AHTO 0250070620Recalls & reported problems
Stryker Corporation · Disposable Suction/Irrigator Tube Set with Tip · Class 2
All models of AHTO (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Stryker suction/irrigator tube set is a disposable tube used for suction and irrigation during surgery. Its 59 reports mostly describe contaminated packaging, plastic or rubber fragments, and leaking or loss of pressure, with no deaths and 1 injury mentioned. There are 2 recalls, including an open one about leaking from the handpiece during use and a closed one about high bioburden levels in some lots.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 59
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 59 adverse event reports that list this exact device (DI 07613327061659), received between Jun 14, 2019 and Aug 31, 2026 — classified as 1 injury, 58 malfunctions.
The most frequently reported specific device problem is “Device Contaminated During Manufacture or Shipping” (18 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 8, 2026Class IIOpen, Classified
There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.
Z-2863-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Nov 17, 2017Class IITerminated
Routine bioburden testing of certain lots were found to have levels higher than internal acceptable rates.
Z-0836-2018 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Contaminated During Manufacture or Shipping18
- Material Fragmentation11
- Decrease in Pressure8
- Material Puncture/Hole8
- Mechanical Problem6
- Contamination /Decontamination Problem5
- Leak/Splash5
- Delivered as Unsterile Product4
Patient problems
- No Clinical Signs, Symptoms or Conditions50
- No Known Impact Or Consequence To Patient9
- Insufficient Information6
- Foreign Body In Patient4
Recent adverse event reports
The 25 most recent of 59 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2026Device Contaminated During Manufacture or Shipping
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50532314 ↗MalfunctionJul 27, 2026Material Fragmentation
It was reported that during a procedure using a Stryker AHTO 250-070-620 irrigation/suction set connected to a B. Braun Ecobag 0.9% NaCl 3000 ml tubing, the scrub nurse noticed a piece of gray plastic in the cup after purging the system.No serious consequences for the patient.Please note, the NaCl irrigation solution bag is not a Stryker product. It was alleged that this issue occurred when using the Stryker AHTO tubeset.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50000055 ↗MalfunctionJun 3, 2026Material Fragmentation
IT WAS REPORTED THAT DURING AN EMERGENCY SURGICAL PROCEDURE (LAPAROSCOPIC APPENDECTOMY), THE SURGICAL TEAM IDENTIFIED A LOOSE PIECE OF RUBBER IN THE ABDOMINAL CAVITY. THIS FRAGMENT ORIGINATED FROM A PUNCTURE IN THE STOPPER OF THE 0.9% NACL IRRIGATION SOLUTION BAG (REF (B)(4) AND BATCH 25493767) DURING CONNECTION TO THE STRYKER AHTO IRRIGATION TUBING (REF (B)(4), BATCH UNKNOWN). THE FRAGMENT WAS RETRIEVED. DURING THE PROCEDURE, WHEN A SECOND BAG OF THE SAME REFERENCE NUMBER WAS PUNCTURED, THE CORE SAMPLING PHENOMENON WAS OBSERVED ONCE AGAIN. THE FRAGMENT SPILLED OUT WITH THE NACL SOLUTION WHILST FILLING A CUP. NO SERIOUS CONSEQUENCES FOR THE PATIENT. PLEASE NOTE, THE NACL IRRIGATION SOLUTION BAG IS NOT A STRYKER PRODUCT. IT WAS ALLEGED THAT THIS ISSUE OCCURRED WHEN USING THE STRYKER AHTO TUBESET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25383650 ↗MalfunctionJun 3, 2026Material Fragmentation
IT WAS REPORTED THAT DURING AN EMERGENCY SURGICAL PROCEDURE (LAPAROSCOPIC APPENDECTOMY), THE SURGICAL TEAM IDENTIFIED A LOOSE PIECE OF RUBBER IN THE ABDOMINAL CAVITY. THIS FRAGMENT ORIGINATED FROM A PUNCTURE IN THE STOPPER OF THE 0.9% NACL IRRIGATION SOLUTION BAG (REF (B)(4) AND BATCH 25493767) DURING CONNECTION TO THE STRYKER AHTO IRRIGATION TUBING (REF (B)(4), BATCH UNKNOWN). THE FRAGMENT WAS RETRIEVED. DURING THE PROCEDURE, WHEN A SECOND BAG OF THE SAME REFERENCE NUMBER WAS PUNCTURED, THE CORE SAMPLING PHENOMENON WAS OBSERVED ONCE AGAIN. THE FRAGMENT SPILLED OUT WITH THE NACL SOLUTION WHILST FILLING A CUP. NO SERIOUS CONSEQUENCES FOR THE PATIENT. PLEASE NOTE, THE NACL IRRIGATION SOLUTION BAG IS NOT A STRYKER PRODUCT. IT WAS ALLEGED THAT THIS ISSUE OCCURRED WHEN USING THE STRYKER AHTO TUBESET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25383648 ↗MalfunctionFeb 9, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT THE PACKAGE CONTAMINATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24300023 ↗MalfunctionFeb 4, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A LONG BLACK HAIR WAS FOUND INSIDE THE PACKAGING..
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24260868 ↗MalfunctionFeb 3, 2026Material Fragmentation
IT WAS REPORTED THAT DURING THE PROCEDURE, THE STRYKER SUCTION IRRIGATION DEVICE BROKE AT THE BASE OF THE SUCTION TIP. NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24252857 ↗MalfunctionJan 5, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT THE PACKAGE WAS CONTAMINATED ON THE INSIDE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23957579 ↗MalfunctionNov 26, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT BLACK ITEM WAS FOUND IN THE AHTO TUBESET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23660665 ↗MalfunctionNov 26, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT WHEN THE INFUSION WAS PUNCTURED, A PIECE OF RUBBER WAS PUNCHED OUT AND ENTERED THE INFUSION FLUID. THE INFUSION WHERE THE PIECE OF RUBBER WAS PUNCHED OUT IS A B. BRAUN BAG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23660120 ↗MalfunctionNov 17, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT WHILE PULLING PRODUCTS FOR CASE SETUP, THE DISPOSABLE SUCTION/IRRIGATOR TUBE SET WITH TIP WAS NOTED TO BE CONTAMINATED WITH MULTIPLE BLACK/BLUE/WHITE SPECKS IN BOTTOM OF PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23577054 ↗MalfunctionNov 17, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING CASE SETUP, THE DISPOSABLE SUCTION/IRRIGATOR TUBE SET WITH TIP WAS NOTED TO BE CONTAMINATED WITH MULTIPLE BLACK/BLUE/WHITE SPECKS IN BOTTOM OF PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23576002 ↗MalfunctionNov 17, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING CASE SETUP, THE DISPOSABLE SUCTION/IRRIGATOR TUBE SET WITH TIP WAS NOTED TO BE CONTAMINATED WITH MULTIPLE BLACK/BLUE/WHITE SPECKS IN BOTTOM OF PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23575958 ↗MalfunctionOct 30, 2025Fluid/Blood Leak
SUCTION IRRIGATOR WAS LEAKING IRRIGATION FLUID FROM HANDLE HUB SITE, EVEN WHEN NOT IN USE.NO IDENTIFIED HARM TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23424940 ↗MalfunctionOct 24, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT HAIR WAS DISCOVERED INSIDE THE PACKAGING AFTER OPENING THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23382243 ↗MalfunctionOct 23, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE DEVICE HAD COMPROMISED PACKAGING. THEREFORE, THERE WAS BREACH IN THE STERILE PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23374952 ↗MalfunctionOct 23, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE DEVICE HAD COMPROMISED PACKAGING. THEREFORE, THERE WAS BREACH IN THE STERILE PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23374950 ↗MalfunctionOct 23, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT HAIR WAS DISCOVERED INSIDE THE PACKAGING AFTER OPENING THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23374522 ↗MalfunctionOct 23, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT HAIR WAS DISCOVERED INSIDE THE PACKAGING AFTER OPENING THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23374521 ↗MalfunctionOct 16, 2025Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THAT THE PISTON TAKES A LONG TIME TO RETURN, RESULTING IN CONTINUOUS SUCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23318906 ↗MalfunctionAug 20, 2025Contamination; Device Contaminated During Manufacture or Shipping; Contamination; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A SMALL RED PLASTIC PIECE WAS FOUND INSIDE OF THE PACKAGE. THERE WAS NO SIGNS OF DAMAGE FOUND ON THE SUCTION IRRIGATOR TUBE SET. UNSURE OF WHERE THE RED PLASTIC PIECE CAME FROM. DEEMED CONTAMINATED DUE TO UNKNOWN FOREIGN OBJECT ORIGIN. NEW PRODUCT WAS OBTAINED AND USED. PER OR MATERIALS MANAGEMENT: RED PIECE APPEARS TO BE BROKEN OFF FROM INTERNAL COMPONENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22850258 ↗MalfunctionJul 9, 2025Leak/Splash; Decrease in Pressure; Leak/Splash; Decrease in Pressure
IT WAS REPORTED THAT THE DEVICE LOSES WATER THROUGH THE BLUE BUTTON, AND THE RED ONE GETS STUCK AND STAYS OPEN WHILE VACUUMING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22461514 ↗MalfunctionJul 9, 2025Leak/Splash; Decrease in Pressure; Leak/Splash; Decrease in Pressure
IT WAS REPORTED THAT THE DEVICE LOSES WATER THROUGH THE BLUE BUTTON, AND THE RED ONE GETS STUCK AND STAYS OPEN WHILE VACUUMING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22461510 ↗MalfunctionJul 9, 2025Leak/Splash; Decrease in Pressure; Leak/Splash; Decrease in Pressure
IT WAS REPORTED THAT THE DEVICE LOSES WATER THROUGH THE BLUE BUTTON, AND THE RED ONE GETS STUCK AND STAYS OPEN WHILE VACUUMING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22461500 ↗MalfunctionJul 9, 2025Leak/Splash; Decrease in Pressure; Leak/Splash; Decrease in Pressure
IT WAS REPORTED THAT THE DEVICE LOSES WATER THROUGH THE BLUE BUTTON, AND THE RED ONE GETS STUCK AND STAYS OPEN WHILE VACUUMING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22461499 ↗
Other LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) devices
- Olympus WA50042AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)1 recall6,135 reports
- Olympus WA50052AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)2,249 reports
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5Olympus Medical Systems Corp. · LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)1 recall1,279 reports
- Olympus WA50040AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)1 recall954 reports
- Olympus WA50050AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)895 reports
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5Olympus Medical Systems Corp. · Laparoscope, Gynecologic (And Accessories)624 reports
- Epix Laparoscopic Scissors CB030Applied Medical Resources Corporation · Laparoscope, Gynecologic (And Accessories)239 reports
- HD EndoEYE LAPARO-THORACO VIDEOSCOPE OLYMPUS LTF TYPE VHOlympus Medical Systems Corp. · LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)168 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page