIS RV LEAD WAS EXPLANTED DUE TO NOISE, HIGH PACING IMPEDANCE, INCREASING SHOCK IMPEDANCE AND IS-1 PIN DISCONNECTED FROM RING. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25451372 ↗
  • MalfunctionJun 9, 2026Ambient Noise Problem

    NOISE IS EVIDENT ON RV CHANNEL AND FF CHANNEL. SOME NOISE IS REPORTED ON ATRIAL CHANNEL. SHORT INTERVAL COUNTER SHOWS ABOVE 249. FULL LEAD EVALUATION IS RECOMMENDED. LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25451254 ↗
  • InjuryJun 4, 2026Ambient Noise Problem

    LEAD WAS EXPLANTED DUE TO ATRIAL AND VENTRICULAR NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25401395 ↗
  • InjuryMay 21, 2026Ambient Noise Problem; Device Dislodged or Dislocated; High Capture Threshold

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT RESULTING IN HIGH RV THRESHOLD AND RA NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25245772 ↗
  • InjuryMay 20, 2026Ambient Noise Problem

    THIS LEAD WAS EXPLANTED DUE TO NOISE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25234686 ↗
  • InjuryMay 5, 2026Inappropriate/Inadequate Shock/Stimulation; Ambient Noise Problem

    NOISE ON FF, RA, AND RV CHANNELS LEADING TO INAPPROPRIATE SHOCK DELIVERED. LEAD CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 25088507 ↗
  • InjuryApr 30, 2026Over-Sensing; Ambient Noise Problem

    NOISE ON LEAD. RV NOISE CAUSED OVER SENSING. THE LEAD WAS CAPPED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25051726 ↗
  • MalfunctionApr 22, 2026Ambient Noise Problem

    HMSC REPORTED EPISODES SHOWING NOISE ON THE RV CHANNEL. TREND GRAPHS SHOW INCREASES IN SHORT INTERVAL COUNTER. ALL OTHER LEAD TRENDS APPEAR STABLE. ADVISED FURTHER LEAD EVALUATION. LEAD CURRENTLY REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24952735 ↗
  • InjuryApr 17, 2026Over-Sensing; Ambient Noise Problem

    LEAD WAS CAPPED DUE TO NOISE OVERSENSING. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24915839 ↗
  • InjuryApr 14, 2026Ambient Noise Problem

    NOISE SEEN DURING THRESHOLD TESTING. PACE/SENSE PORTION OF LEAD CAPPED AND CSP LEAD IMPLANTED. SHOCKING PORTION OF LEAD REMAINS ACTIVE AND ATTACHED TO NEW HFT DEVICE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872345 ↗
  • MalfunctionApr 14, 2026Fracture; Ambient Noise Problem

    PATIENT HAD WHAT APPEARED TO BE TRUE EPISODES OF AT/VT FOLLOWED BY A JUMP IN RV SHORT INTERVALS AND NST/VF EPISODES SHOWING LEAD NOISE ON ALL CHANNELS (B)(6) 2023. LAST MESSAGE ON HOME MONITORING WAS (B)(6) 2023. NO IMPEDANCE TRENDS OBSERVED. THE NOISE COULD BE REPRODUCED AND A FRACTURE IS POSSIBLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872191 ↗
  • InjuryApr 13, 2026Ambient Noise Problem

    THIS RV LEAD WAS CAPPED DUE TO NOISE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24867780 ↗
  • MalfunctionApr 9, 2026Ambient Noise Problem

    NOISE ON THE RV AND FF CHANNELS CAN BE SEEN. THE NOISE ON THE RV CHANNEL BEGAN CAUSING INAPPROPRIATE NST DETECTIONS ON (B)(6) 2026. TWO INAPPROPRIATE VF DETECTIONS WERE MADE DUE TO NOISE, BUT SHOCKS WERE ABORTED IN BOTH CASES. INTERMITTENT NOISE COULD BE RECREATED IN PERSON ON (B)(6) 2026. NO ADVERSE EVENTS REPORTED. LEAD REMAINS IMPLANTED. IT WAS REPORTED THE PATIENT HAD A HEART CATHETERIZATION ON (B)(6) 2026. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24835850 ↗
  • MalfunctionMar 30, 2026Ambient Noise Problem

    NOISE IS EVIDENT ON RV CHANNEL AND FAR FIELD CHANNEL. LEAD CURRENTLY REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24720498 ↗
  • InjuryMar 26, 2026Ambient Noise Problem

    THIS RV LEAD WAS EXPLANTED DUE TO PATIENT HAD A MOTOR VEHICLE ACCIDENT WITH LEFT SHOULDER INJURY AND SUBSEQUENT LEAD NOISE. ICD THERAPIES HAD BEEN PROGRAMMED OFF SINCE FALL 2025. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698862 ↗
  • MalfunctionMar 24, 2026Ambient Noise Problem

    SHORT INTERVAL COUNT ELEVATED ON RV LEAD COMPLEXES APPEAR ON BOTH RV CHANNEL AND FAR FIELD CHANNEL. NO ADVERSE PATIENT EVENTS WERE REPORTED. LEAD REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24675710 ↗

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →